    Honeywell Process Solutions 4905 Style Conductivity Cells Installation and Maintenance Manual 70-82-25-18 Revision 6 ...

  • Honeywell 70-82-25-18 - page 2

    ii 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 2007 by Honeywell Inc. Revision 6 – 12/07 While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a par ...

  • Honeywell 70-82-25-18 - page 3

    12/07 4905 Series Conductivity Cells – Installation and Maintenance iii About This Document Abstract This document is intended to support the installation, operation and m aintenance of the 4905 Series of Conductivity Cells. Revision Notes The following list provides notes concerning all revisions of this document. Rev. ID Date Notes 0 12/96 This ...

  • Honeywell 70-82-25-18 - page 4

    iv 4905 Series Conductivity Cells – Installation and Maintenance 12/07 Contents 1. INTRO DUCTION ................................................................................................... 1 1.1 Overview ........................................................................................................................................ ...

  • Honeywell 70-82-25-18 - page 5

    12/07 4905 Series Conductivity Cells – Installation and Maintenance v Figures Figure 1-1 4905 Type Conductivity Cell Mounted in a 1-1/4” Schedule 80 Tee Using an Adapter Bushing ________________________________________________________________________ 1 Figure 3-1 Typical Conductivity Measuring In stallation _____________________________________ ...

  • Honeywell 70-82-25-18 - page 6

    vi 4905 Series Conductivity Cells – Installation and Maintenance 12/07 ...

  • Honeywell 70-82-25-18 - page 7

    Introduction 12/07 4905 Series Conductivity Cells – Installation and Maintenance 1 1. Introduction 1.1 Overview These cells form the sensing network for industrial an alyzers and transm itters de signed to make continuous measurem ents of electrolyti c conductivity. The cel ls are primaril y suited to m easurements i n effluents of ion-exchangers ...

  • Honeywell 70-82-25-18 - page 8

    Introduction 2 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 The cell constant is selected according to the range of the measuring instrum ent used and the solution measured. In general, a high-const ant cell is used for soluti ons having low electrical resi stance (high conductivity) and a low-consta nt cell is used for so ...

  • Honeywell 70-82-25-18 - page 9

    Specifications 12/07 4905 Series Conductivity Cells – Installation and Maintenance 3 2. Specifications 2.1 Specifications for 04905 Series Parameter Description Cell Constant 04905 Series: 0.01, 0.1, 1.0, 10 and 50 as specified Electrode Material Nickel, Platinum or Monel as specified Maximum Pressure Limit 1724 kPa @ 140°C (250 psig @ 284°F) M ...

  • Honeywell 70-82-25-18 - page 10

    Specifications 4 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 2.2 Specifications for 276127 Flow Chamber Parameter Description Maximum Flow 2 gpm @ 40psig and atmospheric discharge Maximum Pressure 200 psig @ 25°C Maximum Temperature 140°C (284°F) at atmospheric pressure Dimensions 1-1/2” (3.8 cm) octagon x 8-3/4” ( ...

  • Honeywell 70-82-25-18 - page 11

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 5 3. Installation 3.1 Overview The conductivity cell is secured permanentl y to the 1” N.P.T. bushing which i s used for all types of mountings. Al though the physical appearance of the various cel ls is the same (except for length), the cell construction differs a ...

  • Honeywell 70-82-25-18 - page 12

    Installation 6 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 3.4 Flow-Type Mounting The cross-channel or guard-tube hole in the cell m ust always be covered by t he solution and the solution level must be 1-1/2 inches above these holes. When mounting t he cell in a pipe tee such as shown in Figure 3-1, have the solution ent ...

  • Honeywell 70-82-25-18 - page 13

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 7 3.6 Insertion-Type Mounting The cell can be inserted into a 1” N.P.T. threaded op ening, but it is imperative that the tan k or chamber be full under all process conditions. Make certain the li quid head is above the cell location. A vertical insertion (from abov ...

  • Honeywell 70-82-25-18 - page 14

    Installation 8 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 001 0.1 1 10 50 X" Ap pr ox. 5. 4" 5. 4" 5. 4" 5. 8" 6. 9 mm 138 138 138 146 175 0. 940" ( 24mm) Di a 1. 5" (38 mm ) Octagon 1" NPT 1. 687" (42. 8mm ) Cell Cell cab le i s appr ox. 0.2 5 0” (6.4 mm) O.D. max wi th 4 c ...

  • Honeywell 70-82-25-18 - page 15

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 9 Figure 3-5 Mounting Dimensions for 04905 Series w ith Quick Disconnect Option CABLE CONNECTOR WILL PASS THRU A 0.625(15,9) MIN. DIA PIPE OR CONDUIT OPENING QUICK DISCONNECT COAX ELECTRODE CABLE APPROX. 0.270(6,8) DIA. CABLE LENGTHS AVAILABLE: 2, 3, 6, 15 METERS MAX ...

  • Honeywell 70-82-25-18 - page 16

    Electrical Connections 10 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4. Electrical Connections 4.1 Overview The terminal board connections for the various Honeywell m easuring instrum ents are given in the appropriate Fig ures in this section. To avoid the possibility of AC pickup in the cell lead s, separate them from a ...

  • Honeywell 70-82-25-18 - page 17

    Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 11 Direct Ce ll to Analyz er Insta llation Cell to Analyz er thr ough Junction Box NOTE S: 1. FOR PURE W ATER S A M PLES I N NO N-C ONDUCTI V E (PL ASTIC, G LASS, ET C) PIPING, G ROUND THE B LACK CELL E L ECTRODE L EAD NEAR THE CEL L. ALTE RNAT ELY, CO NNEC ...

  • Honeywell 70-82-25-18 - page 18

    Electrical Connections 12 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4.2.2 Model 4905 Series with Integral Cable to APT Series Analyzer/Transmitter 04905 series cells with l eads connected to an APT4000 04905 series cells with l eads connected to an APT2000 Figure 4-3 Model 4905 Series to APT Series Analyzer/Transmitter ...

  • Honeywell 70-82-25-18 - page 19

    Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 13 4.3 Instrument Wiring for 4905 Cells with Quick Disconnect Cable 4.3.1 Wiring Model 4905 with Qu ick Disconnect Cable to UDA2182 Figure 4-4 Wiring Diagram for 4905 Cells w ith Quick Disconnect Cable Connected to UDA2182 Analyzer 4.3.2 Wiring Model 4905 w ...

  • Honeywell 70-82-25-18 - page 20

    Electrical Connections 14 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 COAX YELLOW Note: Ignore black and blue wires. COAX YELLOW Note: Ignore black and blue wires. Figure 4-6 Wiring Diagram for 4905 Cells w ith Quick Disconnect Cable Connected to APT2000 ...

  • Honeywell 70-82-25-18 - page 21

    Maintenance 12/07 4905 Series Conductivity Cells – Installation and Maintenance 15 5. Maintenance 5.1 Introduction If abnormal readings occur, this m ay indicate poor re sponse because the cell is not filled with process solution. Check the cell installation. Note that a grayish dull surface on the cell plastic (normally glassy) can result from e ...

  • Honeywell 70-82-25-18 - page 22

    Maintenance 16 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 5.4 Platinizing the Cell Electrodes Only the electrodes having const ants 10 and 50 must be replat inized if the velvety-black deposit has been rubbed off the electrodes in service or in cleani ng or if platinized electrodes are recom me nded and this black deposi ...

  • Honeywell 70-82-25-18 - page 23

    Replacement Parts and Accessories 12/07 4905 Series Conductivity Cells – Installation and Maintenance 17 6. Replacement Parts and Accessories Description Part Number Flow Cell Housing, PES 276127 Junction Box 31316260 Legacy updates cap 50028816-001 Extension Cables for Se nsors with Quick Disconnect Option 2m (6.56 ft) 3m (9.84 ft) 6m (19.69 ft) ...

  • Honeywell 70-82-25-18 - page 24

    ...

  • Honeywell 70-82-25-18 - page 25

    ...

  • Honeywell 70-82-25-18 - page 26

    Honeywell Process Solutions Honeywell, Inc. 512 Virginia Drive Fort Washington, Pennsylvania 19034 70-82-25-18 1207 Printed in USA http://hpsweb.honey well.com ...

Fabricant Honeywell Catégorie Cell Phone

Les documents que nous recevons du fabricant de l'appareilHoneywell 70-82-25-18 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Honeywell
- manuels d’utilisations 70-82-25-18
- fiches produit Honeywell
- dépliants
- ou étiquettes-énergie Honeywell 70-82-25-18
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Honeywell 70-82-25-18.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Honeywell 70-82-25-18, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Honeywell 70-82-25-18. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Honeywell 70-82-25-18.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Honeywell 70-82-25-18, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Honeywell 70-82-25-18 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Honeywell 70-82-25-18.

Le manuel d’utilisation complet Honeywell, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Honeywell 70-82-25-18 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Honeywell 70-82-25-18, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lHoneywell 70-82-25-18 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Honeywell 70-82-25-18 - qui devraient nous aider dans les premières étapes lors de l'utilisation Honeywell 70-82-25-18
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Honeywell 70-82-25-18
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Honeywell 70-82-25-18 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Honeywell 70-82-25-18?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Honeywell 70-82-25-18, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Honeywell 70-82-25-18 il est probable qu’il a envie de partager la façon de le résoudre.