Summary
  • Invacare IRC5LXAW - page 1

    Dealer: This manual must be given to the end user. User: Before using this concentrator, re ad this manual and save for future reference. Operator's Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Platinum ™ Oxygen Concentrators Standard, with /SensO 2 ® and HomeFill ® Model IRC5L ...

  • Invacare IRC5LXAW - page 2

    WARNING Platinum ™ 2 Part No 113486 6  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE ...

  • Invacare IRC5LXAW - page 3

    TABLE OF CONTENTS Part No 1134866 3 Platinum ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 5 RECYCLING INFO RMATION ............................. 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Maintenance ......................... ...................................................... ........... 9 Ra ...

  • Invacare IRC5LXAW - page 4

    TABLE OF CONTENTS Platinum ™ 4 Part No 113486 6 Routine Maintenanc e ....................... ................................................. 27 SECTION 7—TROUBLES HOOTING GUIDE ............ 31 LIMITED WARRANTY .................................... 33 ...

  • Invacare IRC5LXAW - page 5

    SPECIAL NOTES Part No 1134866 5 Platinum™ SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for  definit ...

  • Invacare IRC5LXAW - page 6

    RECYCLING INFORMATION Platinum™ 6 Part No 1134866 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  pro duct ? ...

  • Invacare IRC5LXAW - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1134866 7 Platinum™ SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GU IDELINES contains important information for the safe operation and use of this product. In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Platinum  co ...

  • Invacare IRC5LXAW - page 8

    SECTION 1—GENERAL GUIDELINES Platinum™ 8 Part No 1134866  WARNINGS The  use  of  oxygen  therapy  requires  that  special  care  be  take n  to  reduce  the  risk  of  fire.  Any  materials  that  will  burn  in  air ,  and  some  that  will  not,  are  ...

  • Invacare IRC5LXAW - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1134866 9 Platinum™ Fill  humidifier  with  wa t e r  to  the  lev el  shown  by  the  manufacturer .  DO  NOT  overfill. Keep  unit  at  least  7.5  cm  (3  in)  aw ay  from  w alls,  draperies,  furniture,  and  the  like. ...

  • Invacare IRC5LXAW - page 10

    SECTION 1—GENERAL GUIDELINES Platinum™ 10 Part No 1134866 DO  NOT  reach  for  prod uct  that  has  fallen  into  wat e r .  Unplug  IMMEDIA TEL Y . This  device  is  to  be  used  only  in  accordance  with  the  prescription  of  a  physician  and  this  Owner ...

  • Invacare IRC5LXAW - page 11

    SECTION 2—FEATURES Part No 1134866 11 Platinum™ SECTION 2—FEATURES Oxygen Purity and Fault Indicator Lights (SensO 2 model)/Fault and Power Indicator Lights (LX model) Circuit Breaker Power Switch Elapsed Time Meter (Hour Meter) Cabinet Filter (one of two shown) Flowmet er Oxygen Out let FRONT VIEW TOP VIEW ACCESSOR IES (NOT SHOWN): HomeFill ...

  • Invacare IRC5LXAW - page 12

    SECTION 3—SHIPPING AND HANDLING Platinum™ 12 Part No 1134866 SECTION 3—SHIPPING AND HANDLING NOTE:  For  this  procedure,  refe r  to  FIGURE  3.1  on  page 13. Before  yo u  install  and  operate  the  concentrator ,  please  refer  to  Platinum  Series  Owner ʹ s  Manual. ...

  • Invacare IRC5LXAW - page 13

    SECTION 3—SHIPPING AND HANDLING Part No 1134866 13 Platinum™ FIGURE 3.1 Unpacking ...

  • Invacare IRC5LXAW - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 14 Part No 1134866 SECTION 4—TYPICAL PRODUCT PARAMETERS Regulatory Listing: EN/IEC 60601-1 Alternating Current Type BF equipment Unit running Unit not running Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulation Electrical Requirements: 230 VAC ± 10% (253 VAC/207 ...

  • Invacare IRC5LXAW - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 15 Platinum™ *Oxygen Output Concentration Levels: All 5LXO2AWQ/5LXAW/ 5LXO2AW/9LXO2AWQ models * (Stated concentration levels achieved after in itial warm-up period (approximately 30 minutes). IRC5LXAW, IRC5 LXO2AW and IRC5LXO2AWQ 93% minimu m at 1 to 3 L/min. 91% minimu m at 4 L/min. 87% mini ...

  • Invacare IRC5LXAW - page 16

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 16 Part No 1134866 Change in maximum recommended flow when back pressure of 7k Pa is applied: 0.7 L/min. Filters: Cabinet (2), Outl et HEPA and Compressor Inlet. Safety System: Current overload or line surge shutdown. High temperature co mpressor shutdown. High Pressure Alarm w/compressor shutdown. ...

  • Invacare IRC5LXAW - page 17

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 17 Platinum™ Regulatory Listing : IRC5LXAW/IR C5LXO2AW/ IRC5LXO2AWQ/IRC9LXO2AWQ ETL tested as complying with: EN55011 EN61000-3-2 EN61000-3-3 IEC 60601-1, A1, A2 IEC 60601-1-2 IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ Models Only ISO8359 MDD CE marked models: IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ E ...

  • Invacare IRC5LXAW - page 18

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 18 Part No 1134866 SENSO 2 INDICATO RS LX INDICATORS LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied technician. SYSTEM ...

  • Invacare IRC5LXAW - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 19 Platinum™ SECTION 5—OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  in  the  home.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ...

  • Invacare IRC5LXAW - page 20

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 20 Part No 1134866 Recommended Guidelines for Optimum Performance NOTE:  Ensure  that  your  concentrator  is  at  least  7.5  cm  (3  in)  aw ay  from  walls,  drap erie s  or  furniture  to  assur e  sufficient  air  flow .  Av o i ...

  • Invacare IRC5LXAW - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 21 Platinum™ Connect Humidifier (If So Prescribed) NOTE:  For  this  procedure,  refe r  to  FIGURE  5.1,FIGURE  5.2,  FIGURE  5.3  on  page 22  and  FIGURE  5.4  on  page 23. 1. Fill  humidifier  with  distilled  wa t e r  t ...

  • Invacare IRC5LXAW - page 22

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 22 Part No 1134866 3. Remove  the  filter  access  door  located  on  the  side  of  the  concentrator .  Insert  a  flathead  screwdriver  in  the  plate  groove  on  th e  top  edg e  of  the  filter  access  door  ...

  • Invacare IRC5LXAW - page 23

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 23 Platinum™ 6. Attach  oxygen  tubing  from  the  humidifier  bottle  to  oxygen  outlet  connector  on  the  oxygen  concentrator . 7. Attach  the  cannula/patient  supply  tubing  to  the  humidifier  bottle  outlet. 8. After ...

  • Invacare IRC5LXAW - page 24

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 24 Part No 1134866 FIGURE 5.5 Power Switch Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.6  on  page 25. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  you r  physician  or  therapist.  WARNING DO NOT ...

  • Invacare IRC5LXAW - page 25

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 25 Platinum™ NOTE:  The  use  of  some  accessories  such  as  the  pediatric  flowstand  and  the  HomeFill  compr essor  will  deactivate  the  Low  Flow  Alarm. FIGURE 5.6 Flowrate SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model ...

  • Invacare IRC5LXAW - page 26

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 26 Part No 1134866 FIGURE 5.7 SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model) GREEN  light  (O 2 ) ‐ Normal  Operation. YELLOW  light  () ‐ Call  supplier  IMMEDIA TEL Y .  Yo u  may  continue  to  use  the  concentrator  unless  in ...

  • Invacare IRC5LXAW - page 27

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 27 Platinum™ FIGURE 5.8 Indicators (LX Model) Main Power Loss Alarm System The  alarm  system  contains  no  battery ‐ and  is  thereby  maintenance  free.  The  alarm  system  is  powered  by  a  capacitor  that  is  continuousl ...

  • Invacare IRC5LXAW - page 28

    SECTION 6—MAINTENANCE Platinum™ 28 Part No 1134866 SECTION 6—MAINTENANCE NOTE:  The  Invacare  concentrators  are  specifically  designed  to  minimize  rou ti n e  pr eventive  maintenance.  In  places  with  high  dust  or  soot  levels ,  maintenance  may  need  to  be ...

  • Invacare IRC5LXAW - page 29

    SECTION 6—MAINTENANCE Part No 1134866 29 Platinum™ 3. Dry  the  filters  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter Cabinet  WARNING Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean  the  cabinet  with  a  mild  household  cle ...

  • Invacare IRC5LXAW - page 30

    SECTION 6—MAINTENANCE Platinum™ 30 Part No 1134866 FIGURE 6.2 Preventive Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS FOR IR ...

  • Invacare IRC5LXAW - page 31

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1134866 31 Platinum™ SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord NOT plugged in. 1. Insert plug i nto outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit breakers or fus ...

  • Invacare IRC5LXAW - page 32

    SECTION 7—TROUBLESHOOTING GUIDE Platinum™ 32 Part No 1134866 YELLOW or RED light illumin ated. Alarm: Continuous On red light only. * Only applies t o SensO 2 models. 1. Low oxygen p urity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min. or less.* 4. Unit overheating due to blocked air intake. 5. Insufficien ...

  • Invacare IRC5LXAW - page 33

    LIMITED WARRANTY Part No 1134866 33 Platinum™ LIMITED WARRANTY A vailable  At: Belgium  &  Luxemburg: Inv acare  nv ,  Autobaan  22,  B ‐ 8210  Loppem Te l :  (32)  (0)50  83  10  10,  Fax:  (32)  (0)50  83  10  11 belgium@inv acare.com Danmark:  Inv acare  A/S,  Sdr .  Ri ...

  • Invacare IRC5LXAW - page 34

    LIMITED WARRANTY Platinum™ 34 Part No 1134866 Italia: Inv acare  Mecc  San  s.r .l.,  Vi a  dei  Pini  62,  I ‐ 36016  Thiene  (VI) Te l :  (39)  0445  38  00  59,  Fax:  (39)  0445  38  00  34 italia@inv acare.com  Nederland: Inv acare  BV ,  Celsiusstraat  46,  NL ? ...

  • Invacare IRC5LXAW - page 35

    LIMITED WARRANTY Part No 1134866 35 Platinum™ Switzerland: Mobitec  Rehab  AG,  Benkenstrasse  260,  CH ‐ 4108  Wit t e rs w il Te l . :  (41)  (0)61  487  70  80,  Fax.:  (41)  (0)61  487  70  81 office@mobitec ‐ rehab.ch switzerland@inv acare.com United  Kingdom: Inv acare  Ltd,  S ...

  • Invacare IRC5LXAW - page 36

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Serv ices 440-329-6593 800-832-4707 Manufacturer: Invacare Medical Equipment (Suzhou) Co., Ltd. Unit 2B, Export Processing Zone No. 200 Suhong Road SIP (Suzhou Industrial Park) Suzhou, Jiangsu, PRC 215021 Tel: 86-512-62586192 F ...

メーカー Invacare カテゴリー Oxygen Equipment

Invacare IRC5LXAWのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- IRC5LXAWの取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare IRC5LXAWの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare IRC5LXAWの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare IRC5LXAWの取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare IRC5LXAWのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare IRC5LXAWの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare IRC5LXAWデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare IRC5LXAWを使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare IRC5LXAWユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare IRC5LXAWの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare IRC5LXAWの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare IRC5LXAWに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare IRC5LXAWデバイスの基本機能を使うにあたってのヒント。 Invacare IRC5LXAWのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare IRC5LXAWに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare IRC5LXAWのメーカー/サービスへの連絡先に関する情報。