Manual Honeywell 70-82-25-18

26 pages 0.58 mb
Download

Go to site of 26

Summary
  • Honeywell 70-82-25-18 - page 1

    Honeywell Process Solutions 4905 Style Conductivity Cells Installation and Maintenance Manual 70-82-25-18 Revision 6 ...

  • Honeywell 70-82-25-18 - page 2

    ii 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 2007 by Honeywell Inc. Revision 6 – 12/07 While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a par ...

  • Honeywell 70-82-25-18 - page 3

    12/07 4905 Series Conductivity Cells – Installation and Maintenance iii About This Document Abstract This document is intended to support the installation, operation and m aintenance of the 4905 Series of Conductivity Cells. Revision Notes The following list provides notes concerning all revisions of this document. Rev. ID Date Notes 0 12/96 This ...

  • Honeywell 70-82-25-18 - page 4

    iv 4905 Series Conductivity Cells – Installation and Maintenance 12/07 Contents 1. INTRO DUCTION ................................................................................................... 1 1.1 Overview ........................................................................................................................................ ...

  • Honeywell 70-82-25-18 - page 5

    12/07 4905 Series Conductivity Cells – Installation and Maintenance v Figures Figure 1-1 4905 Type Conductivity Cell Mounted in a 1-1/4” Schedule 80 Tee Using an Adapter Bushing ________________________________________________________________________ 1 Figure 3-1 Typical Conductivity Measuring In stallation _____________________________________ ...

  • Honeywell 70-82-25-18 - page 6

    vi 4905 Series Conductivity Cells – Installation and Maintenance 12/07 ...

  • Honeywell 70-82-25-18 - page 7

    Introduction 12/07 4905 Series Conductivity Cells – Installation and Maintenance 1 1. Introduction 1.1 Overview These cells form the sensing network for industrial an alyzers and transm itters de signed to make continuous measurem ents of electrolyti c conductivity. The cel ls are primaril y suited to m easurements i n effluents of ion-exchangers ...

  • Honeywell 70-82-25-18 - page 8

    Introduction 2 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 The cell constant is selected according to the range of the measuring instrum ent used and the solution measured. In general, a high-const ant cell is used for soluti ons having low electrical resi stance (high conductivity) and a low-consta nt cell is used for so ...

  • Honeywell 70-82-25-18 - page 9

    Specifications 12/07 4905 Series Conductivity Cells – Installation and Maintenance 3 2. Specifications 2.1 Specifications for 04905 Series Parameter Description Cell Constant 04905 Series: 0.01, 0.1, 1.0, 10 and 50 as specified Electrode Material Nickel, Platinum or Monel as specified Maximum Pressure Limit 1724 kPa @ 140°C (250 psig @ 284°F) M ...

  • Honeywell 70-82-25-18 - page 10

    Specifications 4 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 2.2 Specifications for 276127 Flow Chamber Parameter Description Maximum Flow 2 gpm @ 40psig and atmospheric discharge Maximum Pressure 200 psig @ 25°C Maximum Temperature 140°C (284°F) at atmospheric pressure Dimensions 1-1/2” (3.8 cm) octagon x 8-3/4” ( ...

  • Honeywell 70-82-25-18 - page 11

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 5 3. Installation 3.1 Overview The conductivity cell is secured permanentl y to the 1” N.P.T. bushing which i s used for all types of mountings. Al though the physical appearance of the various cel ls is the same (except for length), the cell construction differs a ...

  • Honeywell 70-82-25-18 - page 12

    Installation 6 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 3.4 Flow-Type Mounting The cross-channel or guard-tube hole in the cell m ust always be covered by t he solution and the solution level must be 1-1/2 inches above these holes. When mounting t he cell in a pipe tee such as shown in Figure 3-1, have the solution ent ...

  • Honeywell 70-82-25-18 - page 13

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 7 3.6 Insertion-Type Mounting The cell can be inserted into a 1” N.P.T. threaded op ening, but it is imperative that the tan k or chamber be full under all process conditions. Make certain the li quid head is above the cell location. A vertical insertion (from abov ...

  • Honeywell 70-82-25-18 - page 14

    Installation 8 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 001 0.1 1 10 50 X" Ap pr ox. 5. 4" 5. 4" 5. 4" 5. 8" 6. 9 mm 138 138 138 146 175 0. 940" ( 24mm) Di a 1. 5" (38 mm ) Octagon 1" NPT 1. 687" (42. 8mm ) Cell Cell cab le i s appr ox. 0.2 5 0” (6.4 mm) O.D. max wi th 4 c ...

  • Honeywell 70-82-25-18 - page 15

    Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 9 Figure 3-5 Mounting Dimensions for 04905 Series w ith Quick Disconnect Option CABLE CONNECTOR WILL PASS THRU A 0.625(15,9) MIN. DIA PIPE OR CONDUIT OPENING QUICK DISCONNECT COAX ELECTRODE CABLE APPROX. 0.270(6,8) DIA. CABLE LENGTHS AVAILABLE: 2, 3, 6, 15 METERS MAX ...

  • Honeywell 70-82-25-18 - page 16

    Electrical Connections 10 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4. Electrical Connections 4.1 Overview The terminal board connections for the various Honeywell m easuring instrum ents are given in the appropriate Fig ures in this section. To avoid the possibility of AC pickup in the cell lead s, separate them from a ...

  • Honeywell 70-82-25-18 - page 17

    Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 11 Direct Ce ll to Analyz er Insta llation Cell to Analyz er thr ough Junction Box NOTE S: 1. FOR PURE W ATER S A M PLES I N NO N-C ONDUCTI V E (PL ASTIC, G LASS, ET C) PIPING, G ROUND THE B LACK CELL E L ECTRODE L EAD NEAR THE CEL L. ALTE RNAT ELY, CO NNEC ...

  • Honeywell 70-82-25-18 - page 18

    Electrical Connections 12 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4.2.2 Model 4905 Series with Integral Cable to APT Series Analyzer/Transmitter 04905 series cells with l eads connected to an APT4000 04905 series cells with l eads connected to an APT2000 Figure 4-3 Model 4905 Series to APT Series Analyzer/Transmitter ...

  • Honeywell 70-82-25-18 - page 19

    Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 13 4.3 Instrument Wiring for 4905 Cells with Quick Disconnect Cable 4.3.1 Wiring Model 4905 with Qu ick Disconnect Cable to UDA2182 Figure 4-4 Wiring Diagram for 4905 Cells w ith Quick Disconnect Cable Connected to UDA2182 Analyzer 4.3.2 Wiring Model 4905 w ...

  • Honeywell 70-82-25-18 - page 20

    Electrical Connections 14 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 COAX YELLOW Note: Ignore black and blue wires. COAX YELLOW Note: Ignore black and blue wires. Figure 4-6 Wiring Diagram for 4905 Cells w ith Quick Disconnect Cable Connected to APT2000 ...

  • Honeywell 70-82-25-18 - page 21

    Maintenance 12/07 4905 Series Conductivity Cells – Installation and Maintenance 15 5. Maintenance 5.1 Introduction If abnormal readings occur, this m ay indicate poor re sponse because the cell is not filled with process solution. Check the cell installation. Note that a grayish dull surface on the cell plastic (normally glassy) can result from e ...

  • Honeywell 70-82-25-18 - page 22

    Maintenance 16 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 5.4 Platinizing the Cell Electrodes Only the electrodes having const ants 10 and 50 must be replat inized if the velvety-black deposit has been rubbed off the electrodes in service or in cleani ng or if platinized electrodes are recom me nded and this black deposi ...

  • Honeywell 70-82-25-18 - page 23

    Replacement Parts and Accessories 12/07 4905 Series Conductivity Cells – Installation and Maintenance 17 6. Replacement Parts and Accessories Description Part Number Flow Cell Housing, PES 276127 Junction Box 31316260 Legacy updates cap 50028816-001 Extension Cables for Se nsors with Quick Disconnect Option 2m (6.56 ft) 3m (9.84 ft) 6m (19.69 ft) ...

  • Honeywell 70-82-25-18 - page 24

    ...

  • Honeywell 70-82-25-18 - page 25

    ...

  • Honeywell 70-82-25-18 - page 26

    Honeywell Process Solutions Honeywell, Inc. 512 Virginia Drive Fort Washington, Pennsylvania 19034 70-82-25-18 1207 Printed in USA http://hpsweb.honey well.com ...

Manufacturer Honeywell Category Cell Phone

Documents that we receive from a manufacturer of a Honeywell 70-82-25-18 can be divided into several groups. They are, among others:
- Honeywell technical drawings
- 70-82-25-18 manuals
- Honeywell product data sheets
- information booklets
- or energy labels Honeywell 70-82-25-18
All of them are important, but the most important information from the point of view of use of the device are in the user manual Honeywell 70-82-25-18.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Honeywell 70-82-25-18, service manual, brief instructions and user manuals Honeywell 70-82-25-18. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Honeywell 70-82-25-18.

Similar manuals

A complete manual for the device Honeywell 70-82-25-18, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Honeywell 70-82-25-18 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Honeywell 70-82-25-18.

A complete Honeywell manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Honeywell 70-82-25-18 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Honeywell 70-82-25-18, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Honeywell 70-82-25-18, that we can find in the current document
3. Tips how to use the basic functions of the device Honeywell 70-82-25-18 - which should help us in our first steps of using Honeywell 70-82-25-18
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Honeywell 70-82-25-18
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Honeywell 70-82-25-18 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Honeywell 70-82-25-18?

Use the form below

If you did not solve your problem by using a manual Honeywell 70-82-25-18, ask a question using the form below. If a user had a similar problem with Honeywell 70-82-25-18 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)