    IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-320 MODEL: 61-322 MODEL: 61-324 The Service Information provid es the following in formation: • Precautions and safety information • Specificatio ns • Performance test procedure • Calibration and cali bration adjustm ent procedure • Basic maintenance (cleaning, replacing the battery and fu ...

  • IDEAL INDUSTRIES 61-320 - page 2

    TABLE OF CONTENTS Page # Introduction 1 Precautions an d Safety Inf ormation 1 Symbols 1 Safety 2 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 5 Resistance, Diode, Continuity Specifications 5 Capacitance , Frequency S pecifications 6 Physical and environm ent characteristics 7 Certification and compliance ...

  • IDEAL INDUSTRIES 61-320 - page 3

    Page 1 Introduction Warning To avoid shoc k or injury , do not per form the verification tests or cal ibration pr ocedures described in this manual unless you are qualified to do so. The information provided in th is document is for the use of qualified personnel only. Caution The 61-320, 61-322, and 61-3 24 contain part s that can be damaged by st ...

  • IDEAL INDUSTRIES 61-320 - page 4

    Page 2 SAFETY Review the fol lowing safety preca utions to avoid inj ury and pre vent dam age to this product or any product s connected to it. To avoid potential h azards, use the product only as specified. CAUTION. These statements iden tify conditions or practices that could result in d amage to the e quipm ent or other pr operty. WARNING. These ...

  • IDEAL INDUSTRIES 61-320 - page 5

    Page 3 SPECIFICATIONS All specificatio ns are warrante d unless n oted typi cal and appl y to the 61-32 0, 61-322 an d 61-324. Stated accuracies are at 23 ºC ± 5 ºC at less than 80% relative humidity and without the ba ttery indicator di splayed. General specifications Characteristic s Description Display c ount 6,000 Numeric updat e rate 1.5 ti ...

  • IDEAL INDUSTRIES 61-320 - page 6

    Page 4 Measurement Characteristics Accuracy is ± (% reading + number of digits) at 23 ° C ± 5 ° C , less than 80% R.H. DC Volts Range Resolution Accuracy Over voltage protection 600.0m V 100µV 6.000V 1mV 60.00V 10mV 600.0V 100mV 1000V 1V ±(0.5% + 2dgt) 1000V rms AC Volts Range Resolution Accuracy Over voltage protection 600.0m V 100µV Unspec ...

  • IDEAL INDUSTRIES 61-320 - page 7

    Page 5 DC Current Range Resolution Accuracy Over voltage protection 600.0µA 1m V 6000µA 1mV 60.00m A 10mV 600.0m A 100mV ±(1.0%+2 dgt ) 600V rms1A (600) Fast blow fuse AC Current Range Resolution Accuracy Over voltage protection 600.0µA 1m V 6000µA 1mV 60.00m A 10mV 600.0m A 100mV ±(1.5%+2 dgt ) 50Hz ~ 50 0Hz * 600V rms 1A (600) Fa st blow fu ...

  • IDEAL INDUSTRIES 61-320 - page 8

    Page 6 Frequency Range Sensitivity Accuracy Overload protection 6000Hz 60.00KHz 600.0KHz 100mV rm s * 6.000MHz 25 0mV rms 60.00MHz 1V rms Frequency : 0.01%±1di git 600V rms * Less t han 20Hz, the sensitivity is 1.5V rms. Capacitance Range Accura cy Overload protection 6.000nF 60.00nF 600.0nF 6.000µF 60.00µF 600.0µF 6.000m F * ±(1.9% + 8 dgt) 6 ...

  • IDEAL INDUSTRIES 61-320 - page 9

    Page 7 Physical and E nvironmental C haracteristi cs Characteristic s Description Dime nsions (H x W x D) 158mm x 76m m x 38mm 164mm x 82mm x 44mm (with holst er) Weight (with battery) 0.3Kg With holster 0.5Kg Environmental Characteristics Descripti on Temperature operating 0 to + 5 0ºC Non-Operating -20 to + 60 º C Humidity (operating) < 80% ...

  • IDEAL INDUSTRIES 61-320 - page 10

    Page 8 Required Equipment Required equipment is listed in Table B. I f the recom mended m odels are not avail able, equi pment wi th equivale nt specifications may be used. Repai rs or ser vicing sh ould be performed o nly by q ualified per sonnel. Table B. Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltag ...

  • IDEAL INDUSTRIES 61-320 - page 11

    Page 9 Basic Maintenance Warning To avoid shoc k, rem ove the test leads and an y input signals before opening the case or re placing the battery or fuses. Opening the Meter Case Caution To avoid unintentional short circ uit, always place the uncovered m eter assembly on a protective surface . When the case of the m eter is open, circuit connection ...

  • IDEAL INDUSTRIES 61-320 - page 12

    Page 10 Testing Fuses To test the internal fuses of the meter: 1. Turn the rotary selector switch to the Ω positio n. 2. To test FS1, plug a test lead into V Ω Hz input terminal and touch the probe to th e mA input terminal. The display sh ould indicate bet ween 0.0 to 0.2 Ω . FS1 (1 A 600V - Bussmann BBS-1 recomm ended). If display reads hig ...

  • IDEAL INDUSTRIES 61-320 - page 13

    Page 11 Performance Tests The following performance tests ve rify the complete ope ra bility of the meter and c heck the accuracy of each meter function against the meter’s specifications. Accuracy specifications are valid for a period of one y ear after calibration, when measured at an operating temp er atu r e of 1 8ºC to 28ºC and at a maximu ...

  • IDEAL INDUSTRIES 61-320 - page 14

    Page 12 Table 1 AC Voltage Test: Step Input Frequency Reading 1 5.800V 50Hz 5.743 to 5.857 2 5.800V 500Hz 5.743 to 5.857 3 58.00V 50Hz 57.43 to 58.57 4 58.00V 500Hz 57.43 to 58.57 5 580.0V 50Hz 574.3 to 585.7 6 580.0V 500Hz 574.3 to 585.7 7 750V 50Hz 738 to 762 8 750V 500Hz 738 to 762 6. Turn the rotary switch to “V ” position. 7. Set the calib ...

  • IDEAL INDUSTRIES 61-320 - page 15

    Page 13 Testing the Capacitance Function The meter measures capacitance by c h arging the capac itor with a known direct current, measuring the resultant voltage, and calcu lating the capacitance. If the same capacitan ce is m easured on an impedance bridge, a different rea ding may result. This variance is likely to be greater at higher frequencie ...

  • IDEAL INDUSTRIES 61-320 - page 16

    Page 14 Table 5 AC mA Test: Step Source Frequency Reading 1 58.00mA 50Hz 56.93 to 59.07 2 58.00mA 500Hz 56.93 to 59.07 3 580.0m A 50Hz 569.3 to 590.7 4 580.0m A 500Hz 569.3 to 590. 7 6. Turn the rotary switch to mA . 7. Apply the inputs for steps 1-2 in Table 6. 8. For each input, compare the reading on the meter display to the read ing in Table 6. ...

  • IDEAL INDUSTRIES 61-320 - page 17

    Page 15 Table 8 Frequency Test: Step Source Level Reading 1 5.600KHz 100m V rms 5.598 to 5.602 2 56.00KHz 100m V rms 55.98 to 56 .02 3 560.0KHz 100m V rms 559.8 to 56.02 4 5.600M Hz 250mV rm s 5.598 to 5.602 5 56.00M Hz 1V rms 55.98 to 56.02 Calibration Calibrate the meter once a year to ensure that it performs according to specifications. Per for ...

  • IDEAL INDUSTRIES 61-320 - page 18

    Page 16 Figure 3 - Calibrati on Adjustment Point s Form number TM61320-2-4 Rev 2 September 2002 ...

  • IDEAL INDUSTRIES 61-320 - page 19

    Form number TM61320-2-4 Rev 2 September 2002 Page 17 ...

メーカー IDEAL INDUSTRIES カテゴリー Automobile Accessories

IDEAL INDUSTRIES 61-320のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 61-320の取扱説明書
- IDEAL INDUSTRIESの製品カード
- パンフレット
- またはIDEAL INDUSTRIES 61-320の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、IDEAL INDUSTRIES 61-320の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、IDEAL INDUSTRIES 61-320の取り付け説明書、サービスマニュアル、簡易説明書、またはIDEAL INDUSTRIES 61-320のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、IDEAL INDUSTRIES 61-320の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

IDEAL INDUSTRIES 61-320デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがIDEAL INDUSTRIES 61-320を使用するのを助ける技術的文書のことです。説明書は通常、全てのIDEAL INDUSTRIES 61-320ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

IDEAL INDUSTRIESの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. IDEAL INDUSTRIES 61-320の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははIDEAL INDUSTRIES 61-320の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。IDEAL INDUSTRIES 61-320に関してこのドキュメントで見つけることができる全てのヒントの目次
3. IDEAL INDUSTRIES 61-320デバイスの基本機能を使うにあたってのヒント。 IDEAL INDUSTRIES 61-320のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。IDEAL INDUSTRIES 61-320に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるIDEAL INDUSTRIES 61-320のメーカー/サービスへの連絡先に関する情報。

IDEAL INDUSTRIES 61-320についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもIDEAL INDUSTRIES 61-320の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがIDEAL INDUSTRIES 61-320で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (1)

PEP GIL

Puedo rechazar llamadas improcedentes con el PANASONIC KX-TG6511SPM?