  • Emerson LTM-300 - page 2

    L TM-300 Magnetostrictive Level T ransmitter Instruction and Maintenance Manual T able of Contents Section 1. T ransmitter Overview Page 1.0 General Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Level T ransmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Emerson LTM-300 - page 3

    The L TM-300 is an elec tronic field instrument, suitable for installation i n h a z a r d o u s a n d n o n - h a z a r d o u s i n d u s t r i a l a r e a s . Te s t i n g a n d certification has been obtained from different agencies for installation in specific areas. This instrument is a two wire , loop powered smart transmitter , designed ...

  • Emerson LTM-300 - page 4

    The L TM -300 is an assembly of two major components: • The S ensor T ube Assembly . This a 5/8” diameter stainless steel probe, sealed on one end, with the magnetostri ctive wavegui de in its center . In addition to the magnetostrictive wa veguide, the tube also houses the optional temperature sensor and the detec tor electronics. The tube is ...

  • Emerson LTM-300 - page 5

    Level/Interface. A second float ma y be added below the first one, and the second output will be calibrated automa tically. The second time interval is measured in the same manner as the first and added to the first to derive the position of the heavier floa t. The same six inch dead zone applies to the measur ement of the second level althoug ...

  • Emerson LTM-300 - page 6

    Ambient T emp. Range -20°C ... +40°C (-4°F ... 104°F) Humidity Limits: SAMA PMC 31.1-5.2 Vibration Limits: SA MA PMC 31.1-5.3 RFI Limits: SAMA PMC 31.1-20 t o 1000 MHz up to 30V/m Material: 316ss standard, optional Hastalloy , Monel, Kynar c oated Operating T emperature: -50 to 302°F (-50 t o 150°C ) Maximum Pressur e: 2000 psig @ 300°F Rang ...

  • Emerson LTM-300 - page 7

    Displays Level Measur ement Display - in inches Level measur ement Display - in centimeters Level Measur ement Display - in millimeters Level Measur ement Display - in feet Level Measur ement Display - in meters Inter face Measurements - same decimal positions as L evel Only if configuration set to L evel+Inter face or All T emperature Measurement ...

  • Emerson LTM-300 - page 8

    6.4 Star t-Up f or Gage -M ounted L TM T ransmitters If the transmitter does not appear to calibrate properly, or has an erratic output, check the deadband of the sensor tube and ensure that the float is within the active region of the probe. The active region of the sensor tube is t ypically mar ked with 20mA and 4mA stickers when the unit is cal ...

  • Emerson LTM-300 - page 9

    All MagT ech produc ts are warran ted against defects in materials and workmanship of one year from date of shipment. The level gage chamber and process connections are guaran teed for the life of the tank or vessel to which it is attached . Floats are guarant eed for two years . MagT ech will repair or r eplace at its discretion those products tha ...

  • Emerson LTM-300 - page 10

    When the HART Communicator is connected to the transmitter and turned on, the first screen that is seen by the user is the root menu screen. If the transmitter is configured as L evel Only or Level plus T emperature, the scr een to the left shows the information displayed to the user . However if the tr ansmitter is configured as L evel plus Int ...

  • Emerson LTM-300 - page 11

    If the user selects Operation Status from the Diagnostics screen, he is able to view all of the status bits indicating the health of the transmitter operation. If the entry is labeled OFF , that means that the entr y is in normal operating conditions. If the entr y is labeled ON, that means that the transmitter diagnostics has detected an abnormal ...

  • Emerson LTM-300 - page 12

    The HART Communicator inf orms the user that the maintenance flags have been r eset when the transmitter has replied to the reset command. The user must acknowledge this information by pressing OK to continue . The user is then notified that the loop may now be returned to automatic contr ol. After the user presses OK the HART Communicator r etur ...

  • Emerson LTM-300 - page 13

    If it is desired to trim the digital level value, he may selec t a single point zero trim or a full level trim that requires the input of both the zer o l e v e l v a l u e a n d s o m e o t h e r l e v e l value preferably near the end of the level measurement tube . If the transmitter is configured with temperature measur ement, he may select a ...

  • Emerson LTM-300 - page 14

    Starting at the Calibration screen, select the Zero L evel T rim procedure . This procedure trims the Level digital v alue measurement at the position that the user desires to be the zer o refer ence position. If the user selects Zero Level T rim from the Calibration scr een, this screen will appear reminding the user that the loop should be remove ...

  • Emerson LTM-300 - page 15

    The HART Communicator now requests that the level floa t be set to the desired level upper trim point. The second trim point can be any other level point but it is recommended that this point be as far away fr om the zero point as possible. When the float is at this position, the transmitter will alway s report the digital value entered in the us ...

  • Emerson LTM-300 - page 16

    The user is then notified that the loop may now be returned to automatic contr ol. After the user presses OK the HART Communicator r eturns to the Calibration screen. The HART Communicator is now at the starting screen for the Zero T emper T rim . F ull T emperature T rim Starting at the Calibration screen, select the F ull Level Trim procedure . ...

  • Emerson LTM-300 - page 17

    The user is then notified that the loop may now be returned to automatic contr ol. After the user presses OK the HART Communicator r eturns to the Calibration screen. The HART Communicator is now at the starting screen for the F ull T emper T rim. Basic Setup Proc edures If a transmitter has been configured to measure both L evel and Inter face , ...

  • Emerson LTM-300 - page 18

    The HART Communicator will display this screen r eminding the user that the loop should be removed fr om automatic control before pr oceeding. When this has been done, the user must press OK to continue . The user is then notified that the loop may now be returned to automatic contr ol. After the user presses OK the HART Communicator r eturns to t ...

  • Emerson LTM-300 - page 19

    The next screen gives the user a choice to set the 4mA or 20mA values or to Exit the procedur e. Here the user will select the 4mA menu position. The HART Communicator r equests the user to set the Lev el (or Inter face) to the position that is to provide the 4mA loop curr ent output. When that has been done, the user must press OK to continue with ...

  • Emerson LTM-300 - page 20

        ? ...

Fabricante Emerson Categoría Laser Level

Los documentos del dispositivo Emerson LTM-300 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Emerson
- manuales de instrucciones LTM-300
- hojas de producto Emerson
- folletos informativos
- o etiquetas energéticas Emerson LTM-300
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Emerson LTM-300.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Emerson LTM-300, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Emerson LTM-300. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Emerson LTM-300.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Emerson LTM-300, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Emerson LTM-300. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Emerson LTM-300.

El manual de instrucciones completo de Emerson debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Emerson LTM-300 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Emerson LTM-300, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Emerson LTM-300 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Emerson LTM-300 - que deben facilitarnos los primeros pasos durante el uso de Emerson LTM-300
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Emerson LTM-300
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Emerson LTM-300 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Emerson LTM-300?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Emerson LTM-300, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Emerson LTM-300 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)