  • National Instruments NI PCI-5114 - page 1

    NI PXI/PCI-5114 Specifications 8-Bit 250 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5114 (NI 5114) high-speed digitizer. Unless otherw ise noted, these specifications are valid for the following conditions: • All filter settings • All impedance selections • Sample clock set to 250 MS/s T ypical values are repre ...

  • National Instruments NI PCI-5114 - page 2

    NI PXI/PCI-5114 Specifications 2 ni.com Waveform Specifications ........... .................... ...................... ...................13 Calibration .............. .................... .................... .................... ................... ..14 Power .............. .................... ................ .................... ............. ...

  • National Instruments NI PCI-5114 - page 3

    © National Instruments Corporat ion 3 NI PXI/PCI-5 114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Of fset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offset Range (V) 0.04 ±0.8 0.04 ±0.8 0.1 ±0.8 0.1 ±0.8 0.2 ±0.8 0.2 ±0.8 0.4 ±0.8 0.4 ±0.8 1 ±6.5 1.0 ± ...

  • National Instruments NI PCI-5114 - page 4

    NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (T ypical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/F all T ime, T ypical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.5 ns Bandwidth Limit Filter 20 MHz Noise Filter — A C Coupling * Cutof f (–3 dB), T ypical 12 ...

  • National Instruments NI PCI-5114 - page 5

    © National Instruments Corporat ion 5 NI PXI/PCI-5 114 Specifications Specif ication Va l u e Commen ts Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR), T ypical Range (V pk-pk ) 10 MHz, –1 dBFS input signal Includes the 2 nd through the 5 th harmonics Measured from DC to 125 MHz 20 MHz bandwidth limit filter of f All ...

  • National Instruments NI PCI-5114 - page 6

    NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Va l u e Comments Sources Internal, Onboard Clock (i nternal VCXO)* External, CLK IN (front panel SMB con nector) * Internal Sample Clock is locked to the Reference Clock or deri ved from the onboard VCXO Onboard Clock (Internal VCXO) Sample Rate Range Real-T ime Sampling ...

  • National Instruments NI PCI-5114 - page 7

    © National Instruments Corporat ion 7 NI PXI/PCI-5 114 Specifications Phase-Locked Loo p (PLL) Reference Clock External Sample Clock Sources CLK IN (front panel SMB connector) — Frequenc y Range 50 MHz to 250 MHz Di vide by n decimation av ailable where 1 ≤ n ≤ 65,535 For more informati on about Sample Clock and decimation, refer to the NI H ...

  • National Instruments NI PCI-5114 - page 8

    NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) T r igger Reference (Stop) T rigger Specif ication Va l u e Input V oltage Range Sine wa ve: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wa ve: 0.2 V pk-pk to 2.8 V pk-pk Maximum Input Overload 7V rms with | Peaks | ≤ 10 V Imped ...

  • National Instruments NI PCI-5114 - page 9

    © National Instruments Corporat ion 9 NI PXI/PCI-5 114 Specifications Analog Trigger (Edge, Window, and Hysteresis Trigger Types) Sources CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) — T rigger Level Resolution 8 bits (1 in 256) T rigger Level Range CH 0, CH 1 TRIG (External Trigger) — 100% ...

  • National Instruments NI PCI-5114 - page 10

    NI PXI/PCI-5114 Specifications 10 ni.com TRIG (External T rigger , Front Panel Connector) T ypes Specific Line Any Line Specific Field — Standards SDTV : M-NTSC, B/G-P AL, SECAM, M-P AL EDTV : 480i/59.94 fps, 480i /60 fps, 480p/59.94 Fps, 480p/60 Fps, 576i/50 f ps, 576p/50 Fps HDTV : 720p/ 50 Fps, 720p/59.94 Fps, 720p/6 0 Fps, 1080i/50 fps, 1080i ...

  • National Instruments NI PCI-5114 - page 11

    © National Instruments Corporat ion 11 NI PXI/PCI-5 114 Specifications PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) Specif ication Va l u e Connector 9-pin mini-circu lar DIN Direction Bi-directional As an Input (Trigger) Destinations Start T rigger (Acquisitio n Arm) Reference (Stop) Tri gger Arm Reference Trigger ...

  • National Instruments NI PCI-5114 - page 12

    NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis. For more inform ation about TClk synchronizati on, refer to the NI-TClk Synchronization Help , wh ich is located within the N I Hi ...

  • National Instruments NI PCI-5114 - page 13

    © National Instruments Corporat ion 13 NI PXI/PCI-5 114 Specifications W aveform Specifications Specif ication Va l u e Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channel — 64 MB per Channel Option 64 megasamples per channel 256 MB per Channel Option 256 megasamples per channel Minimum Record Length 1 Sample — Num ...

  • National Instruments NI PCI-5114 - page 14

    NI PXI/PCI-5114 Specifications 14 ni.com Calibration Power Specif ication Va l u e Self-Calibrati on Self-calibratio n is done on software command. The calibration corrects for gain, of fset, compensated 1 M Ω attenuator , triggerin g, and timing adjustment errors for all input ranges. External Calibration (Factory Calibration) The external calibr ...

  • National Instruments NI PCI-5114 - page 15

    © National Instruments Corporat ion 15 NI PXI/PCI-5 114 Specifications Software Specification Va l u e Dri ver Software NI PXI-5114 : NI-SCOPE 2.9 or later NI PCI-5114 : NI-SCOPE 3.1 or later NI-SCOPE is a n IVI-compliant dri ver that allo ws you to configure, control, and calibrate the NI 5114. NI-SCOPE provides application programming interfaces ...

  • National Instruments NI PCI-5114 - page 16

    NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note T o ensure that the NI PXI-5114 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only . Specif ication Va l u e Operating T emperature 0 ºC to +55 ºC in all N ...

  • National Instruments NI PCI-5114 - page 17

    © National Instruments Corporat ion 17 NI PXI/PCI-5 114 Specifications NI PCI-5114 Note T o ensure that the NI PCI-5114 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage. T o maximize airflo w and extend the life of the de vice, leav e any adjacent ...

  • National Instruments NI PCI-5114 - page 18

    NI PXI/PCI-5114 Specifications 18 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follow ing standards of safety for electrical e quipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safet ...

  • National Instruments NI PCI-5114 - page 19

    © National Instruments Corporat ion 19 NI PXI/PCI-5 114 Specifications Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er. NI recognizes that eliminating certa in hazardous substances from our products is beneficial not only to t he environment but also to ...

  • National Instruments NI PCI-5114 - page 20

    NI PXI/PCI-5114 Specifications 20 ni.com Dimensions and W eight NI PXI-5114 Front Panel Indicators Label Function Fo r mo re information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5114 to the controller . AC T I VE The ACTIVE LED indicates th e status ...

  • National Instruments NI PCI-5114 - page 21

    © National Instruments Corporat ion 21 NI PXI/PCI-5 114 Specifications NI PCI-5114 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 421 g (14.8 oz) 2.0 cm (0. 8 in.) 3 5.5 cm (14.0 in.) 11. 3 cm (4.4 in.) ...

  • National Instruments NI PCI-5114 - page 22

    NI PXI/PCI-5114 Specifications 22 ni.com Where to Go for Support The National Instruments W eb site is your complete resource for technical support. At ni.com/support you ha ve access to e verything from troubleshooting and ap plication dev elopment self-help resources to email and phone assistance from NI Application Engin eers. A Declaration of C ...

  • National Instruments NI PCI-5114 - page 23

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s. For patents covering Natio ...

Manufacturer National Instruments Category Graphics Tablet

Documents that we receive from a manufacturer of a National Instruments NI PCI-5114 can be divided into several groups. They are, among others:
- National Instruments technical drawings
- NI PCI-5114 manuals
- National Instruments product data sheets
- information booklets
- or energy labels National Instruments NI PCI-5114
All of them are important, but the most important information from the point of view of use of the device are in the user manual National Instruments NI PCI-5114.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals National Instruments NI PCI-5114, service manual, brief instructions and user manuals National Instruments NI PCI-5114. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product National Instruments NI PCI-5114.

A complete manual for the device National Instruments NI PCI-5114, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use National Instruments NI PCI-5114 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of National Instruments NI PCI-5114.

A complete National Instruments manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual National Instruments NI PCI-5114 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the National Instruments NI PCI-5114, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the National Instruments NI PCI-5114, that we can find in the current document
3. Tips how to use the basic functions of the device National Instruments NI PCI-5114 - which should help us in our first steps of using National Instruments NI PCI-5114
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with National Instruments NI PCI-5114
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of National Instruments NI PCI-5114 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning National Instruments NI PCI-5114?

Use the form below

If you did not solve your problem by using a manual National Instruments NI PCI-5114, ask a question using the form below. If a user had a similar problem with National Instruments NI PCI-5114 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)