    NI PXI/PCI-5114 Specifications 8-Bit 250 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5114 (NI 5114) high-speed digitizer. Unless otherw ise noted, these specifications are valid for the following conditions: • All filter settings • All impedance selections • Sample clock set to 250 MS/s T ypical values are repre ...

  • National Instruments NI PCI-5114 - page 2

    NI PXI/PCI-5114 Specifications 2 ni.com Waveform Specifications ........... .................... ...................... ...................13 Calibration .............. .................... .................... .................... ................... ..14 Power .............. .................... ................ .................... ............. ...

  • National Instruments NI PCI-5114 - page 3

    © National Instruments Corporat ion 3 NI PXI/PCI-5 114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Of fset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offset Range (V) 0.04 ±0.8 0.04 ±0.8 0.1 ±0.8 0.1 ±0.8 0.2 ±0.8 0.2 ±0.8 0.4 ±0.8 0.4 ±0.8 1 ±6.5 1.0 ± ...

  • National Instruments NI PCI-5114 - page 4

    NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (T ypical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/F all T ime, T ypical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.5 ns Bandwidth Limit Filter 20 MHz Noise Filter — A C Coupling * Cutof f (–3 dB), T ypical 12 ...

  • National Instruments NI PCI-5114 - page 5

    © National Instruments Corporat ion 5 NI PXI/PCI-5 114 Specifications Specif ication Va l u e Commen ts Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR), T ypical Range (V pk-pk ) 10 MHz, –1 dBFS input signal Includes the 2 nd through the 5 th harmonics Measured from DC to 125 MHz 20 MHz bandwidth limit filter of f All ...

  • National Instruments NI PCI-5114 - page 6

    NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Va l u e Comments Sources Internal, Onboard Clock (i nternal VCXO)* External, CLK IN (front panel SMB con nector) * Internal Sample Clock is locked to the Reference Clock or deri ved from the onboard VCXO Onboard Clock (Internal VCXO) Sample Rate Range Real-T ime Sampling ...

  • National Instruments NI PCI-5114 - page 7

    © National Instruments Corporat ion 7 NI PXI/PCI-5 114 Specifications Phase-Locked Loo p (PLL) Reference Clock External Sample Clock Sources CLK IN (front panel SMB connector) — Frequenc y Range 50 MHz to 250 MHz Di vide by n decimation av ailable where 1 ≤ n ≤ 65,535 For more informati on about Sample Clock and decimation, refer to the NI H ...

  • National Instruments NI PCI-5114 - page 8

    NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) T r igger Reference (Stop) T rigger Specif ication Va l u e Input V oltage Range Sine wa ve: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wa ve: 0.2 V pk-pk to 2.8 V pk-pk Maximum Input Overload 7V rms with | Peaks | ≤ 10 V Imped ...

  • National Instruments NI PCI-5114 - page 9

    © National Instruments Corporat ion 9 NI PXI/PCI-5 114 Specifications Analog Trigger (Edge, Window, and Hysteresis Trigger Types) Sources CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) — T rigger Level Resolution 8 bits (1 in 256) T rigger Level Range CH 0, CH 1 TRIG (External Trigger) — 100% ...

  • National Instruments NI PCI-5114 - page 10

    NI PXI/PCI-5114 Specifications 10 ni.com TRIG (External T rigger , Front Panel Connector) T ypes Specific Line Any Line Specific Field — Standards SDTV : M-NTSC, B/G-P AL, SECAM, M-P AL EDTV : 480i/59.94 fps, 480i /60 fps, 480p/59.94 Fps, 480p/60 Fps, 576i/50 f ps, 576p/50 Fps HDTV : 720p/ 50 Fps, 720p/59.94 Fps, 720p/6 0 Fps, 1080i/50 fps, 1080i ...

  • National Instruments NI PCI-5114 - page 11

    © National Instruments Corporat ion 11 NI PXI/PCI-5 114 Specifications PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) Specif ication Va l u e Connector 9-pin mini-circu lar DIN Direction Bi-directional As an Input (Trigger) Destinations Start T rigger (Acquisitio n Arm) Reference (Stop) Tri gger Arm Reference Trigger ...

  • National Instruments NI PCI-5114 - page 12

    NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis. For more inform ation about TClk synchronizati on, refer to the NI-TClk Synchronization Help , wh ich is located within the N I Hi ...

  • National Instruments NI PCI-5114 - page 13

    © National Instruments Corporat ion 13 NI PXI/PCI-5 114 Specifications W aveform Specifications Specif ication Va l u e Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channel — 64 MB per Channel Option 64 megasamples per channel 256 MB per Channel Option 256 megasamples per channel Minimum Record Length 1 Sample — Num ...

  • National Instruments NI PCI-5114 - page 14

    NI PXI/PCI-5114 Specifications 14 ni.com Calibration Power Specif ication Va l u e Self-Calibrati on Self-calibratio n is done on software command. The calibration corrects for gain, of fset, compensated 1 M Ω attenuator , triggerin g, and timing adjustment errors for all input ranges. External Calibration (Factory Calibration) The external calibr ...

  • National Instruments NI PCI-5114 - page 15

    © National Instruments Corporat ion 15 NI PXI/PCI-5 114 Specifications Software Specification Va l u e Dri ver Software NI PXI-5114 : NI-SCOPE 2.9 or later NI PCI-5114 : NI-SCOPE 3.1 or later NI-SCOPE is a n IVI-compliant dri ver that allo ws you to configure, control, and calibrate the NI 5114. NI-SCOPE provides application programming interfaces ...

  • National Instruments NI PCI-5114 - page 16

    NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note T o ensure that the NI PXI-5114 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only . Specif ication Va l u e Operating T emperature 0 ºC to +55 ºC in all N ...

  • National Instruments NI PCI-5114 - page 17

    © National Instruments Corporat ion 17 NI PXI/PCI-5 114 Specifications NI PCI-5114 Note T o ensure that the NI PCI-5114 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage. T o maximize airflo w and extend the life of the de vice, leav e any adjacent ...

  • National Instruments NI PCI-5114 - page 18

    NI PXI/PCI-5114 Specifications 18 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follow ing standards of safety for electrical e quipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safet ...

  • National Instruments NI PCI-5114 - page 19

    © National Instruments Corporat ion 19 NI PXI/PCI-5 114 Specifications Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er. NI recognizes that eliminating certa in hazardous substances from our products is beneficial not only to t he environment but also to ...

  • National Instruments NI PCI-5114 - page 20

    NI PXI/PCI-5114 Specifications 20 ni.com Dimensions and W eight NI PXI-5114 Front Panel Indicators Label Function Fo r mo re information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5114 to the controller . AC T I VE The ACTIVE LED indicates th e status ...

  • National Instruments NI PCI-5114 - page 21

    © National Instruments Corporat ion 21 NI PXI/PCI-5 114 Specifications NI PCI-5114 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 421 g (14.8 oz) 2.0 cm (0. 8 in.) 3 5.5 cm (14.0 in.) 11. 3 cm (4.4 in.) ...

  • National Instruments NI PCI-5114 - page 22

    NI PXI/PCI-5114 Specifications 22 ni.com Where to Go for Support The National Instruments W eb site is your complete resource for technical support. At ni.com/support you ha ve access to e verything from troubleshooting and ap plication dev elopment self-help resources to email and phone assistance from NI Application Engin eers. A Declaration of C ...

  • National Instruments NI PCI-5114 - page 23

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s. For patents covering Natio ...

メーカー National Instruments カテゴリー Graphics Tablet

National Instruments NI PCI-5114のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- NI PCI-5114の取扱説明書
- National Instrumentsの製品カード
- パンフレット
- またはNational Instruments NI PCI-5114の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、National Instruments NI PCI-5114の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、National Instruments NI PCI-5114の取り付け説明書、サービスマニュアル、簡易説明書、またはNational Instruments NI PCI-5114のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、National Instruments NI PCI-5114の製品を使用するにあたって最も人気のある説明書を閲覧できます。

National Instruments NI PCI-5114デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがNational Instruments NI PCI-5114を使用するのを助ける技術的文書のことです。説明書は通常、全てのNational Instruments NI PCI-5114ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

National Instrumentsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. National Instruments NI PCI-5114の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははNational Instruments NI PCI-5114の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。National Instruments NI PCI-5114に関してこのドキュメントで見つけることができる全てのヒントの目次
3. National Instruments NI PCI-5114デバイスの基本機能を使うにあたってのヒント。 National Instruments NI PCI-5114のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。National Instruments NI PCI-5114に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるNational Instruments NI PCI-5114のメーカー/サービスへの連絡先に関する情報。

National Instruments NI PCI-5114についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもNational Instruments NI PCI-5114の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがNational Instruments NI PCI-5114で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)