    www.rosemount.com ¢00825-0100-4101¤ Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 Sta r t End Step 1: Mount the T ransmitter Step 2: Consider Housing Rotation Step 3: Set the Jumpers Step 4: Connect the Wiring and Power Step 5: V erify Configuration Step 6: T rim the T ransmitt er Safety Instrumented Systems Product C ...

  • Emerson Process Management 2051 - page 2

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 2 © 2010 Rosemo unt Inc. All rights re served. All marks property o f owner. Rosemount and the Rosemount logotype are registered trad emarks of Rosemount Inc. Rosemount Inc. 8200 Market B oulevard Chanhassen, M N USA 55317 T (US) (800 ) 9 99 -9307 T (Intnl) (952) 906-8888 F ...

  • Emerson Process Management 2051 - page 3

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 3 S TE P 1: M OUNT THE T RANSMITTER A. Applications Liquid Flow Application s 1. Place taps to the side of the line. 2. Mount beside or below the taps. 3. Mount the transmitter so that the drain/vent valves are oriented upward. Gas Flow Applicatio ns 1. Place taps in the top ...

  • Emerson Process Management 2051 - page 4

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 4 S TE P 1 CONTI NUED ... B. Optional Mounting Brackets When installing the transmitter to one of the optional mounting brackets, torque the bracket bolts to 125 in.-lbs. (0,9 N-m). Rosemount 2051C Panel Mount (1) (1) Panel bolts are customer supplied. Pipe Mount Coplanar Fla ...

  • Emerson Process Management 2051 - page 5

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 5 S TE P 1 CONTI NUED ... C. Bolting Considerations If the transmitter installation requires assembly of the process flanges , manifolds, or flange adapters, follow these assembly guidelines to ensure a tight seal for optimal performance characteristics of the transmitte rs. ...

  • Emerson Process Management 2051 - page 6

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 6 S TE P 1 CONTI NUED ... Figure 2. T orque values for the fl ange an d f lange adapter bolts D. O-rings with Flange Adapters W ARNING Whenever the flanges or adapters are removed, vi su ally inspect the o-rings. Replace them if there are any signs of damage, such as nicks or ...

  • Emerson Process Management 2051 - page 7

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 7 E. Inline Gage T ransmitte r Orientat ion The low side pressure port (atmospheric referenc e) on the inline gage transmitter is located in the neck of the transmitter , behind the housing. T he vent path is 360° around the transmitter between the housing and sensor . (See ...

  • Emerson Process Management 2051 - page 8

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 8 Figure 4. 2051 T ransmitter Electronics Board 4-20 mA HART Without LCD Meter With LCD Display 1-5 Vdc HART Low Power Without LCD Meter With LCD Display Security Alarm Alarm Security 4101_QIG_RevEA.fm Page 8 Tuesd a y, June 29, 2010 9:26 AM ...

  • Emerson Process Management 2051 - page 9

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 9 S TE P 4: C ONNECT THE W IRING AND P OWER Use the following steps to wire the transmitter: 1. Remove the housing cover on the FIELD TERMINALS side. 2. Connect the positive lead to the “+” termi nal (PWR/COMM) and the negative lead to the “–” terminal. Figure 5. 4 ...

  • Emerson Process Management 2051 - page 10

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 10 Figure 7. Wiring NOTE Do not connect the powered signal wiring to the test terminals. Power could damage the test diode in the test connection. Shielded twisted pa ir cable should be used for best results. Use 24 A WG or larg er wire and do not exceed 5,000 feet (1500 mete ...

  • Emerson Process Management 2051 - page 11

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 11 Power Supply for 4-20 mA HART T ransmitter operates on 10.5 - 42.4 Vdc. The dc power supply should provide power with less than two percent ripple. Figure 8. Load Limitati on The total resistance load is the sum of the resistance of the signal leads and the load resistanc ...

  • Emerson Process Management 2051 - page 12

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 12 S TE P 5: V ERIFY C ONFIGURATI ON NOTE: A check (P) indicates the basic configuration parameters. At minimum, these parameters should be verified as part of the configuration and startup procedure. T able 1. Field Communicator Fast Key Se quence Function 4-20 mA HART 1-5 V ...

  • Emerson Process Management 2051 - page 13

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 13 S TE P 6: T RIM THE T RANSMITTE R NOTE T ransmitte rs are shipped fully calibrated per reques t or by the factory default of full scale (span = upper range limit). Zero T rim A zero trim is a single-point adjustment used for compensating mounting posit ion ef fects. When ...

  • Emerson Process Management 2051 - page 14

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 14 S AFETY I NSTRUMENTED S YSTEMS The following section applies to 2051 tr ansmitters used in SIS applicatio ns. NOTE T ransmitter output is not safety-rated during t he following: configuration changes, multidrop, loop test. Alternative means should be used to ensure process ...

  • Emerson Process Management 2051 - page 15

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 15 Operation and Maintenance Proof T est and Ins pection The following proof tests are recommended. Proof test results and corrective actions taken must be documented at www .emersonprocess.com/rosemount/safety/certtechdocumentation.htm in the event that an error is found in ...

  • Emerson Process Management 2051 - page 16

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 16 2051 Safety Failure V alues Safety accuracy: 2% (1) Safety response time- 1.5 sec Product Life 50 years – based on worst case component wear-out mechanisms – not based on wear-out process wetted materials P RODUCT C E RTIFICATIONS Approved Manufacturi ng Locations Emer ...

  • Emerson Process Management 2051 - page 17

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 17 HART Protocol Hazardou s Lo ca ti ons Certifications North American Certifications FM Approvals E5 Explosion-Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition-Proof for Class II, Division 1, Groups E, F , and G . Du st-Ignition-Proof for Cl ass III, Divisio ...

  • Emerson Process Management 2051 - page 18

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 18 N1 A TEX T ype n Certification No. Baseefa08A TEX0130X II 3 G Ex nAnL IIC T4 (–40  T a  +70 °C) U i = 42.4 Vdc max IP66 Special Conditions for Safe Use (X): When the optional transient protection terminal block is installed, the apparatus is not capable of withsta ...

  • Emerson Process Management 2051 - page 19

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 19 IECEx Certifications I7 IECEx Intrinsic Safety Certification No. IECExBAS08.0045X Ex ia IIC T4 (–60  T a  +70 °C) 11 8 0 Special Conditions for Safe Use (X): When the optional transient protection terminal block is installed, the apparatus is not capable of withs ...

  • Emerson Process Management 2051 - page 20

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 20 TIIS Certifications E4 TIIS Flame-Proof Ex d IIC T6 Inmetro Certifications E2 Flame-Proof BR-Ex d IIC T6/T5 I2 Intrinsic Safety BR-Ex ia IIC T4 GOST (Russia) Certifications IM Intrinsic Safety Ex ia IIC T4 EM Flame-Proof Ex d IIC T5/T6 China (NEPSI) Certifications E3 Flame ...

  • Emerson Process Management 2051 - page 21

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 21 Combinations of Certifications S tainless steel certification t ag is provided when optional appr oval is specified. Once a device labeled with multiple approval types is installed, it shou ld not be r einst alled using any other approval types. Permanently mark the appro ...

  • Emerson Process Management 2051 - page 22

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 22 4101_QIG_RevEA.fm Page 22 Tu esday, June 29, 2010 9:26 AM ...

  • Emerson Process Management 2051 - page 23

    Quick Installation Guid e 00825-0100-4101, Rev EA June 2010 Rosemount 2051 23 4101_QIG_RevEA.fm Page 23 Tu esday, June 29, 2010 9:26 AM ...

  • Emerson Process Management 2051 - page 24

    Quick Installation Guide 00825-0100-4101, Rev EA June 2010 Rosemount 2051 24 4101_QIG_RevEA.fm Page 24 Tu esday, June 29, 2010 9:26 AM ...

Производитель Emerson Process Management Категория Carbon Monoxide Alarm

Документы, которые мы получаем от производителя устройства Emerson Process Management 2051 мы можем разделить на несколько групп. Это в частности:
- технические чертежи Emerson Process Management
- инструкции обслуживания 2051
- паспорта изделия Emerson Process Management
- информационные брошюры
- энергетические этикетки Emerson Process Management 2051
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Emerson Process Management 2051.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Emerson Process Management 2051, инструкции обслуживания, короткие инструкции или инструкции пользователя Emerson Process Management 2051. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Emerson Process Management 2051.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Emerson Process Management 2051, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Emerson Process Management 2051 пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Emerson Process Management 2051.

Полная инструкция обслуживания Emerson Process Management, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Emerson Process Management 2051 - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Emerson Process Management 2051, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Emerson Process Management 2051, которое найдем в данном документе
3. Советы по использованию основных функций устройства Emerson Process Management 2051 - которые должны облегчить нам первые шаги во время использования Emerson Process Management 2051
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Emerson Process Management 2051
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Emerson Process Management 2051 в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Emerson Process Management 2051?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Emerson Process Management 2051, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Emerson Process Management 2051 со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)