Инструкция обслуживания Teledyne HFC-303/307

35 страниц 0.7 mb
Скачать

Перейти на страницу of 35

Summary
  • Teledyne HFC-303/307 - page 1

    INSTRUCTION MANUAL HFM-301/305/306 FLOW METERS HFC-303/307 FLOW CONTROLLERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED ...

  • Teledyne HFC-303/307 - page 2

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 2 of 35 Manual Print History The print history shown below lists the printing da tes of all revisions and addenda created for this manual. The revision level letter increases alphabetic ally as the manual undergo es subsequent updates. Addenda, which are released between revisions, c ontain impor ...

  • Teledyne HFC-303/307 - page 3

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 3 of 35 Table of Contents 1. GENERAL INFO RMATIO N ............................................................................................................................................ 4 1.1. F EATURES ........................................................................................ ...

  • Teledyne HFC-303/307 - page 4

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 4 of 35 1. General Information The Hastings 300 Series hi gh range mass flow meters are des igned to accurately measu re mass flow at a range between 25 slm and 10 000 slm. The Hastings 300 Series high range ma ss flow controllers are designed to accurately measure an d control mass flow at a ran ...

  • Teledyne HFC-303/307 - page 5

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 5 of 35 1.2. Specifications Accuracy .......................................................... HFM-301/3 05 & HFC-303/307: ±1% full scale (F.S.) ........................................................................................................... HFM-306; ±3% full sc ale (F.S.) Repea ...

  • Teledyne HFC-303/307 - page 6

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 6 of 35 1.3. Optional 4-20 mA Current Output An option to the standard 0 - 5 VDC out put is the 4 - 20 mA current output that is proportional to flow. The 4 - 20 mA signal is produced fro m the 0 - 5 VDC outp ut of the flow meter. The current loop output is useful for remote applications where pi ...

  • Teledyne HFC-303/307 - page 7

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 7 of 35 2. Operation This section contains the neces sary steps to assist in getting a new flow meter/controller into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install the instrument. 2.1. Receiving Inspection Carefully unpack the Hasti ...

  • Teledyne HFC-303/307 - page 8

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 8 of 35 sulfur hexafluoride) or at high pr essures (> 2 50 psig) the instrument must be installed horizontally. When mounted in a different orientation the instru ment should be re-zeroed at zero flow with the system pressurized to the expected operating p ressure. One of the smallest of the i ...

  • Teledyne HFC-303/307 - page 9

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 9 of 35 Figure 2-1 Table 2-1 “H” Pin-out Pin # 1 Do not use 2 Do not use 3 Do not use 4 Do not use 5 Signal Comm on 6 Output 0 -5 VDC (4-20m A) 7 Case Ground 8 Valve Override 9 -15 VDC 10 Do not use 11 +15 VDC 12 Signal Comm on 13 External Input 14 Set Point 0-5 VDC (4-20m A) 15 Do not use ...

  • Teledyne HFC-303/307 - page 10

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 10 of 35 Figure 2-2 Table 2-2 “U” Pin-out Pin # 1 Signal Comm on 2 Do not use 3 Do not use 4 +15 VDC 5 6 Output 0 -5 VDC (4-20m A) 7 Signal Comm on 8 Case Ground 9 Valve Override 10 11 -15 VDC 12 External In put 13 Signal Com mon 14 Signal Com mon 15 Set Point 0- 5 VDC ( 4-20mA) ...

  • Teledyne HFC-303/307 - page 11

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 11 of 35 2.6. 4 – 20 mA Connections The 300 Series flow meters have a 4 - 20 mA current output option available as an alternative to the standard 0 - 5 VDC output. This current output is useful for remote applicat ions where noise pick-up could substantially affect t he stability of the voltage ...

  • Teledyne HFC-303/307 - page 12

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 12 of 35 2.7. Operation The standard instrument output is a 0 - 5 VDC and the sign al is propor tional to the flow i.e., 0 volts = zero flow and 5 volts = 100% of rated flow. The 4 - 20 mA option is also prop ortional to flow, 4 m A = zero flow and 20 mA = 100 % of rated flow. 2.7.1. Operating Co ...

  • Teledyne HFC-303/307 - page 13

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 13 of 35 S pa n E rror - v s - P res sure f o r 0 . 0 1 4 " S enso rs -3 % -2 % -1 % 0% 1% 2% 3% 0 100 200 300 400 500 600 700 800 900 1000 P ressu re ( p si g ) Er ror s hif t ( % r ea d ing ) Me a n Ma x Mi n Figure 2-4 Sp an E rro r -v s- P ressu re fo r 0. 017" S en so rs -2 % -1 % ...

  • Teledyne HFC-303/307 - page 14

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 14 of 35 S pan E rror v s P re ss ure for 0. 02 6" S e nsor - 20.0% - 18.0% - 16.0% - 14.0% - 12.0% - 10.0% -8 . 0% -6 . 0% -4 . 0% -2 . 0% 0.0% 2.0% 0 100 200 300 400 500 600 700 800 900 1000 P ressure( psi g ) Sp an E r ro r ( % reading ) Figure 2-6 ...

  • Teledyne HFC-303/307 - page 15

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 15 of 35 Span Error v s Pre ssure -1.0 % -0.8 % -0.6 % -0.4 % -0.2 % 0. 0% 0. 2% 0. 4% 0. 6% 0. 8% 1. 0% 0 50 100 150 200 250 300 Pressure( psig) S pan Err or ( % r eadin g) M e a n erro r ma x mi n Figure 2-7 2.7.4. Blending of Gases In the blending of two gases, it is possible to mainta in a fi ...

  • Teledyne HFC-303/307 - page 16

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 16 of 35 2.8. Range Changing Changing the range of a flow controller can be done in the field, but calibration is required. It is recommended however, that the unit be sent back to the factory along with the new range desired, gas and operating parameters . Consult factory for more information. 2 ...

  • Teledyne HFC-303/307 - page 17

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 17 of 35 of the flow meter signal to within ± 0.001 V olts of the command signal. The command signal will not match the flow signal if th ere is insufficient gas pres sure to generate the desired flow. If the comm and signal exceeds 5 Volts the controller will continue to incre ase the flow unti ...

  • Teledyne HFC-303/307 - page 18

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 18 of 35 2.14. Temperature Coefficients As the temperature of the instrument changes fr om the calibration temper ature, errors will be introduced into the output of the instrument. The Te mperature Coefficient of Zero describes the change in the output that is seen at zero flow. This error is ad ...

  • Teledyne HFC-303/307 - page 19

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 19 of 35 3. Theory and Function 3.1. Overall Functional Description The 300 Series flow meters consist of a sensor, base, shunt, control valve (303/307) and electronic circuitry. The sensor is config ured to measure gas flow rate fr om 25-10000 slm, dependin g on the customer’s desired overall ...

  • Teledyne HFC-303/307 - page 20

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 20 of 35 ( ) PP m C T T ud p h e a t e r a m b i e n t −= − • 2 conducted from the upstr eam and downstrea m coils to the foam insulation surroundin g them is assumed to be equal, based on the symmet ry of the sensor construction. Since the sensor tube inlet and outlet are linked by an alum ...

  • Teledyne HFC-303/307 - page 21

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 21 of 35 1.75"(44.5 mm) diameter. The base and endcap seal is an O-ring gland config uration, which uses Viton as a standard sealing mate rial. The HFM-305/ 306 is made from a larger 3.0” b y 3.0” (76.2 mm by 76.2 mm) cross section of 316 stainless steel and has a le ngth of 5.3” (134. ...

  • Teledyne HFC-303/307 - page 22

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 22 of 35 () Δ P LQ D Q D KK ce =+ + 128 8 4 2 24 μ π ρ π The end effects for a typical laminar fl ow element in air account for approxim ately 4% of the total pressure drop. For hydrogen, howeve r, which has a density that i s about 14 times less than air and has a viscosity that is much gre ...

  • Teledyne HFC-303/307 - page 23

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 23 of 35 3.7. Control Valve The control valve is an “automatic metering solenoid” valve (see Figure 3-3). While most solenoid valves operate in either the fully open or closed positions, the automatic metering solen oid valve is designed to control flow. A spring is used to hold a magnetic pl ...

  • Teledyne HFC-303/307 - page 24

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 24 of 35 3.8. Electronic Circuitry The 300 Series flow meters employs a thermal tr ansfer principle (capillary tube described in sect ion 3.2) to measure the flow through the sensor which is p rop ortional to the total flow through the instrument. The sensor develops a differential v oltage outpu ...

  • Teledyne HFC-303/307 - page 25

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 25 of 35 4. Maintenance 4.1. Authorized Maintenance With proper care in installation and use , the flow meter will require little or no mainten ance. If maintenance does become necessary, most of the inst rument can be cleaned or repaired in the field. Some procedures may require reca libration. ...

  • Teledyne HFC-303/307 - page 26

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 26 of 35 Symptom: Sensor has proper resistan ce readings , but little or no output with flow. Cause: Plugged sensor. Action: Shut off gas supply and disconnect the powe r to the flow meter. Remove cover and PC board from unit. Remov e and inspect sensor. If sensor has ev idence of cloggi ng, clea ...

  • Teledyne HFC-303/307 - page 27

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 27 of 35 4.4. End Cap Removal The end cap on the inlet side must be removed to ga in access to shunt assemb ly. First remove power and shut off the supply of gas t o the instrument. Disconnect the fittings on the inlet and outlet sides of the transducer and remove it from the system plum bing. On ...

  • Teledyne HFC-303/307 - page 28

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 28 of 35 4.5. Printed Circuit Boar d and Sensor Replacement NOTE : This instrument contains static sensitive PC boards. Maintain stat ic protection when handling t he PC boards. In the event that any of the PC boards fail, they are easily removed from the instrument and replaced with a spare. It ...

  • Teledyne HFC-303/307 - page 29

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 29 of 35 5. Outline Drawings 5.1. HFM-301 5.2. HFC-303 ...

  • Teledyne HFC-303/307 - page 30

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 30 of 35 5.3. HFM-305 5.4. HFM-306 ...

  • Teledyne HFC-303/307 - page 31

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 31 of 35 5.5. HFC-307 ...

  • Teledyne HFC-303/307 - page 32

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 32 of 35 6. W ARRANTY 6.1. Warranty Repair Policy Hastings Instruments wa rrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship. This war ran ty does not apply to defects or failures resulting from unauthorized modif ication, mi ...

  • Teledyne HFC-303/307 - page 33

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 33 of 35 7. Appendix 7.1. Appendix 1- Volume tric versus Mass Flow Mass flow measures just w hat it says, the mass or number of mol ecules of the gas flowing through the instrument. Mass flow (or weight per u nit time) uni ts are given in pounds per hour (lb/hour), kilograms per sec (kg/sec) etc. ...

  • Teledyne HFC-303/307 - page 34

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 34 of 35 7.2. Appendix 2 - G as Conversion Factors Gas conversion fact ors (GCF’s) for gasse s metered using Hastings I nstruments product s, can be found by visiting the Hastings Instru ments web site. The web address can be found at the end of this document. The gas conversion factors (GCF&ap ...

  • Teledyne HFC-303/307 - page 35

    Manual: 152-082010 3 01-3 05-306_303-307 Series Page 35 of 35 20 / 1.4047 = 14.238 That is, you set the meter to read 14.238 sccm. Some meters, specifically the high flow meters, are calibrated in air. The flow readings must be corrected for the case where a gas other than air is flowing throu gh the meter. In addition, there must be a correction f ...

Производитель Teledyne Категория Automobile Parts

Документы, которые мы получаем от производителя устройства Teledyne HFC-303/307 мы можем разделить на несколько групп. Это в частности:
- технические чертежи Teledyne
- инструкции обслуживания HFC-303/307
- паспорта изделия Teledyne
- информационные брошюры
- энергетические этикетки Teledyne HFC-303/307
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Teledyne HFC-303/307.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Teledyne HFC-303/307, инструкции обслуживания, короткие инструкции или инструкции пользователя Teledyne HFC-303/307. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Teledyne HFC-303/307.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Teledyne HFC-303/307, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Teledyne HFC-303/307 пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Teledyne HFC-303/307.

Полная инструкция обслуживания Teledyne, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Teledyne HFC-303/307 - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Teledyne HFC-303/307, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Teledyne HFC-303/307, которое найдем в данном документе
3. Советы по использованию основных функций устройства Teledyne HFC-303/307 - которые должны облегчить нам первые шаги во время использования Teledyne HFC-303/307
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Teledyne HFC-303/307
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Teledyne HFC-303/307 в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Teledyne HFC-303/307?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Teledyne HFC-303/307, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Teledyne HFC-303/307 со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)