Manual Invacare IH6300ADH

32 pages 1.94 mb
Download

Go to site of 32

Summary
  • Invacare IH6300ADH - page 1

    Owner’s Operation and Maintenance Manual DEALER: This manual MUST be given to the user of the tub. USER: BEFORE using this tub, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare-ccg.com Supine Bath Tub IH6300 Fixed Height IH6300ADH Adjustable Height AFT ...

  • Invacare IH6300ADH - page 2

    Supine Bath Tub 2 Part No 1096926  WARNING DO NOT USE THIS PRODUCT OR AN Y AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READ ING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTION AL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MA NUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OP TIONAL EQUIPMENT. IF YOU ARE U ...

  • Invacare IH6300ADH - page 3

    TABLE OF CONTENTS Part No 1096926 3 Supine Bath Tub TABLE OF CONTENTS SPECIAL NOTES ................................................................................ 5 LABEL LOCATION ............................................................................ 6 TYPICAL PRODUCT PARAMETERS .................................................... 7 FEATURE ...

  • Invacare IH6300ADH - page 4

    TABLE OF CONTENTS Supine Bath Tub 4 Part No 1096926 TABLE OF CONTENTS SECTION 5—MAINTE NANCE ........................................................... 25 Parts Information ................ ...................... .......................................................................................... ......25 Safety Inspection Checklis ts .... ...

  • Invacare IH6300ADH - page 5

    SPECIAL NOTES Part No 1096926 5 Supine Bath Tub SPECIAL NOTES Signal  words  are  use d  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  table  below  for  defini ...

  • Invacare IH6300ADH - page 6

    LABEL LOCATION Supine Bath Tub 6 Part No 1096926 LABEL LOCATION ! W ARNING ELECTRICAL SHOCK HAZARD. All disassembly and maintenance of the Supine T ub MUST be done by a qualified technician, certified electrician or plumber. Disconnect electrical supply or turn OFF circuit breaker before performing an y maintenance to the Supine T ub. 1097132 Rev . ...

  • Invacare IH6300ADH - page 7

    TYPICAL PRODUCT PARAMETERS Part No 1096926 7 Supine Bath Tub TYPICAL PRODUCT PARAMETERS IH6300 FIXED HEIGHT IH6300ADH ADJUSTABLE HEIGHT Height Tub Rim 32 inches (813 mm) Raised N/A 41 inches (1041 mm ) Lowered N/A 27 inches (686 mm) Overall Tub Height 56 inches (1422 mm) Raised N/A 65 inches (1651 mm ) Lowered N/A 51 inches (1295 mm ) Width Tub Ext ...

  • Invacare IH6300ADH - page 8

    FEATURES Supine Bath Tub 8 Part No 1096926 FEATURES CONTROL PANEL TUB INTERIOR Model IH6300ADH Adjustable Height Tub Tub Fill ON/OFF Disinfectant Flow Mete r Disinfectant Wand Disinfectant Cabinet Water Temperature Thermometer Shower ON/OFF Tub Fill Spou t Disinfec tant On/Off Whirlpool ON/OFF Model IH6300 Fixed Height Tub Tub Controls UP/Down Whir ...

  • Invacare IH6300ADH - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1096926 9 Supine Bath Tub SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUIDELIN ES contai ns important inform ation for the safe operation and use of this product. Check  all  parts  for  shipping  damage  before  using .  In  case  of  damage,  DO  NOT ...

  • Invacare IH6300ADH - page 10

    SECTION 1—GENERAL GUIDELINES Supine Bath Tub 10 Part No 1096926 When  operating  wa t e r  pressure  exceeds  55  p.s.i.,  pressure  regulators  (similar  to  the  ¾ ‐ inch  W atts  U5)  MUST  be  installed. Maintenance and Inspection If  any  part  of  the  system  is  NO ...

  • Invacare IH6300ADH - page 11

    SECTION 1—GENERAL GUIDELINES Part No 1096926 11 Supine Bath Tub The  Supine  Tu b  is  specifically  designed  to  be  used  with  a  bath  lift.  ICCG  recommends  use  of  the  T raverse  Lift  Model  1200  ONL Y .  Bath  lifts  designed  by  other  manufacturer ...

  • Invacare IH6300ADH - page 12

    SECTION 2—DISINFECTING THE TUB Supine Bath Tub 12 Part No 1096926 SECTION 2—DISINFECTING THE TUB  WARNING Disinfect the tub after EACH use to avoid resi dent infectio n and contamination of the tub. DO NOT disinfect the tub when it is occupied otherwise injury may o ccur. Read and understand all info rmation on Invacare brand disinfectant be ...

  • Invacare IH6300ADH - page 13

    SECTION 3—OPERATION Part No 1096926 13 Supine Bath Tub SECTION 3—OPERATION Overview The  tu b  is  designed  to  improve  the  efficiency  and  environmental  safety  of  you r  facility ʹ s  operation.  Howev er ,  if  the  tub  is  not  operated  or  maintained  properly ...

  • Invacare IH6300ADH - page 14

    SECTION 3—OPERATION Supine Bath Tub 14 Part No 1096926 Operating the Tub  WARNING DO NOT stand in tub. Tub surfaces are slippery wh en wet. Serious injury may occur. DO NOT stand on tub frame. Frame is slippery when wet. BEFORE immersing the resident or oper ating the shower/mas sage wand, check that the water temperature is between 95° F and ...

  • Invacare IH6300ADH - page 15

    SECTION 3—OPERATION Part No 1096926 15 Supine Bath Tub Operating Shower/Massage Wand NOTE:  STEPS  1 ‐ 3  are  to  be  performed  while  holding  the  shower/massage  wand  over  the  tu b.  Refer  to  Features  on  page 8  for  additional  illustrations. 1. T u rn  show ...

  • Invacare IH6300ADH - page 16

    SECTION 3—OPERATION Supine Bath Tub 16 Part No 1096926 Adjusting Tub Height - Model IH6300ADH ONLY  WARNING DO NOT stand on tub frame. Frame is slip pery when wet. Foot injury is possible when adjusting tub height. CAUTION DO NOT adjust height more than three cycles (up and down) within a ten minute period. Damage to th e actuator may result. ...

  • Invacare IH6300ADH - page 17

    SECTION 4—INSTALLATION Part No 1096926 17 Supine Bath Tub SECTION 4—INSTALLATION Preparing the Work Area NOTE:  For  this  proced ure,  ref er  to  FIGURE 4.1. 1. Determine  location  for  the  supine  bath  tub  installation. NOTE:  Follow  r ecommended  floor  plan  when  installing  ...

  • Invacare IH6300ADH - page 18

    SECTION 4—INSTALLATION Supine Bath Tub 18 Part No 1096926 Securing the Tub NOTE:  For  this  proced ure,  ref er  to  FIGURE 4.1  on  page 17  and  FIGURE 4.2. 1. Set  the  tub  in  place. 2. Mark  location  of  anchor  holes,  using  pre ‐ drilled  holes  in  leg  mounting ...

  • Invacare IH6300ADH - page 19

    SECTION 4—INSTALLATION Part No 1096926 19 Supine Bath Tub Electrical Hook-up  WARNING Be certain ALL electrical wo rk is performed by a qualif ied electrician and is in compliance with local electrical codes. Before installing and connec ting the tub, make certai n that the electric power source is SHUT OFF to avoid electrical shock. NOTE:  ...

  • Invacare IH6300ADH - page 20

    SECTION 4—INSTALLATION Supine Bath Tub 20 Part No 1096926 FIGURE 4.3 Electrical Hook-up Electrical Hookup 115 V Electrical IH6300 Fixed Height Use 15 Amp GFCI Breaker Only IH6300ADH Adjustable Height Use 15 Amp Hospital Grade GFCI Receptacle Only DETAIL “B” - ELECTRICAL HOOK UP DET AILS DETAIL “A” - ELECTRICAL HOOK UP LOCATION Wall Tub Wa ...

  • Invacare IH6300ADH - page 21

    SECTION 4—INSTALLATION Part No 1096926 21 Supine Bath Tub Electrical Installation  DANGER The National Electric Code re quires the use of a ground faul t circuit interrupter on all therapeutic pools and tubs (h ydrotherapeutic tanks). Failure to follow this requirement could result in seri ous or fatal injury. ELECTRICAL SHOCK HAZARD. All dis ...

  • Invacare IH6300ADH - page 22

    SECTION 4—INSTALLATION Supine Bath Tub 22 Part No 1096926 Plumbing Hook-Up  WARNING Be certain ALL plumbing work is perf ormed by a qualified plumber and is in compliance with local codes. NOTE:  For  this  proced ure,  ref er  to  Refer  to  FIGURE 4.5. NOTE:  For  installation  details  re fe r  to  ...

  • Invacare IH6300ADH - page 23

    SECTION 4—INSTALLATION Part No 1096926 23 Supine Bath Tub Plumbing Installation  WARNING Be certain ALL plumbing work is performed by a qualified plumber. Some state and local pl umbing codes require the install ation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and cold supply lines to prevent back flow contaminati ...

  • Invacare IH6300ADH - page 24

    SECTION 4—INSTALLATION Supine Bath Tub 24 Part No 1096926 FIGURE 4.6 Plumbing Installation and Schematic DETAIL “A” - PLUMBING INSTALLATION Cold Water Inlet Hot Water Inlet Shower/Massage Wand HOT COLD Shower/Mas sage Wand Valve Tub Fill Valve Disinfectant Line Flow meter Disinfectant Valve Jet Pump In House Water Service RADA Mixing Valve Op ...

  • Invacare IH6300ADH - page 25

    SECTION 5—MAINTENANCE Part No 1096926 25 Supine Bath Tub SECTION 5—MAINTENANCE  WARNING Before performing any maintenance to the tub, make certain that the electric power source is shut off to avoid electrical shock. Be certain ALL electrical wo rk is performed by a qualif ied electrician and is in compliance with local codes. If any part of ...

  • Invacare IH6300ADH - page 26

    SECTION 5—MAINTENANCE Supine Bath Tub 26 Part No 1096926 Every Six Months ❑ Ve r i f y  the  wa t e r  fl ow  temperature  with  a  separate  thermometer . ❑ V isually  inspect  plumbing  for  leaks. ❑ Check  that  all  labels  are  present  and  legible.  Replace  if  necess ...

  • Invacare IH6300ADH - page 27

    SECTION 5—MAINTENANCE Part No 1096926 27 Supine Bath Tub Heavy Duty Cleaning Procedure  WARNING DO NOT use this product or any available option al equipment without first completely reading and understanding these inst ructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied w ...

  • Invacare IH6300ADH - page 28

    SECTION 5—MAINTENANCE Supine Bath Tub 28 Part No 1096926 Calibrating the Disinfectant Flowmeter  WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions a nd any additional instructional ma terial such as owner’s manuals, service manuals or instruction sheet s ...

  • Invacare IH6300ADH - page 29

    SECTION 5—MAINTENANCE Part No 1096926 29 Supine Bath Tub 15. Adjust  the  injector  by  performing  one  of  the  foll owing: • Bottom  of  Floating  Ba ll  Reads  Greater  Than  35 ‐  Less  disinfectant  needs  to  be  added  to  the  mixture.  Loosen  the  s ...

  • Invacare IH6300ADH - page 30

    SECTION 5—MAINTENANCE Supine Bath Tub 30 Part No 1096926 Maintenance Record DATE MAINTENANCE PERSON TYPE OF MAINTENANCE PERFORMED NOTES ...

  • Invacare IH6300ADH - page 31

    SECTION 6—TROUBLESHOOTING Part No 1096926 31 Supine Bath Tub SECTION 6—TROUBLESHOOTING NOTE:  If  pr oblems  are  not  rem ed i e d  by  the  suggested  means  above,  contact  an  Invacar e  Dealer . SYSTEM/OPERATION SYMPTOMS FAULTS SOLUTION Adjustable Height Model IH6300ADH Tub will not raise or lower ...

  • Invacare IH6300ADH - page 32

    LIMITED WARRANTY PLEASE NOTE: THE WA RRANTY BELOW HAS BE EN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MA N UFACTURED AFTER JULY 4, 1975. This warranty is extended only to the o riginal purchaser/user of our pro ducts. This warranty gives you speci fic legal rights and yo u may also have other legal ri ghts which vary from state to s ...

Manufacturer Invacare Category Hot Tub

Documents that we receive from a manufacturer of a Invacare IH6300ADH can be divided into several groups. They are, among others:
- Invacare technical drawings
- IH6300ADH manuals
- Invacare product data sheets
- information booklets
- or energy labels Invacare IH6300ADH
All of them are important, but the most important information from the point of view of use of the device are in the user manual Invacare IH6300ADH.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Invacare IH6300ADH, service manual, brief instructions and user manuals Invacare IH6300ADH. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Invacare IH6300ADH.

Similar manuals

A complete manual for the device Invacare IH6300ADH, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Invacare IH6300ADH by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Invacare IH6300ADH.

A complete Invacare manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Invacare IH6300ADH - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Invacare IH6300ADH, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Invacare IH6300ADH, that we can find in the current document
3. Tips how to use the basic functions of the device Invacare IH6300ADH - which should help us in our first steps of using Invacare IH6300ADH
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Invacare IH6300ADH
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Invacare IH6300ADH in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Invacare IH6300ADH?

Use the form below

If you did not solve your problem by using a manual Invacare IH6300ADH, ask a question using the form below. If a user had a similar problem with Invacare IH6300ADH it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)