Manual Graham Field 765000

24 pages 0.64 mb
Download

Go to site of 24

Summary
  • Graham Field 765000 - page 1

    765000 AL TERNA TING PRESSURE MA TTRESS SYSTEM WITH ACTIVE SENSOR TECHNOL OGY USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use. Info: The most current version of this manual can be found online at www.grahameld.com. 765000-INS-LAB-RevC11 ...

  • Graham Field 765000 - page 2

    2 765000-INS-LAB-RevC11 CONTENTS INTRODUCTION..................................................................................... 3 INTENDEDUSEOFTHISDEVICE ........................................................ 3 CONTRAINDICA TION ...................................................................... 3 IMPORT ANTSAFETY ...

  • Graham Field 765000 - page 3

    765000-INS-LAB-RevC11 3 INTRODUCTION Use this manual for set-up and operation of the Lumex ® AltaDyne ® Sensor Plus 765000 Alternating Pr essur e Mattr ess System with Active Sensor T echnology . Read all instructions befor e using the mattr ess system. Save this manual for futur e r efer ence. INTENDED USE OF THIS DEVICE The intended use of t ...

  • Graham Field 765000 - page 4

    4 765000-INS-LAB-RevC11 IMPORT ANT SAFETY PRECA UTIONS The safety statements pr esented in this chapter r efer to the basic safety information that should be obser ved by those using this Mattr ess System. Ther e ar e additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specic to ...

  • Graham Field 765000 - page 5

    765000-INS-LAB-RevC11 5 WARNING T o reduce the risk of bur ns, electrocution, r e, or per sonal injur y:  W ARNING:Donotleavethisproductunattendedwhen pluggedin.  W ARNING: Alwaysuseclosesupervisionwhenthis productisused b y, on,ornearchildrenorthosewho req ...

  • Graham Field 765000 - page 6

    6 765000-INS-LAB-RevC11  W ARNING:Patiententrapmentwithbedsiderails maycauseinjuryordeath.Thebedframeandits components,includingthemattress,bedsiderails, headandfootboard,bedding,andanyaccessories addedtothebed,canallaffectthe? ...

  • Graham Field 765000 - page 7

    765000-INS-LAB-RevC11 7 EQUIPMENT SYMBOLS Manufacturer “BF”symbol: Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Attention! Read the instructions! Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as hous ...

  • Graham Field 765000 - page 8

    8 765000-INS-LAB-RevC11 PRODUCT DESCRIPTION PUMP AND MA TTRESS SY STEM pumpandmattresssystem CPRvalves quickconnectors (mattress-pump connectors) mattress footend symbol pump PUMP: FRONT quickconnector front panel PUMP: REAR mainpower switch powercord receptacle fuse hanger filter ...

  • Graham Field 765000 - page 9

    765000-INS-LAB-RevC11 9 PUMP: FRONT P ANEL Fr ont panel of pump is shown below . Descriptions of buttons and indicators follow . T uningbutton; indicatorsatleftofbutton Powerfailureindicator Alarmmutebutton Serviceindicator Power(ON/OFF)button; indicatoratrightofbutton Panellockbutton; indica ...

  • Graham Field 765000 - page 10

    10 765000-INS-LAB-RevC11 Tuning button: The tuning button ne-tunes the mattr ess system's auto-adjusted pr essur e. Thr ee tuning settings ar e available: 0 System auto-adjusted pr essur e - One level softer than auto-adjusted pr essur e + One level rmer than auto-adjusted pr essur e Info: The tuning function is only available in alte ...

  • Graham Field 765000 - page 11

    765000-INS-LAB-RevC11 11 Maxfir m button: The pump will automatically enter maxrm mode ever y time the power is turned on. Maxrm mode ensur es the pump is able to r each its maximum operating pr essur e. Maxrm mode will last for 20 minutes, and then the system will automatically enter auto-adjust mode. T o disable maxrm mode, pr ess ...

  • Graham Field 765000 - page 12

    12 765000-INS-LAB-RevC11 Low pr essur e indicator: The low pressur e indicator (yellow LED) illuminates, and the alarm sounds, when an abnormally low pr essur e level occurs. If the low pr essur e indicator persistently illuminates inappr opriately: 1. Check to ensur e all connections ar e pr operly and secur ely connected per installation instr ...

  • Graham Field 765000 - page 13

    765000-INS-LAB-RevC11 13 INST ALLA TION UNP ACKING 1. Car efully r emove all components fr om the carton. 2. Inspect all components. 3. If damage is evident to the contents, please notify the carrier and your Graham-Field equipment pr ovider . PUMP AND MA TTRESS SETUP 1. Place the mattr ess on top of the bed frame; ensur e that the foot end symb ...

  • Graham Field 765000 - page 14

    14 765000-INS-LAB-RevC11  W ARNING:Locateallcordssothattheywillnotbe steppedon,trippedover ,orotherwisesubjectedto damageorstress.  W ARNING: Donotuseahouseholdextensioncordifthe electricalcorddoesnotreachthepoweroutlet.Useo ...

  • Graham Field 765000 - page 15

    765000-INS-LAB-RevC11 15 OPERA TION GETTING ST AR TED 1. T urn the main power switch to the ON position. 2. Pr ess the fr ont panel power button. The power indicator will illuminate. Info: The mattress system will automatically enter maxfirm mode for the quickest inflation at set-up; when inflated, it will then automatically enter auto- adjus ...

  • Graham Field 765000 - page 16

    16 765000-INS-LAB-RevC11 TRANSPORT CAP In case of power failur e or transport: Disconnect the quick connector and cap the mattr ess with the transport cap to slow deation. Info: This system is equipped with cell-on-cell technology to provide added support and prevent bottoming out in the event of a power failure. Lower cells will hold air for ...

  • Graham Field 765000 - page 17

    765000-INS-LAB-RevC11 17 MAINTENANCE CLEANING  W ARNING: To reducetheriskofincreasedbacterial growth,infection,illness,orinjuryfromcontamination, thoroughlycleananddrythemattresssystembefore useandasnecessaryduringuseasfollows: Pump : W ipe the pump with ...

  • Graham Field 765000 - page 18

    18 765000-INS-LAB-RevC11 Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug fr om the wall outlet immediately . 2. Use a small scr ewdriver to r emove the cover of the fuse holder (fuse locations ar e shown in pictur e at right). 3. Insert a new fuse of the correct rating (T1A/250V , VDE appr oved ...

  • Graham Field 765000 - page 19

    765000-INS-LAB-RevC11 19 SPECIFICA TIONS Pump Specication Power Supply (Info: see rating label on product) AC 120V 60 Hz 0.17A (for 120V system) Fuse Rating T1A, 250V Cycle time Fixed / 10 minutes Dimensions (L x W x H) 1 1.4" x 7.9" x 4.6" (29.1 x 20 x 11.7 cm) Weight 5 lb (2.2 kg) Environment T emperature Operation: 50°F to ...

  • Graham Field 765000 - page 20

    20 765000-INS-LAB-RevC11 LIMITED WARRANTY GF Health Pr oducts, Inc. warrants the Lumex ® AltaDyne ® Sensor Plus 765000 Alternating Pr essur e Mattr ess System with Active Sensor T echnology for a period of twelve months for defects in workmanship and materials. If a pr oduct is deemed to be under warranty , GF Health Pr oducts, Inc. shall pr o ...

  • Graham Field 765000 - page 21

    765000-INS-LAB-RevC11 21 APPENDIX A: EMC INFORMA TION Guidance and Manufacturer’s De claration- Electromagnetic Emissions: This device is intended for use in the electromagnetic environment specified below . The user of this device should make sure i t is used in such an environmen t. Emissions T est Compliance Electromagnetic Environment-Gui ...

  • Graham Field 765000 - page 22

    22 765000-INS-LAB-RevC11 Guidance and Manufact urer ’ s Declaration- Electromagn etic Immunity: This device is intended for use in the electromagnetic environment spec ified bel ow . T he user of t his device should ma ke sure it is us ed in such an environment. Immunity T est IEC60601 test level Compliance Electromagnetic Environment-Guidance ...

  • Graham Field 765000 - page 23

    765000-INS-LAB-RevC11 23 a The ISM (indu str ial, sc ient ific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 M Hz to 40,70 MHz. b The compliance levels in the ISM frequenc y bands between 150 kHz and 80 MHz and in the frequency ran ge 80 MHz to 2,5 GHz are i ...

  • Graham Field 765000 - page 24

    USA Corporate Headquar ters: GF Health Products, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 770-368-4700 fax: 770-368-2386 www .grahamfield.com © 2011 GF Health Products, Inc. ...

Manufacturer Graham Field Category Camping Equipment

Documents that we receive from a manufacturer of a Graham Field 765000 can be divided into several groups. They are, among others:
- Graham Field technical drawings
- 765000 manuals
- Graham Field product data sheets
- information booklets
- or energy labels Graham Field 765000
All of them are important, but the most important information from the point of view of use of the device are in the user manual Graham Field 765000.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Graham Field 765000, service manual, brief instructions and user manuals Graham Field 765000. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Graham Field 765000.

A complete manual for the device Graham Field 765000, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Graham Field 765000 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Graham Field 765000.

A complete Graham Field manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Graham Field 765000 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Graham Field 765000, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Graham Field 765000, that we can find in the current document
3. Tips how to use the basic functions of the device Graham Field 765000 - which should help us in our first steps of using Graham Field 765000
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Graham Field 765000
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Graham Field 765000 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Graham Field 765000?

Use the form below

If you did not solve your problem by using a manual Graham Field 765000, ask a question using the form below. If a user had a similar problem with Graham Field 765000 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)