Инструкция обслуживания National Instruments NI 9217

28 страниц 0.44 mb
Скачать

Перейти на страницу of 28

Summary
  • National Instruments NI 9217 - page 1

    OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9217 4-Channel, 24-Bit, 100 Ω RTD Analog Input Module ni.com/manuals Deutsch Français ...

  • National Instruments NI 9217 - page 2

    NI 9217 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 217 and incl u des specif ications and term inal assignments for the NI 9217. V isit ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing. For informa tion a ...

  • National Instruments NI 9217 - page 3

    © National Inst ruments Corp. 3 NI 9217 Operating Instructi ons and Specific ations Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines for Hazardous V oltages If hazardo u s voltages are connected to the mod u le, ta ke the follow ing preca u tions. A hazardo ...

  • National Instruments NI 9217 - page 4

    NI 9217 Operating I nstruction s and Specificati ons 4 ni.com Fig u re 1 sho ws the NI 9939 connector backshell. Figure 1. NI 9939 Connector Bac kshell Safety Guidelines fo r Hazardous Loca tions The NI 9217 is s u itable for u se in Class I, Di vision 2, Gro u ps A, B, C , D , T 4 h a z a r d o u s locations; Class I, Zone 2, AEx nC IIC T4, and Ex ...

  • National Instruments NI 9217 - page 5

    © National Inst ruments Corp. 5 NI 9217 Operating Instructi ons and Specific ations only . Follo w these g u ideline s if yo u are installing the NI 9217 in a potentially explosi ve environm e nt. Not follo wing these g u idelines may res u lt in serio u s inj u ry or death. Caution Do not disconnect I/O-side wires or connectors u nless power has ...

  • National Instruments NI 9217 - page 6

    NI 9217 Operating I nstruction s and Specificati ons 6 ni.com Special Conditions for Hazard ous Locations Use in Europe This eq u ipment has been e val u ated as EEx nC IIC T4 eq u ipment u nder DEMK O Certificate No. 03 A TEX 0324020X. Each mod u le is marked II 3G and is s u itable for u se in Zone 2 hazardo u s locations. If yo u are u sing the ...

  • National Instruments NI 9217 - page 7

    © National Inst ruments Corp. 7 NI 9217 Operating Instructi ons and Specific ations Connecting the NI 9217 The NI 9217 has a 16-terminal, de tachable scre w-terminal connector that provides con nections for fo u r R TD channels. Figure 2. NI 9217 T erminal Assignments EX0 RTD0+ RTD0– COM EX1 RTD1+ RTD1– COM EX2 RTD2+ RTD2– COM EX3 RTD3+ RTD3 ...

  • National Instruments NI 9217 - page 8

    NI 9217 Operating I nstruction s and Specificati ons 8 ni.com Yo u can connect any combination of 3-wire or 4-wire R TDs to the NI 9217. Each NI 92 17 channel contains circ u itry that a u tomatically detect s the type of R TD connected to the channel. The NI 9217 scans all fo u r inp u t channels and a u tomatically config u res each channel for t ...

  • National Instruments NI 9217 - page 9

    © National Inst ruments Corp. 9 NI 9217 Operating Instructi ons and Specific ations Figure 3. Input Circuitry for On e Channel of the NI 9217 in 4-Wire Mode Each channel has one terminal, R TD+, to whic h yo u connect the positiv e lead of the R TD signal and one terminal, R TD–, to which yo u connect the negati ve lead of th e R TD signal. Each ...

  • National Instruments NI 9217 - page 10

    NI 9217 Operating I nstruction s and Specificati ons 10 ni.com If yo u are u sing a 4-wire R TD, connect EX to the positi ve lead of the R TD. If yo u are u sing a 3-wire R TD, do not connect the R TD to the EX terminal. F or the best meas u rement res u lts when u sing the NI 9217 with a 3-wire R TD, u se eq u al-length wires between the R TD+ ter ...

  • National Instruments NI 9217 - page 11

    © National Inst ruments Corp. 11 NI 9217 Operating Instructi ons and Specific ations Wiring for High-Vib ration Applications If an application is s u bject to high vibrat ion, National Instr u ments recommends that yo u either u se ferr u les to terminate wires to the detachable screw-term inal connector or u se the NI 9939 backshell kit to protec ...

  • National Instruments NI 9217 - page 12

    NI 9217 Operating I nstruction s and Specificati ons 12 ni.com Sleep Mode This mod u le s u pports a low-po wer sleep mode. S u pport for sleep mode at the system le vel depe nds on the chassis that the mod u le is pl u gged into. Refer to the chassis man u al for information abo u t s u pport for sleep mode. If the chassis s u pports sleep mode, r ...

  • National Instruments NI 9217 - page 13

    © National Inst ruments Corp. 13 NI 9217 Operating Instructi ons and Specific ations Specifications The follo wing specifications a re typical for the range – 40 to 70 °C u nless otherwise noted. All v oltages are relat iv e to COM u nless otherwise noted. All specifications gi ven in °C are specif ic to 100 Ω platin u m R TDs. Input Characte ...

  • National Instruments NI 9217 - page 14

    NI 9217 Operating I nstruction s and Specificati ons 14 ni.com Con version time High-resol u tion mode....... .......... 200 ms per channel, 800 ms total for all channels High-speed mode .......... ........... ... 2 .5 ms per channel, 10 ms total for all channels Te m p e r a t u re acc u racy (incl u ding noise) 1 , 4-wire mode Te m p e r a t u re ...

  • National Instruments NI 9217 - page 15

    © National Inst ruments Corp. 15 NI 9217 Operating Instructi ons and Specific ations Noise High-resol u tion mode......... ........ 0.003 °C High-speed mode .......... ........... ... 0 .02 °C Excitation c u rrent.................. ........... . 1 m A per channel Noise rejection Normal mode (50/60 Hz) High-resol u tion mode .......... 85 dB m in ...

  • National Instruments NI 9217 - page 16

    NI 9217 Operating I nstruction s and Specificati ons 16 ni.com High-resol u tion f ilter response 1,2 1 This im age is provid ed co u rtesy of Linear T echnology Corp. 2 High-speed filter response has the s ame characteristics as the high-resol u t ion filter response except that the first notch is at 14 kHz. –20 –0 –40 –60 –80 –100 – ...

  • National Instruments NI 9217 - page 17

    © National Inst ruments Corp. 17 NI 9217 Operating Instructi ons and Specific ations Overv oltage protection ....... ........... ... ±30 V between inp u ts MTBF .......... ........... ............ ........... ... 891,597 ho u rs at 25 °C; Bellcore Iss u e 2, Method 1, Case 3, Limited P art Stress Method Note Contact NI for Bellcore MTBF specific ...

  • National Instruments NI 9217 - page 18

    NI 9217 Operating I nstruction s and Specificati ons 18 ni.com Physical Characteristics If yo u need to clean the mod u l e, wipe it with a dry to wel. Screw-terminal wiring ..... .............. ... 16 to 28 A WG copper cond u ctor wire with 7 mm (0.28 in.) of ins u lation stripped from the end To r q u e for screw terminals ..... .......... 0.22 t ...

  • National Instruments NI 9217 - page 19

    © National Inst ruments Corp. 19 NI 9217 Operating Instructi ons and Specific ations Safety Maximum Voltage 1 Connect only v oltages that are within the follo wing limits. All terminals-t o-COM....... ........... .......±30 V Isolation Voltages Channel-to-channel... .............. ..........None Channel-to-earth gro u nd Contin u o u s .......... ...

  • National Instruments NI 9217 - page 20

    NI 9217 Operating I nstruction s and Specificati ons 20 ni.com Caution Do not connect the NI 9217 to signals or u se for meas u rements within Meas u rement Categories III or IV . Safety Standards This prod u ct is designed to meet the req u ire ments of the follo wing standards of safety for electrical eq u ipment for meas u rement, control, and l ...

  • National Instruments NI 9217 - page 21

    © National Inst ruments Corp. 21 NI 9217 Operating Instructi ons and Specific ations Hazardous Locations U.S. (UL) .......... ......... ........... ........... . C lass I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4 Canada (C-UL) .............. ........... ........ Class I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone ...

  • National Instruments NI 9217 - page 22

    NI 9217 Operating I nstruction s and Specificati ons 22 ni.com Operating h u midity (IEC 60068-2-56)....... ........... ............ 10 to 90% RH, noncondensing Storage h u midity (IEC 60068-2-56)....... ........... ............ 5 to 95% RH, noncondensing Maxim u m altit u de.......... ......... .......... 2,000 m Poll u tion Degree (IEC 60664) ... ...

  • National Instruments NI 9217 - page 23

    © National Inst ruments Corp. 23 NI 9217 Operating Instructi ons and Specific ations Electromagnetic Compatibility This prod u ct is designed to meet the req u ire ments of the follo wing standards of EMC for electrical eq u ipment for meas u remen t, control, and laboratory u se: • EN 61326 EMC req u irements; Ind u strial Imm u nity • EN 550 ...

  • National Instruments NI 9217 - page 24

    NI 9217 Operating I nstruction s and Specificati ons 24 ni.com ni.com/certification , search by mod u le n u mber or prod u ct line, and click the appropriate link in the Certification col u mn. Environmental Mana gement National Instr u ments is committe d to designing and man u fact u ring prod u cts in an en vironmentally res ponsible manner . N ...

  • National Instruments NI 9217 - page 25

    © National Inst ruments Corp. 25 NI 9217 Operating Instructi ons and Specific ations Calibration Yo u can obtain the calibration cert ificate and information abo u t calibration services for th e NI 9217 at ni.com/calibration . Calibration interval ............... ............ 1 ye ar ⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё ೑ RoHS ˅ Ё೑ ...

  • National Instruments NI 9217 - page 26

    NI 9217 Operating I nstruction s and Specificati ons 26 ni.com Where to Go for Support The National Instr u ments W eb site is yo u r comple te reso u rce for technical s u pport. At ni.com/ s upport yo u hav e access to ev erything from tro u bleshooting and applic ation dev elopment self-help reso u rces to email and phone assistance from NI Appl ...

  • National Instruments NI 9217 - page 27

    © National Inst ruments Corp. 27 NI 9217 Operating Instructi ons and Specific ations K orea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 0 1 800 010 0793, Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 3390150, Port u gal 351 210 311 210, R u ssia 7 495 783 6851, Singa ...

  • National Instruments NI 9217 - page 28

    © 2006–2008 Nati onal Instruments Corp. All rights reserved. 374187C-01 May08 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments t rademarks. Other product and company names ment ioned herein are trad ...

Производитель National Instruments Категория Network Card

Документы, которые мы получаем от производителя устройства National Instruments NI 9217 мы можем разделить на несколько групп. Это в частности:
- технические чертежи National Instruments
- инструкции обслуживания NI 9217
- паспорта изделия National Instruments
- информационные брошюры
- энергетические этикетки National Instruments NI 9217
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания National Instruments NI 9217.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа National Instruments NI 9217, инструкции обслуживания, короткие инструкции или инструкции пользователя National Instruments NI 9217. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия National Instruments NI 9217.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства National Instruments NI 9217, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании National Instruments NI 9217 пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей National Instruments NI 9217.

Полная инструкция обслуживания National Instruments, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией National Instruments NI 9217 - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания National Instruments NI 9217, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся National Instruments NI 9217, которое найдем в данном документе
3. Советы по использованию основных функций устройства National Instruments NI 9217 - которые должны облегчить нам первые шаги во время использования National Instruments NI 9217
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем National Instruments NI 9217
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра National Instruments NI 9217 в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся National Instruments NI 9217?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с National Instruments NI 9217, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с National Instruments NI 9217 со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)