    1529 Chub-E4 Thermometer Readout User’ s Guide Rev. 5A2602 Hart Scientific ...

  • Fluke 1529 Chub-E4 - page 2

    Rev. 5A2602 Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. • Copyright © 2005 • Printed in USA ...

  • Fluke 1529 Chub-E4 - page 3

    T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W arnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 Cautio ...

  • Fluke 1529 Chub-E4 - page 4

    6.5 DC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . 25 6.6 Power On Self-T est . . . . . . . . . . . . . . . . . . . . . . . . . 25 6.7 Display Backlight and Contrast . . . . . . . . . . . . . . . . . . . 25 6.8 T aking Measurements . . . . . . . . . . . . . . . . . . . . . . . . 25 6.8.1 Connecting the Sensor . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 5

    7.4.2.4 Print Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 7.4.2.5 Delete Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 7.4.3 Log Stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 7.4.4 Data Labels . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 6

    8.4.3.8 R OUT e:SCAN:MODE? [MIN|MAX|DEF] . . . . . . . . . . . . . . . . . . . . . . 95 8.4.3.9 R OUT e:SCAN:MODE <num>|MIN|MAX|DEF . . . . . . . . . . . . . . . . . . . . 95 8.4.4 Probe Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 8.4.4.1 CALCulate<chn>:CONV ert:CA T alog? . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 7

    8.4.8 System Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 8.4.8.1 *IDN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 8.4.8.2 *OPT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 8.4.8.3 *RST . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 8

    8.4.12.16 ST A Tus:OPERation:EN ABle <num>|MIN|MAX|DEF . . . . . . . . . . . . . . . . 124 8.4.12.17 ST A Tus:QUEStionable ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 8.4.12.18 ST A Tus:QUEStionable :CONDition? . . . . . . . . . . . . . . . . . . . . . . . . . 125 8.4.12.19 ST A Tus:QUEStionable :EN ABle? [MIN|MAX|DE ...

  • Fluke 1529 Chub-E4 - page 9

    vii Figures Figure 1 Using the Clamp-on Ferrites . . . . . . . . . . . . . . . . . . . . . . . 14 Figure 2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Figure 3 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Figure 4 Thermocouple Connections . . . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 10

    viii Figure 41 Default Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Figure 42 System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 Figure 43 Comm Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 Figure 44 Date T ime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 11

    ix T ables T able1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 T able 2 Con version T ypes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 T able 3 Alphabetical List of Commands . . . . . . . . . . . . . . . . . . . . . 81 T able 4 Statistical Calculation T ypes . . . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 1529 Chub-E4 - page 12

    1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the symbols that may be used on t he instrument or in thi s manual and the meaning of each symbol. Symbol Description AC ( Alternating Current) AC- DC Batter y Complies with European Uni on Directiv es DC (Direct Current ) Double Insulated Electric Shock Fuse PE Ground Hot Surf ace (Bur n Hazard) ...

  • Fluke 1529 Chub-E4 - page 13

    Symbol Description On Canadian Standards Associati on O VERV OL T AGE (Inst allation) CA TEGOR Y II, P ollution Degree 2 per IEC1010- 1 re - f ers to the le vel of Impulse Wit hstand V oltage protection provided. Equipment of O VERV OL T AGE CA TEGORY I I is energy-consuming equipment to be supplied from the fix ed installat ion. Examples include h ...

  • Fluke 1529 Chub-E4 - page 14

    provided with the instrument or equiv alent adapter recommended by the manufacturer of this instrument. • The A C adapter has circuits with high voltage inside that could present danger of electrical shock or fire if exposed. If the A C adapter is damaged in any way or becomes hot, discontinue its use immediately , disconnect it from any A C supp ...

  • Fluke 1529 Chub-E4 - page 15

    • Keep the probe wires clean and aw ay from fluids. 1.3 A uthoriz ed Service Centers Please contact one of the follo wing authorized Service Centers to coordinate service on your Hart product: Fluke Corporation, Hart Scientif ic Division 799 E. Utah V alle y Driv e American Fork, UT 84003-9775 USA Phone: +1.801.763.1600 T elefax: +1.801.763.1010 ...

  • Fluke 1529 Chub-E4 - page 16

    Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Of fice Service Center 60 Alexandra T errace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 T elefax: +65 6799-5588 E-mail: antng@singa.fluke.com When contacting these Service Cente rs for support, please hav e the follo wing information av ailable: • Model Number • Serial ...

  • Fluke 1529 Chub-E4 - page 17

    2 Intr oduction The Hart 1529 is a low-cost, high-accurac y , digital thermometer readout de - signed to be used with 25 and 100 Ω PR Ts, thermistors, and thermocouples. Its unique combination of features makes it suitable for a wide v ariety of applica - tions from laboratory measurement to industrial processes. Features of the 1529 include: • ...

  • Fluke 1529 Chub-E4 - page 18

    3 Specifications and En vir onmental Conditions 3.1 Specifications PRT Thermistor Thermocouple Inputs 2 channels PRT/ther mistor and 2 channels thermocoupl e or 4 channels PRT/thermistor or 4 channels thermocouple PRT/thermistor channel s accept 2,3, or 4 wires Thermocouple channels accept B, E, J, K, N, R, S, T, and Au-Pt thermocouple types Temper ...

  • Fluke 1529 Chub-E4 - page 19

    PRT Thermistor Thermocouple Probe Connection Patented DWF connectors accepts spade lug, bare-wire, or banana plug terminati ons Universal receptacle accepts miniature and standar d connectors Communications RS-232 and IR por ts included, IEEE-488 (GPI B) optional AC Power 100 to 240 VAC, 50-60 Hz, 0.4A DC Power 12 to 16 VDC, 0.5 A ( battery charges ...

  • Fluke 1529 Chub-E4 - page 20

    3.3 W arranty Fluke Corporati on, Hart Scientif ic Di vision (Hart) warrants t his product to be free from defects in materia l and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warrant y extends onl y to the original purchaser and shall not apply t o any product, whic ...

  • Fluke 1529 Chub-E4 - page 21

    4 Quic k Start This section briefly explains the basics of setting up and operating your 1529 Thermometer Readout. 4.1 Unpac king Carefully unpack the 1529 t hermometer readout. V erify that the foll owing items are present: • 1529 Thermometer Readout (The battery inside the 1529 is completely charged when shipped. Initial charging is not require ...

  • Fluke 1529 Chub-E4 - page 22

    4.4.1 Using the Clamp-on Ferrites Clamp-on ferrites a re provided wit h this product for the use of impro ving its electromagnetic (EM) immunity in en vironments of excessiv e EM interference. During EMC testi ng we found that ferrites cl amped around the probe cables and power cord reduced the risk that EM interference af f ects measurements. Ther ...

  • Fluke 1529 Chub-E4 - page 23

    Enter button to continue initialization. If an error message is display ed on power up see Section 11.1, T roubleshooting. 4.7 Measure T emperature After initialization, th e temperature measurements for the configured chan nels are displayed. The appropriate sensor type must be select ed and coeff icients en - tered for the measurements to be accu ...

  • Fluke 1529 Chub-E4 - page 24

    5 P ar ts and Contr ols The functions of the v arious features of the 1529 are described below . 5.1 Fr ont P anel Buttons The front panel butt ons Enter/Menu, Unit , Contrast ,a n d Exit a r eu s e dt os e - lect and alter the functions of the t hermometer readout (see Figure 2). The function of each button is as follows: Enter/Menu - This b utton ...

  • Fluke 1529 Chub-E4 - page 25

    1529 Chub-E4 Thermometer Readout User’s Guide 18 1529 1529-R 1529-T RS-232 IEEE-488 (option) POWER 1 AMP 14.5–16 V CHARGING IR PORT FUSES INTERNAL - 1 A T 125 V FLUKE CORPORA TION HART SCIENTIFIC DIVISION www.hartscientifc.com CHANNEL CONFIGURA TION CH1 CH2 CH3 CH4 – + Figure 3 Back P anel ...

  • Fluke 1529 Chub-E4 - page 26

    5.2 Bac k P anel The back panel consists of the po wer switch, A C adapter connector , charging indicator , IR port, RS-232 port connect or , IEEE-488 port connector (optional), serial label, and probe connectors. The 1529 thermometer readout can be con - figured in three dif ferent ways. The configurati on affects t he probe connectors on the back ...

  • Fluke 1529 Chub-E4 - page 27

    • 2380 Small Thermocouple Connector Kit • 2381 Large Thermocouple Connector Kit • 2374 I/R Dongle • 2385 HP-IR Printer • 2386 HP-IR Printer Adapter • 2387 HP-IR Printer Paper 1529 Chub-E4 Thermometer Readout User’s Guide 20 ...

  • Fluke 1529 Chub-E4 - page 28

    6 General Operation This section e xplains basic operat ion of the 1529 thermometer readout. De - tailed operati on of the 1529 is e xplained in Secti ons 7 and 8. Section 7 e xplains the menu structure and the functions a v ailable in the menu struct ure and Sec - tion 8 e xplains the communicat ions interf ace for operating the 1529 remote ly . 6 ...

  • Fluke 1529 Chub-E4 - page 29

    is fully charged. Only ch ar ge the batte ry when ambient temperature is between 16°C and 30°C (61°F and 86°F) It normally tak es about three hours to fully charge the battery . The instrument can b e operated while the battery is being charged. The char ging indicator on the back panel is green when the battery is fully charged a nd amber when ...

  • Fluke 1529 Chub-E4 - page 30

    All possible c onfigurat ions are sho wn in Figure 3 on page 18. 6.4.1 Connecting a PR T or Thermistor Probe PR T and thermistor probes are attached to the resist ance input module via Hart Scientific mini-D WF connectors. Thes e patented connectors accept bare wire, spade, or mini banana plug terminations. The connectors are col or coded for ease ...

  • Fluke 1529 Chub-E4 - page 31

    1529 Chub-E4 Thermometer Readout User’s Guide 24 Channel 1 Channel 2 Connecting 4-wire probes Channel 1 Channel 2 Connecting 3-wire probes Channel 1 Channel 2 Connecting 2-wire probes Shield Figure 5 Probe Connection Wir ing Diagram ...

  • Fluke 1529 Chub-E4 - page 32

    6.5 DC P ower Sour ce Caution : F or CE compliance and for performance , use only the AC adapter shipped with the instrument by Hart Scientific. If the AC adapter needs to be replaced, contact a Hart Authorized Service Center (see Sec - tion 1.3). The DC power source pro vides power to char ge the battery . It can also be used to po wer the 1529 wh ...

  • Fluke 1529 Chub-E4 - page 33

    6.8.2 Enabling the Channel Channels are set first by selecting either simultaneous or scan mode in the CHANNEL menu CHANNEL MODE function. Simult aneous mode sho ws one to four channels in the upper di splay simultaneously . Scan mode sho ws one channel at a time (in lar ge digit format) scanning through one to four channels depending on the channe ...

  • Fluke 1529 Chub-E4 - page 34

    6.10 Data Logging Data can be logged either on demand or automati cally at a user -selected inter - val . When data is logged, readi ngs are stored to the inst rument's memory and assigned a label for easy recall and or ga nization of data. Up to 100-demand log and 8,160 auto log readings can be stored. When data is being logged, the left bott ...

  • Fluke 1529 Chub-E4 - page 35

    7 Menu Functions Selecting the Enter/Menu button from the front panel accesses the main menu of the 1529 thermometer readout. The main menu consists of the submenus: CHANNEL , PR OBE , FIELDS , LOGGING ,a n d SYSTEM . Each submenu has its o wn set of functions. The Enter/Menu b u t t o ni su s e dt os e l e c ta n ds a v e menu choices. The Unit LR ...

  • Fluke 1529 Chub-E4 - page 36

    and DISPLA Y OPTIONS/A UT O-CAL. (The DISPLA Y OPTIONS changes to A UTO-CAL when the measure peri od is less than 1 second.) 7.1.1 Measure P eriod and Fast Measurement Mode The MEASURE PERIOD functi on allo ws you to control t he period (time) be- tween measurements. Y ou may select between 0.1, 0.2, 0.5, 1, 2, 5, 10 and 30 seconds, 1, 2, 5, 10, an ...

  • Fluke 1529 Chub-E4 - page 37

    The LR buttons are used to select the period between measurements. Press the Enter button to sa ve the new set ting. Press the Exit button to cancel and to exit to the menu. When the measure period is set t o less than 1 second (i.e. 0.1, 0.2, 0.5 seconds) the 1529 enters the Fast Measurement M ode. The menu buttons may seem to be sluggish in this ...

  • Fluke 1529 Chub-E4 - page 38

    Press ENTER to continue … Press the Enter butt on to confi rm and to continue. The display is remains un - changed and all enabled channels remai n enabled channels. Press the Exit but - ton to cancel. When the measure period is reset to 1 second or greater , the following message is displayed indi cating that the 1529 is returning to normal oper ...

  • Fluke 1529 Chub-E4 - page 39

    hold the Exit button to cancel to the main display or press the EXIT button to cancel and to mov e to the ne xt parameter . 7.1.3 Channel Mode The CHANNEL MODE functi on sets the channel displ ay mode. When this function is selected, the bott om portion of the display sho ws the current scan mode and allows the setting to be change d. The follo win ...

  • Fluke 1529 Chub-E4 - page 40

    proving resolut ion. When this function i s selected, the bott om portion of the display shows the av eraging setting and allows the setting to be changed. Select the number of raw measurements to be a veraged to produce the displ ayed mea - surement. The range is from 1 to 10. The factory default is 1. The measurements are av eraged together until ...

  • Fluke 1529 Chub-E4 - page 41

    of the display sho ws the current display settings. Thi s function is the same as the FIELDS menu DISPLA Y OPTIONS function. The RESOL parameter determines the number of decimal places (0 through 4 or A UTO) to be displayed for the temperature measurements. If A UTO is se- lected, the instrument automatically determines the number of d ecimal place ...

  • Fluke 1529 Chub-E4 - page 42

    testing con version settings, viewing and printing probe parameters, and clearing the default probe. The functions that appear in this menu are EDIT PR OBE, COPY PR OBE, TEST PROBE, PRINT PR OBE, and DEF A UL T PR OBE. 7.2.1 Edit Probe The EDIT PR OBE function allo ws you to select a probe channel, enter the probe serial number , and select the con ...

  • Fluke 1529 Chub-E4 - page 43

    channel, the probe serial number and the con version type. The characterization coeff icients and p arameters are displaye d after the con version type is selected. The PR OBE parameter is for selecting an y one of the four input channel s: CHAN1, CHAN2, CHAN3 and CHAN4 or one of t he ten probe memory chan- nels: PMEM0, PMEM1, PMEM2 , PMEM3, PMEM4, ...

  • Fluke 1529 Chub-E4 - page 44

    the next parameter . On the last parameter , press the Exit b u t t o nt oe x i tt ot h e menu. 7.2.1.1 ITS-90 Con version The ITS-90 con version is for PR Ts calibrated and characterized using the Inter - national T emper ature Scale of 1990 equations. The parameters that appear when ITS-90 is selected are R TPW , A4, B4, A, B, C, and D. These sho ...

  • Fluke 1529 Chub-E4 - page 45

    The RANGE parameter is used to select the appropriate resist ance range. For 25 ohm and 100 ohm PR Ts select 100 ohms and for 1000 ohm PR Ts select 10K ohms. The WIRES parameter sets the appropriate number of lead wires from the sen - sor . This parameter can be set to 2-, 3-, or 4-wires. See Figure 5 on page 24 for the wiring con f iguration. The ...

  • Fluke 1529 Chub-E4 - page 46

    and to sav e the ne w settings. Use the LR b uttons to change each digit of the settings. Press and hold t he Exit button to cancel to the main display or press the EXIT button to cancel and to mo ve to the next parameter . On the last pa - rameter , press the Exit button to e xit the menu. 7.2.1.3 PT -100 Con version The parameters for the PT -100 ...

  • Fluke 1529 Chub-E4 - page 47

    The A, B, and C coef ficients can be con verted t o α , δ ,a n d β coefficients using the following formulas: α= + AB 100 δ= − + 100 100 1 A B β= − + 10 100 8 C AB When the CVD con version type is selected, the R0, ALPHA, DEL T A, and BET A coeff icients are displayed on the right portion of the bottom of the dis - play . Use the UD butto ...

  • Fluke 1529 Chub-E4 - page 48

    When the Thermistor T(R) con version type is selected, the coeff icients are d is - played on the right portion of the bottom of the di splay . Use the UD buttons to scroll through the coef fi cients. Use the Enter b utton to e dit and sa ve the ne w setting. Press and hol d the Exit button to cancel to the main display or press the EXIT button to ...

  • Fluke 1529 Chub-E4 - page 49

    7.2.1.10 Standard Thermocouple Con versions Standard thermocouple con versions incl ude types B, E, J, K, N, R, S, T , and Au-Pt. V oltage is con verted to temperature using standard reference functions. (For addit ional information c onsult the publi cation NIST Monograph 175.) Y ou can specify internal or external reference junction compensation ...

  • Fluke 1529 Chub-E4 - page 50

    7.2.2 Copy Pr obe The COPY PR OBE function allo ws you to copy a set of probe parameters from one channel or memory location to a nother . The parameters are copied from the user selected source channel or memory l ocation into the current channel. The COPY PR OBE function can be password protected to restrict access. See Sec - tion 7.5.3, Passw or ...

  • Fluke 1529 Chub-E4 - page 51

    the display sho ws the probe channel or memory location, the probe serial num - ber , and the con version type. Use the LR butt ons to select the channel or memory location. After the channel is selected, press the Enter button to test the temperature calculation for the probe. Y ou can enter an input v alue ( Ω ,K Ω , o r milli volts) and the co ...

  • Fluke 1529 Chub-E4 - page 52

    this function is select ed, the bottom portion of the display sho ws the probe channel. Use the LR butt ons to select the channel, memory locati on, or ALL. Use the UD buttons to scroll between parameters. Press the Enter button to co ntinue and to send the data to the selected por t. A message is displayed stating the channel to print and the port ...

  • Fluke 1529 Chub-E4 - page 53

    selected channel, memory location, or ALL to the fact ory defaults. When this function is selected, the bott om portion of the display sho ws the probe channel. Use the LR buttons to select the channel, memory location, o r ALL. Press the Enter button to reset the probe coeff icients. Press the Exit button to cancel and to e xit to the menu. 7.3 Fi ...

  • Fluke 1529 Chub-E4 - page 54

    functions that appear in this menu are CLEAR ST A TS, SELECT FIELDS, EDIT FIELDS, DEF A UL T FIELDS, and DISPLA Y OPTIONS. 7.3.1 Clear Stats The CLEAR ST A TS function clears the statistical registers and resets all statis- tical calculations for all channels. Refer to Section 7.3.2, Select Fields, to dis- play statistical data. When this function ...

  • Fluke 1529 Chub-E4 - page 55

    Press the Enter button to clear the statistical data. Press the Exit b u t t o nt oa b o r t clearing the statistical data and to e xit to the menu. 7.3.2 Select Fields The SELECT FIELDS function allo ws you to select a display set (1 through 9) to use for displaying i nformation provi ded by the instrument. A displ ay set can be customized to d is ...

  • Fluke 1529 Chub-E4 - page 56

    Keyw ord Field T ype A VE Av erage BA TT Battery Percent Remaining DA T E D at e INP Input in ohms, K ohms, or mV MAX Maximum MEAS Number of Seconds Until Next Measur ement MIN Minimum REF Ref erence (see below ) SER Probe Serial Number SPR Spread STD Standard De viation STN Count -T1, -T2, -T3, -T4 Delta TEM T emperatur e TIME Time TYP Conv ersion ...

  • Fluke 1529 Chub-E4 - page 57

    7.3.3 Edit Fields The EDIT FIELDS f unction allows editing of the display fields for a g i ven dis - play set (1 through 9). Y ou can choose to display one to eight fiel ds. The av ail - able f ields and their corresponding k eyw ords are gi ven in the Field T ypes table abov e. When this functi on is selected, the bot tom portion of the displ ay a ...

  • Fluke 1529 Chub-E4 - page 58

    factory defaul ts. When this function i s selected, the bott om portion of the dis - play instructs you to press Enter to re vert to the default display set. Press the Enter butt on to rev ert to the default displ ay setup. Press the Exit but- ton to e xit to the menu. 7.3.5 Displa y Options The DISPLA Y OPTIONS function allows you to select the di ...

  • Fluke 1529 Chub-E4 - page 59

    bottom portion of t he display sho ws the current display setti ngs. This function is the same as the CHANNEL MODE menu DISPLA Y OPTIONS function. The RESOL parameter determines the number of decimal places (0 through 4 or A UTO) to be displayed for the temperature measurements. If A UTO is se- lected, the instrument automatically determines the nu ...

  • Fluke 1529 Chub-E4 - page 60

    this menu are DEMAND LOG, A UTO LOG, LOG ST A TS, D A T A LABELS, and DEF A UL T LABELS. 7.4.1 Demand Log The DEMAND LOG submenu allo ws measurements to be logged on demand, one reading at a time. Readings are store d to the instrument's memory and as- signed a label for easy recall and or ganization of data. The instruments memory can store u ...

  • Fluke 1529 Chub-E4 - page 61

    mand logging submenu which consists of the functions: ST ORE READING, LOG HISTOR Y , VIEW D A T A, PRINT D A T A, and DELETE D A T A. 7.4.1.1 Store Reading The STORE RE ADING function stores the current measurements for all dis- played channels. When the instrument is in simult aneous mode, all channels that are displayed on t he upper portion of t ...

  • Fluke 1529 Chub-E4 - page 62

    current resolution setting) , and the units as the y are stored into memory . The right side of the bottom displ ay shows the current l abel. T o change the label, press the LR buttons to scroll t hrough and select the la- bel that applies to the demand log data being sto red. F or information on e diting the labels see the D A T A LABELS function ...

  • Fluke 1529 Chub-E4 - page 63

    When this function is selected, the b ottom portion of the display shows the de - mand log history data. Press the UD butt ons to scroll up and do wn through the st ored readings se- quentially on e at a time. Press the LR b uttons to scroll up and do wn through the stored readings sequential ly by groups of 16. Press the Enter or Exit button to re ...

  • Fluke 1529 Chub-E4 - page 64

    HISTOR Y function. When this funct ion is selected, the bottom portion of the display sho ws the demand log dat a. Press the UD butt ons to scroll up and do wn through the st ored readings se- quentially . Press the LR buttons to scroll up and do wn through the stored readings sequentiall y by groups of 16. Press the Enter or Exit b u t t o nt or e ...

  • Fluke 1529 Chub-E4 - page 65

    and the port. Select ALL D A T A to print all demand log entries, or select a spe - cific label to print onl y the demand log data for that label . Use the LR buttons to select the parameter settings. U se the UD buttons to scroll between parameters. Press the Enter button to continue. A message i s displayed stati ng the label of the data to print ...

  • Fluke 1529 Chub-E4 - page 66

    bered. Select ALL D A T A to delete all de mand log readings or select a specific label to delete onl y the demand log data for that la bel. Use the LR buttons to select the label of th e data you want to delete. Press the Enter button to continue and delete t he data. Press the Exit button to can- cel and return to the DEMAND LOG menu. 7.4.2 A uto ...

  • Fluke 1529 Chub-E4 - page 67

    A UTO LOG submenu and consists of the functions: LOGGING OPTIONS, ST AR T STOP , VIEW D A T A, PRINT D A T A, and DELETE D A T A. 7.4.2.1 Logging Options The LOGGING OPTIONS function allo ws the user to defi ne the logging pa- rameters. When this function is select ed, the bottom portion of t he display prompts the user to select the label, period, ...

  • Fluke 1529 Chub-E4 - page 68

    The LOG PER parameter is for selecting t he interv al between measurements. Y ou may select between 0.1, 0.2, 0.5, 1, 2, 5, 10, and 30 seconds, 1, 2, 5, 10, and 30 minutes or 1 hour . The LOG PER must be set to a v alue less t han or equal to the measurement period. The 1529 will not log readings faster than the measurement period. The measurem ent ...

  • Fluke 1529 Chub-E4 - page 69

    Note: The log is not st arted or stopped until the Enter b u t t o ni sp r e s s e dt oc o n - firm the user action. After a log sessio n is started, the bottom por tion of the display shows the log statistics. These statistics can also be viewed from the LOG ST A TS function of the LOGGING menu. The left side displays deman d log statistics and th ...

  • Fluke 1529 Chub-E4 - page 70

    readings currently stored in t he demand or auto log memory . The COUNT pa - rameter sho ws the number of auto l og readings remaining during the current log session. Press the Enter or Exit b u t t o nt oe x i tt h i sd i s p l a ya n dt or e t u r nt ot h eA U T O LOG menu. Only those channels t hat are enabled during the l og session are logged. ...

  • Fluke 1529 Chub-E4 - page 71

    Subsequent entries in t he log consist of the readi ng number , channel, measure - ment value, units, and the time of each entry . Press the LR buttons to jump 100 readings at a time. Hold the UD buttons down t o advance quickly through the data. Press and hold the L b utton to vie w the top reading, and t he R button to vie w the bottom reading . ...

  • Fluke 1529 Chub-E4 - page 72

    Select ALL D A T A to delete all auto log r eadings or select a specif ic label to de - lete only the auto log entries for t hat label. Use the LR buttons to select the label of th e data you want to delete. Press the Enter button to conti nue and to delete the dat a. Press the Exit button to cancel and to return to the A UTO LOG menu. 7.4.3 Log St ...

  • Fluke 1529 Chub-E4 - page 73

    both the demand and auto logs. When this function is selected, the bott om por - tion of the displ ay sho ws the logging stat istics. The left side displays deman d log statistics and the right side d isplays the auto log statistics. The FILLED parameter shows the percentage of the total demand or auto log memory that is currently filled. The FREE ...

  • Fluke 1529 Chub-E4 - page 74

    digits. These labels are used to ident ify logged data when using the demand and auto log functions. Press the LR buttons to select the label to ed it. By default the labels are set to D A T A_01, DA T A_02, … at the factory . Press the Enter button to begin edit- ing the label. The word SELECT changes to EDIT , and the label can no w be edited. ...

  • Fluke 1529 Chub-E4 - page 75

    this function is s elected, the bottom portio n of the display instructs the u ser to press the Enter button to reset the data labels to their default settings. T o exit without resetting the labels, press the Exit button. 7.5 System Menu The SYSTEM menu provides functions for the communi cation ports, the date and time format, the password , the c ...

  • Fluke 1529 Chub-E4 - page 76

    appear in this menu are COMM SETUP , D A TE TIME, P ASSWORD, CALI - BRA TION, and SYSTEM RESET . 7.5.1 Comm Setup The COMM SETUP function is used to setup the communication parameters for the installe d ports. Communications port s are used for communicating with a computer , printer , or other equipment. When this function is selected, the bot- to ...

  • Fluke 1529 Chub-E4 - page 77

    sages stored in the error queue on t he display . Once an error has been vie wed, it is remov ed from the error queue. Press the LR buttons to select t he port to setup. Press the Enter button to display the port parameters. 7.5.1.1 Serial When the SERIAL option is selected, the parameters for the serial port are dis- played on the bottom port ion ...

  • Fluke 1529 Chub-E4 - page 78

    The ECHO parameter allows t he user to set the echo (duplex) mode. Selecting ON enables echoing so that all charact ers receiv ed on the serial port are echoed back ov er the serial port. This mode is useful when using terminal emulation software on a computer to co mmunicate with the instrument. Selecting O FF disables echoing. This mode is common ...

  • Fluke 1529 Chub-E4 - page 79

    The GPIB parameter allows the user to enable or disable the IEEE-488 port. The ADDR parameter allo ws the user to specify the address of this instrument on the IEEE-488 bus. The def ault address is 22 and the range is 1 to 30. The TERM parameter allows the user to s elect the termination character trans - mitted after each transmission. The defau l ...

  • Fluke 1529 Chub-E4 - page 80

    The D FORM parameter allows the user to set the date f ormat used when dis - play the date. The choices are M -D-YY , M-D-YYYY , D/M/YY , and D/M/YYYY . Press the LR buttons to change the parameter setting. Use the UD b uttons to scroll between the parameters. Press the Enter b u t t o nt os a v et h en e ws e t - ting. Press the Exit butt on to ex ...

  • Fluke 1529 Chub-E4 - page 81

    of the display . The password-protected parameters are P ASSWRD and PRB SEC. The P ASSWRD parameter allo ws the user to change the password. Press the Enter button t o edit the P ASSWRD parameter . Press the LR buttons to highlight th e digit to change. U se the UD b uttons to scroll through the num - bers 0-9. Press the Enter but ton to sav e the ...

  • Fluke 1529 Chub-E4 - page 82

    By default the password i s set to ‘1529’ when the instrument ships from the factory . W e recommend that the user cha nge the default password to protect the password-protected sett ings. Note: Keep your passw ord in a secure location and do not forget it. T o enter the password, press the Enter button. Press the LR b uttons to high - light th ...

  • Fluke 1529 Chub-E4 - page 83

    The C0K, C10K, and C100K parameters apply to the thermistor probe types and 10K Ω resistance range. Use these parameters to adjust the mea - surement at 0K,10K, and 100K Ω respecti vely . These values are in K ohms . The D A TE parameter is the date the current calibration was performed. The DUE parameter is the date the current calibration expir ...

  • Fluke 1529 Chub-E4 - page 84

    By default the password i s set to ‘1529’ when the instrument ships from the factory . W e recommend that the user cha nge the default password to protect the password-protected sett ings. Note: Keep your passw ord in a secure location and do not forget it. T o enter the password, press the Enter button. Press the LR b uttons to high - light th ...

  • Fluke 1529 Chub-E4 - page 85

    8 Digital Comm unications Interface 8.1 Overview The communication feature allows an e xternal de vice, such as a computer , to communicate with the 1529 t o obtain measurement data and cont rol operating conditions. Communication is accomp lished by issuing commands to the 1529 through RS-232, IR, or IEEE-488 communication ports. 8.2 Comm unicatio ...

  • Fluke 1529 Chub-E4 - page 86

    All commands sent to the 1529 through the serial interface must be te rminated with a carriage return or linefeed character . 8.2.2 GPIB Communications The GPIB interface allo ws the Model 1529 to be connected to an IEEE-488 bus along with other inst ruments. A system controller can remotely control the op - eration of the Model 1529 and read measu ...

  • Fluke 1529 Chub-E4 - page 87

    Command Description Refer To *CLS Clears the status r egisters Section 8.4.12 *ESE? Returns the Event Stat us Enable Register Section 8.4.12 *ESE <num>|MIN|MAX| DEF Sets the Event Status Enabl e Register Section 8.4.12 *ESR? Returns the Event Stat us Register Section 8.4.12 *IDN? Returns the instr ument identifi cation string that indicates t ...

  • Fluke 1529 Chub-E4 - page 88

    Command Description Refer To CALibrate:AUTO Executes the automatic cal ibration procedur e when using fast measurement mode. Section 8.4 .5 CALibrate<chn>:DATE:CALi brate? [MIN|MAX|DEF] Returns the channel calibr ation date for t he specified channel Section 8.4 .5 CALibrate<chn>:DATE:CALibrate (<year>,<month>,<day>) | ...

  • Fluke 1529 Chub-E4 - page 89

    Command Description Refer To FORMat:STAMp? Returns the extended measurement data format setting Section 8.4.1 FORMat:STAMp <bool> Enables extended measurement data f ormat with the FETCh?, MEASure?, and READ? commands Section 8.4 .1 INITiate Provided for co mpatibility but has no effect Section 8.4.2 INITiate:CONTinuous? Provided for compat i ...

  • Fluke 1529 Chub-E4 - page 90

    Command Description Refer To ROUTe:SCAN [<chn list>] Sets the input channels for scanning Section 8.4. 3 ROUTe:SCAN:MODE? [MIN|M AX|DEF] Returns the current scan mode Section 8.4.3 ROUTe:SCAN:MODE <num>|MIN|MAX| DEF Sets the scan mode Section 8.4.3 SENSe:AVERage:COUNt? [MIN| MAX|DEF] Returns the moving average filter setting Section 8.4 ...

  • Fluke 1529 Chub-E4 - page 91

    Command Description Refer To SYSTem:COMMunicate:SERial:FEED <bool>|MIN|MAX|DEF Sets the state o f the serial port printing settin g Section 8 .4.9 SYSTem:COMMunicate:SERial:LINefeed? [MIN|MAX|DEF] Returns the state of the serial port l inefeed setting Section 8.4.9 SYSTem:COMMunicate:SERial:LINefeed <bool>|MIN|MAX|DEF Sets the state o f ...

  • Fluke 1529 Chub-E4 - page 92

    Commands consist of a command header and, if necessary , parameter data. All commands must be terminated with either a carriage return (ASCII 0D he x or 13 decimal) or ne w line character (ASCII 0A he x or 1 0 decimal ). Command headers consist of one or more mnemonics separated by colons (:). Mnemonics may use letter characters, the underscore cha ...

  • Fluke 1529 Chub-E4 - page 93

    Logging Commands – commands for automati c and demand logging. System Commands – commands for general system configuration parameters. Communication Interface Commands – commands for contro lling the inter - face conf iguration(serial port, IR, and IEEE). Date and Tim e Commands – commands fo r setting the date, time and the date and time f ...

  • Fluke 1529 Chub-E4 - page 94

    • <num> indicates numeric value is required. • <chn list> indicates a list of channel numbers is required. • <bool> indicates a Boolean value (0 or 1) is required. The mneumonics OFF and ON are also accepted for 0 and 1, respectively . • <con v> indicates a conv ersion mnemoic is required. • <param> indicates ...

  • Fluke 1529 Chub-E4 - page 95

    The CALCulate suf fix, <chn>, specifies the channel (1 to 4). The A VERage suff ix, <n>, specifies the calculation type. See T able 4 on page 126 for the Sta - tistical Calculation T y pes. 8.4.1.3 CALCulate:A VERage<n>:TYPE? Returns the ke yword for the specified f ield type number . CALC:A VER2 :TYPE STD The A VERage suffi x, &l ...

  • Fluke 1529 Chub-E4 - page 96

    day the measurement was made W ith th is format k ohms and milliv olts are not con verted to ohms and volts. This command may return the same reading if a ne w reading has not yet been obtained. If only ne w readings are desired, the FORMat:ST AMp command can be used to enable the e xtended format option and any readings i n which the val ue of the ...

  • Fluke 1529 Chub-E4 - page 97

    The SENSe suff ix, <chn>, specifies the channel (1 to 4). The parameter is ohms for low range PR Ts, K ohms for high range PR Ts and thermistors, and milli volts for thermocouples. The second returned parameter is alw ays 0 for PR Ts and themistors and the reference junction temperature for thermocouples. 8.4.2 Measurement Control Commands Th ...

  • Fluke 1529 Chub-E4 - page 98

    0 This command only applie s to PR T/Thermistor channels. Ex ecuting this com - mand on Thermocouple channels generates a -294, "Incompatible type", error message. The SENSE suf fix, <chn>, specif ies the channel (1 to 4), or memory channel (5 to 14). Specifying the MIN parameter re turns a value of 0. Specifying the MAX parameter r ...

  • Fluke 1529 Chub-E4 - page 99

    8.4.2.8 SENSe<chn>:RESistance:WIRE <num>|MIN|MAX|DEF Sets the probe wire setting . SENS2:RES:WIRE MIN This command only applie s to PR T/Thermistor channels. Ex ecuting this com - mand on Thermocouple channels generates a -294, "Incompatible type", error message. The SENSE suf fix, <chn>, specif ies the channel (1 to 4), ...

  • Fluke 1529 Chub-E4 - page 100

    8.4.3 Channel Commands The channel commands are used for query ing and setting the current channel, the channels to scan, the type of scanning, and the state of the channel scan - ning. The channel commands are not password protected. 8.4.3.1 ROUT e:CLOSe? <chn> Returns the state of the specified channel. R OUT :CLOS? 1 0 The <chn> para ...

  • Fluke 1529 Chub-E4 - page 101

    8.4.3.6 ROUT e:SCAN? Returns a comma delimited list of the enabled chan nels. R OUT :SCAN? (@2,3,4) The list of channels is enclosed in parenthesis and preceded with the @ symbol. 8.4.3.7 ROUT e:SCAN <chn list> Sets the channels for scanning. R OUT :SCAN 2,3 The <chn list> parameter is a comma d elimited list of the channels (1 to 4) to ...

  • Fluke 1529 Chub-E4 - page 102

    CALC2:CONV :CA T? “RES”,“ITS”, ”ITS5”, ”PT”, ”CVD”, ”TRES”, ”TTEM”, ”YSI” The CALCulate suf fi x, <chn>, specifies the channel (1 to 4) or memory chan - nel (5 to 14). Each item in the response lis t is enclosed in double quotes and separated from other items by a comma. See T able 7 on page 128, for the av ail ...

  • Fluke 1529 Chub-E4 - page 103

    type. The acceptable con version types depend on the channel type. The con ver - sion types and their names are l isted in T able 7 on page 128. This command may be password-protected. T o access this command: • If the SYST :P ASS:CONV? command returns a 0 (probe password-protec - tion is disabled), the password is not required. • If the SYST : ...

  • Fluke 1529 Chub-E4 - page 104

    CALC2:CONV :P AR:V AL R TPW , 100.0145, A4, 0.0045 The CALCulate suf fi x, <chn>, specifies the channel (1 to 4) or memory chan - nel (5 to 14). The <param> parameter is the name of the parameter as g i ven in T able 8, Con ve rsion T ypes and Parameters. If the <param> parameter is not ap - plicable to the current con version t y ...

  • Fluke 1529 Chub-E4 - page 105

    can consist of up to eight charact ers that include an y letters, numeric digits, and the underscore ‘_’. This command may be password-protected. T o access this command: • If the SYST :P ASS:CONV? command returns a 0 (probe password-protec - tion is disabled), the password is not required. • If the SYST :P ASS:CONV? command returns a 1 (pr ...

  • Fluke 1529 Chub-E4 - page 106

    2000,1,1. Specifying the MAX parameter returns a v alue of 2099,12,31. Spec - ifying the DEF parameter returns a v alue of 2000,1,1. 8.4.5.3 CALibrate<chn>:D A TE:CALibrate (<year>,<month>,<day>)|MIN|MAX|DEF Sets the calibration date f or the specified channel. CAL1:D A TE:CAL 2000,8,29 The CALibrate suf fix, <chn>, sp ...

  • Fluke 1529 Chub-E4 - page 107

    8.4.5.6 CALibrate<chn>:P ARameter :OFFSet<n>? [MIN|MAX|DEF] Returns the calibration offset par ameter for the specifi ed channel. CAL1:P AR:OFFS2? –1.2 The CALibrate suf fix, <chn>, specif ies the channel (1 to 4). This command does not apply to memory channels. For PR T/Thermistor channels, the cal ibra - tion offset suf fix &l ...

  • Fluke 1529 Chub-E4 - page 108

    Thermistors (2). For Thermocouple c hannels, the calibration scale suf fix <n> is ignored and may be omitted. Note: For PR T/Thermistor channels, the value re - turned by this command is always in ohms. For PR T/Thermisto r channels, specifying the MIN parameter retu rns a v alue of -9 when the scale suf fix <n> is 1 (for PR Ts) and -90 ...

  • Fluke 1529 Chub-E4 - page 109

    suff ix <n> is 1 (for PR Ts) and 9000 when the linearity suf fix <n> is 2 (for Thermistors). Specifying the DE F parameter returns a value of 0. 8.4.5.11 CALibrate <c hn>: P ARameter:LINearity<n> <num>|MIN|MAX|DEF Sets the calibration linearity parameter for th e specif ied channel. CAL1:P AR:LIN1 5 This command only a ...

  • Fluke 1529 Chub-E4 - page 110

    the specified channel. Specifying t he MIN, MAX, or DEF parameter sets the RJC parameter to the minimum, maxim um, or default v alue respecti vely . This command is password-protected. T o access this command, the password must be successfully entered using the SYST :P ASS:CEN <pass> command (see Section 9.4.11, P assword Commands). 8.4.6 Dis ...

  • Fluke 1529 Chub-E4 - page 111

    MIN, MAX, or DEF sets th e lamp to the minimum, maximum, or default value respecti vely . 8.4.6.5 DISPlay:DECimal:FORMat? [MIN|MAX|DEF] Returns the current decimal format setting. DISP:DEC:FORM? 0 A returned value of 0 i ndicates a period (.) is used as the decimal separator . A returned value of 1 indicates a comma (,) is used as the decimal separ ...

  • Fluke 1529 Chub-E4 - page 112

    The WINDow s uff ix, <n>, specif ies the display set (1 to 9). The FIELd suf fix, <n>, specifi es the fi eld (1 to 8) within the displ ay set. The response consists of two numbers separated by a comma. The fi rst number represents the channel associated with the f ield. The second number represents the parameter to be displayed in the s ...

  • Fluke 1529 Chub-E4 - page 113

    The <num> parameter is the number of measurements to be stored in the auto log during the auto log session. Specifying the MIN, MAX, or DEF parameter sets the number of measurements to aut o log to the minimum, maxi mum, or de - fault respecti vely . 8.4.7.4 LOGging:A UT omatic:FREE? Returns the number of a uto log entries free a nd stored. L ...

  • Fluke 1529 Chub-E4 - page 114

    8.4.7.8 LOGging:A UT omatic:PRINt [<num>|ALL [,<port>]] Prints the auto log entries. LOG:A UT :PRIN ALL D A T A_25 1 22.676C 12:19:42 09-05-00 D A T A_25 2 9.960 KO 12: 19:44 09-05-00 D A T A_25 3 23.220C 12:19:46 09-05-00 D A T A_25 4 23.245C 12:19:48 09-05-00 D A T A_25 1 22.765C 12:19:52 09-05-00 D A T A_25 2 9.962 KO 12: 19:54 09-05 ...

  • Fluke 1529 Chub-E4 - page 115

    8.4.7.12 LOGging:AUT omatic:TIMe <num>|MIN|MAX|DEF Sets the auto log interval setting. LOG:A UT :TIM 1 The <num> parameter can be one of the follo wing values: 0.1, 0.2, 0.5, 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800 or 3600. The units are seconds. If a v alue other than one listed is used, the instrument aut omatically selects the neare ...

  • Fluke 1529 Chub-E4 - page 116

    Note: Every time a demand log reading is stored, the first entry in the log for that stored reading is the data label and the date. All subsequent entries in the demand log consist of the channel, measurement v alue, units, and time. 8.4.7.16 LOGging:DEMand:LABel? [MIN|MAX|DEF] Returns the current demand log label setting. LOG:DEM:LAB? 6 Specifying ...

  • Fluke 1529 Chub-E4 - page 117

    8.4.7.20 LOGging:DEMand:ST ORe Stores the current measurement into the demand log. LOG:DEM:ST OR If simultaneous scan mode is enabled, one reading is stored i n the log for each enabled channel. 8.4.7.21 LOGging:DEMand:V ALue? <num>|MIN|MAX|DEF Returns the specif ied demand log entry . LOG:DEM:V AL? 10 D A T A_25,2,9.961,KO,14,59,35,2000,9,5 ...

  • Fluke 1529 Chub-E4 - page 118

    *IDN? HAR T ,1529,A09001,1.11 8.4.8.2 *OPT? Returns a list of input modules instal led. *OPT? PR T ,TC Input module types are reported by pos ition, top to bottom, of the tw o input modules. The returned exampl e is for a 1529 equipped with one 2-channel PR T input module and one 2-channel Thermocouple input module. 8.4.8.3 *RST Sets the instrument ...

  • Fluke 1529 Chub-E4 - page 119

    8.4.8.6 SYST em:ERRor? Returns a system error message if any are present in the system error queue. SYST :ERR? 0,“No error” Each error condition produces only one error message at a time. The error queue may contain up to ten messages. The messages are reported in the order they occured. Reading a message from the queue remo ve s the message so ...

  • Fluke 1529 Chub-E4 - page 120

    8.4.9 Comm unication Interface Commands The communication interface commands are used for queryin g and setting se - rial port, IR, and IEEE configuration parameters. 8.4.9.1 SYST em:COMMunicate:IR:FEED? [MIN|MAX|DEF] Returns the state of t he IR port printing. Example: SYST :COMM:SER:FEED? Response: 0 A val ue of 1 is returned if IR port printing ...

  • Fluke 1529 Chub-E4 - page 121

    sets the IR mode to IRD A mode. Specifying the MIN, MAX, or DEF parameter sets the IR mode to th e minimum, maximum, or d ef ault value respectiv ely . 8.4.9.5 SYST em:COMMunicate:IR:TIMe? [MIN|MAX|DEF] Returns the IR sample period sett ing. Example: SYST :COMM:IR:TIM? Response: 1 Specifying the MIN parameter returns a value of 0.1. Specify ing the ...

  • Fluke 1529 Chub-E4 - page 122

    SYST :COMM:SER:FDUP? 1 A val ue of 1 is returned if serial port echo (duple x) is ON. A v alue of 0 is re - turned if serial port echo (duple x) is OFF . Specifying the MIN parameter re - turns a value of 0. Specifying the MA X parameter returns a value of 1. Specifying the DEF parameter returns a value of 1. 8.4.9.10 SYST em:COMMunicate:SERial:FDU ...

  • Fluke 1529 Chub-E4 - page 123

    8.4.9.14 SYST em:COMMunicate:SERial:LINefeed <bool>|MIN|MAX|DEF Sets the state of the serial po rt linefeed setting. SYST :COMM:SER:LIN ON The <bool> parameter turns li nefeed on (1 or ON) or off (0 or OFF). Spec - ifying the MIN, MAX , or DEF parameter sets the linefeed to the m inimum, maximum, or default respecti vely . 8.4.9.15 SYST ...

  • Fluke 1529 Chub-E4 - page 124

    The <bool> parameter turns t he ke ypad lockout on (1 or ON) or of f (0 or OFF). Specifying the MIN, MAX, or DEF parameter sets the k eypad lock out to the minimum, maximum, or default respecti vely . 8.4.9.19 SYST em:PO W er:B A TT er y? Returns the current battery ch ar ge le vel. SYST :POW :B A TT? 90 The returned val ue indicates the curr ...

  • Fluke 1529 Chub-E4 - page 125

    The <num> parameter specifies the number of the desired date format. See T a - ble 11 on page 130, Date Formats, for a list of acceptable time formats. Spec - ifying the MIN, MAX, or DEF parameter sets the date format to the minimum, maximum, or default respecti vely . 8.4.10.3 DISPlay:TIME:FORMat? [MIN|MAX|DEF] Returns a number indicating th ...

  • Fluke 1529 Chub-E4 - page 126

    8.4.10.7 SYST em:TIME? Returns the current system time. SYST :TIME? 11,43,23 The response is returned in the format, <hour>,<minute>,<second>. Note: The system time is a lw ays returned using 24-hour format rega rdless of the current time for mat setting. 8.4.10.8 SYST em:TIME <hour>,<minute>,<second> Sets the cu ...

  • Fluke 1529 Chub-E4 - page 127

    SYST :P ASS:CEN:ST A T? 0 The response is either 0 (disabled) or 1 (enabl ed). A response of 0 indicates password-protected settings cannot be accessed. A response of 1 indicates the password was successful ly entered and password-protect ed settings can be accessed. 8.4.11.4 SYST em:P ASSwor d:CONV ersion? [MIN|MAX|DEF] Returns the current probe p ...

  • Fluke 1529 Chub-E4 - page 128

    8.4.12 Status Commands The status commands are u sed for querying and settin g the state of the instru - ment regi sters. For details on t he operation of the status re gisters refer to the IEEE-488.2 and SCPI standards. 8.4.12.1 *CLS Clears the status registers. *CLS 8.4.12.2 *ESE? Returns the Event Status Enable Register . *ESE? 255 8.4.12.3 *ESE ...

  • Fluke 1529 Chub-E4 - page 129

    8.4.12.8 *TST? Performs a self-test and reports an y errors that are found. *TST? 0 8.4.12.9 ST A T us:MEASure? Reads and clears the Measurement Status Register . ST A T :MEAS? 1 The assignment of the b its in this re gister is as follows: bit 0: channel 1 measurement complete bit 1: channel 2 measurement complete bit 2: channel 3 measurement compl ...

  • Fluke 1529 Chub-E4 - page 130

    ST A T :MEAS:ENAB 15 8.4.12.13 ST A T us:OPERation? Reads and clears the Operation Status Register . ST A T :OPER? 16 The assignment of the b its in this re gister is as follows: bits 0-3: not used bit 4: measurement complete bits 5-15: not used 8.4.12.14 ST A T us:OPERation:CONDition? Returns the Operation Status Conditio n Re gister . ST A T :OPE ...

  • Fluke 1529 Chub-E4 - page 131

    bit 4: measurement in valid bits 5-15: not used 8.4.12.18 ST A T us:QUEStionab le:CONDition? Returns the Questio nable Status Condition Register . ST A T :QUES:COND? 16 The assignment of the b its in this re gister is as follows: bits 0-3: not used bit 4: measurement in valid bits 5-15: not used 8.4.12.19 ST A T us:QUEStionable:ENABle? [MIN|MAX|DEF ...

  • Fluke 1529 Chub-E4 - page 132

    8.4.13 Statistical Calculation T ypes The statistical calculation types are listed in the following table. 8.4.14 Field T ypes The bottom portion of t he main display is capabl e of sho wing up to eight f ields. These fields are selectable from the following table. Number Keyw ord Fiel d T ype 0 ---- None 1D A T E D a t e 2T I M E T i m e 3 TYP Con ...

  • Fluke 1529 Chub-E4 - page 133

    Number Keyw ord Fiel d T ype 18 ALOG Logging Indicator 19 BA TT Bat tery Percent Remaini ng 20 MEAS Number of Seconds Until Ne xt Measurement Note: When the field type reference (REF) is selected, th e displayed ke yword changes depending on the probe type. T he reference display ke ywords are de - tailed in the following table. Probe Type Keyword ...

  • Fluke 1529 Chub-E4 - page 134

    8.4.15 Con version T ypes and P arameters T able 7 lists the probe types, con version types, mnemonics, and serial port re - sponses. T able 8 lists the con versioin types and their parameters. Probe Type Conversion Type Mnemonic Port Response SPRT, PRT, RTD, Thermistor ITS-90 (default ) ITS-SR5 PT-100 CVD THRM R(T) THRM T(R) YSI-400 R(W) ITS-90 IT ...

  • Fluke 1529 Chub-E4 - page 135

    Conversion Type Parameters TC-K VIN[mV] TC-B TC-E TC-J TC-N TC-R TC-S TC-T TC-AU/PT TC-POLY RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 <none> RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 RJC, RJT, T1, ADJ1, T2, ADJ2, T3, ADJ3 RJC, RJT, T ...

  • Fluke 1529 Chub-E4 - page 136

    8.4.17 Date and Time Formats The following tables list the acceptable date and time formats and their number setting. Number Format 0 MM-DD-YY 1 MM-DD-YYYY 2 DD/MM/YY 3 DD/MM/YYYY Number Format 0 H:MM:SSAM/PM (12-hour f ormat) 1 HH:MM:SS (24-hour f ormat) 1529 Chub-E4 Thermometer Readout User’s Guide 130 Ta b l e 1 1 Date Formats Ta b l e 1 2 Tim ...

  • Fluke 1529 Chub-E4 - page 137

    9 Calibration Each channel of the 1529 is cali brated independently of the other channels and has its o wn set of calibration parameters. The 1529 should be calibrated at reg - ular intervals t o ensure that it continues to measure with proper accuracy . Cali - bration should only be done by qualif ied, authorized personnel. When t he 1529 is recal ...

  • Fluke 1529 Chub-E4 - page 138

    The calibration parameters C0, C100, and C400 can be adjusted to opti mize the accuracy . These password-protected parameters are accessed from the CALI - BRA TION function of the SYSTEM menu. 9.1.2 PR T Calibration Procedure Calibration requires adjustment of the C0, C100, and C400 parameters at three specific i nput resistances. If the resistance ...

  • Fluke 1529 Chub-E4 - page 139

    3. Connect a 400 Ω resistor (6 ppm accuracy) to the input and measure its resistance. Note the av erage error in the measurement. Adjust the C400 parameter by subtracting the measured error. F or example, if the input is exactly 400.0000 Ω and the readout shows 399.9913 Ω , adjust the C400 parameter by adding 0.0087. 4. Record the date in the D ...

  • Fluke 1529 Chub-E4 - page 140

    2. Connect a 10 k Ω resistor (25 ppm accuracy) to the input and measure its resistance. Note the av erage error in the measurement. Adjust the C10K parameter by subtracting the measured error. F or example, if the input is exactly 10.00000 k Ω and the readout shows 10.00295 k Ω , adjust the C10K parameter by subtracting 2.95. 3. Connect a 100 k ...

  • Fluke 1529 Chub-E4 - page 141

    Par ameter Description C0 Adjust s the vol tage accuracy of at 0 m V C100 Adjusts the v oltage accuracy of at 100 mV CRJ Adjusts the internal RJC accuracy D A TE Records the date the thermocouple channel was calibrat ed DUE Records the date the thermocouple channel calibr ation e xpires The calibration parameters can be acce ssed from the CALIBRA T ...

  • Fluke 1529 Chub-E4 - page 142

    sured error . For e xample, if the thermocouple temperature is actually 0.0°C but the 1529 reads 0.184°C, the CRJ parameter should be adjusted by subtracting 0.184. If it was previously 0.000 it should now be –0.184. 6. Record the date with the D A TE parameter . Set the DUE date to the de - sired calibration interval. 7. Repeat the procedure f ...

  • Fluke 1529 Chub-E4 - page 143

    10 Maintenance • The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. A void operating the instrument in oily , wet, dirty , or dusty en vironments. ...

  • Fluke 1529 Chub-E4 - page 144

    11 T r oubleshooting In case you run into dif ficulty while operati ng the 1529 system, this sect ion provides some suggestions t hat may help you solve the problem. Belo w are se veral si tuations that may arise fol lowed by possi ble causes and suggested ac - tions you might tak e. 11.1 T roub leshooting The follo wing table lists possibl e error ...

  • Fluke 1529 Chub-E4 - page 145

    Problem Solution Incorrect resi stance or voltage reading While attempti ng to measure resistance or voltage, the display shows an incorr ect value or no value at all ( ..........) . Consider the follo wing possibilities: Poor or incorrect connecti on of the probe. A common mistake is to connect the wi res of the probe to the wrong ter minals. Chec ...

  • Fluke 1529 Chub-E4 - page 146

    Message Recommended Action SYSTEMS: IR POR T F AILURE Cycle the power . SYSTEMS: BA TTER Y NOT INST ALLED Cycle the pow er . Check that t he battery is properly connected. Replace the battery . CHANNELS: F AILURE Cycle the pow er . CALIBRA TION: DUE FOR RECAL Check that the instrument date is correct. Recalibrat e the instrument and set the recal i ...

  • Fluke 1529 Chub-E4 - page 147

    a) Set up the terminal program to use the correct COM port and baud rate. Conf igure the communicat ion settings to use 8 data bits, 1 st op bit, no parity and no flo w control. b) If you want to sa ve the do wnloaded data to a file, conf igure your ter - minal program to capture the inco ming text and sa ve it to a text f ile. Re - fer to your ter ...

  • Fluke 1529 Chub-E4 - page 148

    The instrument was designed specif ically as a test and measuring device. Com - pliance to the EMC di recti ve is t hrough IEC 61326-1 Electrical equipment for measur ement, contr ol and laborat ory use – EMC r equir ements (1998). As noted in the IEC 61326-1, the instrument can ha ve va rying confi gurations. The instrument w as tested in a typi ...

  • Fluke 1529 Chub-E4 - page 149

    4. Ho w long does the battery last from full char ge? About 8 hours from full charge with the backlight off. 5. Ho w do I know when the battery is at full charge? The charging indica - tor LED located on the back panel displays amber while charging and then turns green when the batteries are fully charged (only when the AC adapter is connected). Th ...

  • Fluke 1529 Chub-E4 - page 150

    2. What are the implications of the fast measurement mode in terms of ac - curacy? Accuracy is reduced when using the Model 1529 in fast mea - surement mode. Refer to Section 8.1.1 for a better description of the f ast measurement mode and its limitations. 3. Why can I only see one channel in the fast measurement mode? Because of the time required ...

  • Fluke 1529 Chub-E4 - page 151

    11.5.4 Output 1. Ho w do I get data from the Model 1529 to my PC? Data can be trans - ferred to your PC via an RS-232 interf ace or by IR communications through the use of the optional IR dongle and software available through Hart. 2. Is the IR output on the Model 1529 compatible with the IR port on my desktop laser printer? The IR port on the Mode ...

  • Fluke 1529 Chub-E4 - page 152

    Inde x ! 2-wire probe 39 - 40 3-wire probe 39 - 40 4-wire probe 39 - 40 A A C adapter 21, 25 accessories 19 ASTM E1137 40 AU T O L O G 6 0 AU T O - C A L 3 5 av erage 33 B backlight 25, 35, 53 battery 21, 25 charging 21 baud rate 71 butt ons 17, 141 C calibration 131, 141 CALIBRA TION 75 CALIBRA TION EXPIRED 141 Callendar -V an Dusen 40 channel sel ...

  • Fluke 1529 Chub-E4 - page 153

    E EDIT FIELDS 51 EDIT PR OBE 36 EMC 142 EN ABLE CHANNEL 32 en vironment 10 error 140 exi t 17, 29 F fast measurement 26, 31 features 7, 17 field data 49 - 51 FIELDS menu 47 filter 33 G GPIB 79 GPIB port 72 I IEC-751 40 IEEE-488 72, 79 IR period 72 IR port 72, 142 IRD A 72 ITS-90 38 - 39 L lamp 35, 53 linefeed 71 log count 62 LOG HISTOR Y 56 log per ...

  • Fluke 1529 Chub-E4 - page 154

    serial number 37 type 37 wires 23, 39 - 41 PR OBE menu 35 PR T 2 3 PT -100 40 R rack mounting 19 range resistance 39 - 41 reference junction 43, 50 reset 77 resistance 41, 50 resolution 35, 53 RJC 43 RS-232 71, 79 RT D 4 0 S scan 33 SELECT FIELDS 49 self-test 25, 140 sensor 23 serial interface 79 serial number 37, 69 serial period 71 serial port 71 ...

Manufacturer Fluke Category Thermometer

Documents that we receive from a manufacturer of a Fluke 1529 Chub-E4 can be divided into several groups. They are, among others:
- Fluke technical drawings
- 1529 Chub-E4 manuals
- Fluke product data sheets
- information booklets
- or energy labels Fluke 1529 Chub-E4
All of them are important, but the most important information from the point of view of use of the device are in the user manual Fluke 1529 Chub-E4.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Fluke 1529 Chub-E4, service manual, brief instructions and user manuals Fluke 1529 Chub-E4. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Fluke 1529 Chub-E4.

Similar manuals

A complete manual for the device Fluke 1529 Chub-E4, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Fluke 1529 Chub-E4 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Fluke 1529 Chub-E4.

A complete Fluke manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Fluke 1529 Chub-E4 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Fluke 1529 Chub-E4, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Fluke 1529 Chub-E4, that we can find in the current document
3. Tips how to use the basic functions of the device Fluke 1529 Chub-E4 - which should help us in our first steps of using Fluke 1529 Chub-E4
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Fluke 1529 Chub-E4
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Fluke 1529 Chub-E4 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Fluke 1529 Chub-E4?

Use the form below

If you did not solve your problem by using a manual Fluke 1529 Chub-E4, ask a question using the form below. If a user had a similar problem with Fluke 1529 Chub-E4 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)