  • Omron M6W Wide Cuff - page 1

    Check following components! HEM-7213-E_B_M01_120123.pdf Manufacturer OMRON HEAL THCARE Co., Ltd. 53, Kunotsubo, T erado-cho, Muko, Kyoto, 617-0002 JAP AN EU-representative OMRON HEAL THCARE EUROPE B.V . Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www .omron-healthcare.com Production facility Site de production Herstellungswerk Stabilimento di p ...

  • Omron M6W Wide Cuff - page 2

    Content s Thank you for purchasing the OMRON M6W Automatic Blood Pressure Monitor . The OMRON M6W is a compact, fully automatic blood pressure monitor , operating on the oscillometric principle.It measures your blood pressure and pulse rate simply and quickly . For comfortable controlled in fl ation without the need of pressure pre-setting or re-i ...

  • Omron M6W Wide Cuff - page 3

    1 EN Import an t Safety I nformation Consult your doctor prior to using in pregnancy or if d iagnosed with arrhyt hmia or arterios clerosis. Please read t his section carefully b efore using the unit. War ni n g : • Indicates a pot entially hazardous situation which, if not avo ided, could result in deat h or serious injury . (General Usage) • ...

  • Omron M6W Wide Cuff - page 4

    2 Impor tant Safet y Inform ation (AC Adapter Usage) • Use only the original AC adapter (optional) designed f or this unit. Use of unsupported adapters may damage and/or may be hazardous to the unit. • Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap. • Do not use the AC adapter if the unit or the power cord ...

  • Omron M6W Wide Cuff - page 5

    3 EN 1. Overview Main unit Display Memory button O/I ST ART ( ) button Up/Down ( ) buttons USE R ID Se lectio n Sw itch (A / B) Date/T ime setting ( ) button Air Jack Battery compartment AC adapter jack ( for optional AC adapter) Arm cuff Arm cuff (Wide Range Cuff: arm circumferenc e 22-42 cm) Marker Air tube Air plug Display Date/T ime display USE ...

  • Omron M6W Wide Cuff - page 6

    4 2. Prep aration 2.1 Inst alling/Replaci ng the Batteries 1. Remove the battery cover . 2. Insert four “AA” batteries as indicated in the battery comp artment and then replace the battery cover . Not es: • If the ba ttery low symbol ( ) appea rs on the display , turn the uni t off t hen replace all batter ies at the same time. • The me asu ...

  • Omron M6W Wide Cuff - page 7

    5 2. Prep arat ion EN 2.2 Setting the Date and T ime 1. Pres s th e butt o n. 2. Set the monitor to the correct date and time before t aking a measureme nt for the first time. 3. Press t he O/I ST ART butt on to store the setting. Not es: • If the batteries have been removed for 30 seco nds or more, the date and time setting will need to be reset ...

  • Omron M6W Wide Cuff - page 8

    6 3. Using the Unit 3.1 Appl ying the Ar m Cuff Remo ve tight- fittin g clothin g or tigh t rolled up sleev e from y our uppe r arm. Do no t place th e cuff over th ick clothe s. 1. Insert the air plug into the ai r jack . 2. Put your arm through the cuff loop . 3. Position the arm correctly . The bo ttom ed ge of the cuff should b e 1 to 2 c m abo ...

  • Omron M6W Wide Cuff - page 9

    7 3. Using the Unit EN 3.2 How to Sit Correctly T o take a m easurem ent, you n eed to b e relax ed and com fortably s eated, under comfor table room t emperatur e. No b athing, dr inking alcoh ol or caffein e, smoking, exerci sing or eating 30 minutes before taking a m easurem ent. • Sit on a chair with you r feet flat on the f loor . • Sit up ...

  • Omron M6W Wide Cuff - page 10

    8 3. Using the Unit 3.3 T a king a Re adin g Not es: • T o can cel a m easurem ent, press th e O/I ST A RT b utton t o turn the uni t of f and to rel ease the air in t he arm cuff. • Remain still wh ile t aking a meas urement. The moni tor is design ed to take measure ments and store the measureme nt valu es in the memo ry for two peop le using ...

  • Omron M6W Wide Cuff - page 11

    9 3. Using the Unit EN Dual Check System Press the O/I ST ART button, then the Dual Check System lamp lights and the unit starts detecting a device error . If the unit is accurate and functioning correctly , the Dual Check System lamp remains lit dur ing the meas uremen t. If an e rror is detected , the D ual Chec k Sy stem lam p flash es an d “E ...

  • Omron M6W Wide Cuff - page 12

    10 3. Using the Unit • Y our blood pressure monitor includes an irregular heart beat feature. Irregular hear tbeats can influence the results of t he measur ement. The irregular heartbeat algorithm automatically determines if the m easurement is usable or needs t o be repeated. If the measurement results are affected by irregular heartbeats but t ...

  • Omron M6W Wide Cuff - page 13

    11 3. Using the Unit EN 3.4 Using the Memo ry Func tion The m onitor au tomatical ly stores the res ult up to 1 00 sets for e ach user (A and B). It can al so ca lculate an avera ge readi ng based o n the m easurem ents from the last three rea dings taken within 10 min utes. I f there ar e only tw o read ings in memor y for that per iod, th e avera ...

  • Omron M6W Wide Cuff - page 14

    12 3. Using the Unit T o V iew the R eadings S tored in Memory 1. Press the button, while t he average value is dis played. The M emory nu mber appe ars for a secon d before the pul se rate is disp layed . The newe st set i s numb ered “1”. 2. Press the or button to view the readings stored in memory . : T o the older readings : T o the more re ...

  • Omron M6W Wide Cuff - page 15

    13 EN 4. T roubl eshooting and Maint enance 4.1 The Ico ns and Error Mes sages Error Display Cause Remedy Irregular heartbeats are detected. Remove the arm cuff. W ait 2-3 minutes and then t ake another measureme nt. Repeat the steps in section 3.3. If this error continues to appear , contact your doctor . Movement durin g measurement. Carefully re ...

  • Omron M6W Wide Cuff - page 16

    14 4. Troub leshoot ing and Maintena nce Note: The irregular hear tbeat symbol ( ) may also be displayed with er ror messages. Air pl ug d is con nec ted . Insert the plug securely . Refe r to section 3.1. Arm cuf f not applied correctly . Apply the arm cuff correctly . Refe r to section 3.1. Air is leaking from the arm cuf f. Repla ce the cuff wit ...

  • Omron M6W Wide Cuff - page 17

    15 4. Troub leshoot ing and Maintena nce EN 4.2 T roubleshoo ting Problem Cause Remedy The reading is extremely low (or high). Arm cuff not applied correctly . Apply the arm cuff correctly . Refer to section 3.1. Movement or talking during measurement. Remain still and do not talk during measurement. Refer to section 3.3. Clothing is interfering wi ...

  • Omron M6W Wide Cuff - page 18

    16 4. Troub leshoot ing and Maintena nce Nothing happens when you pres s the buttons. The batteries are empty . Replace the batteries with new ones. Refer to section 2.1. The batteries have been inserted incorrectly . Insert the batteries wi th the corre ct (+/ -) polarit y . Refer to section 2.1. Other problems. • Press t he O/I ST A RT button a ...

  • Omron M6W Wide Cuff - page 19

    17 4. Troub leshoot ing and Maintena nce EN 4.3 Maintenance T o protec t you r unit from damage , please o bserve the follo wing: • Do not subject the main unit and the cuf f to extreme te mperatures, humidity , moisture or direct sunlight. • Do not fold the cuff or tubing tightly . • Do not disassemble the u nit. • Do not subject the unit ...

  • Omron M6W Wide Cuff - page 20

    18 4. Troub leshoot ing and Maintena nce 4.4 St or a g e Keep th e unit i n its stor age case when no t in us e. 1. Unplug the air plu g from the air jack. 2. Gently fold the air tube into the arm cuff. Note: Do not bend t he air tube exc essively . 3. Place t he arm cuff and main unit in the storage case. Do no t store th e unit in the foll owing ...

  • Omron M6W Wide Cuff - page 21

    19 EN 5. Optional Part s Using th e Optio nal AC Adapter 1. Insert the AC a dapter plug into the AC adapter jack on the rear s ide of the main unit. 2. Plug the AC adapter into an electrical outlet. T o dis connec t the AC adapter , unplu g the AC adapter f rom th e electric al out let firs t and then remo ve the AC adapter plug fro m the m ain uni ...

  • Omron M6W Wide Cuff - page 22

    20 6. T echnical Dat a Note: Subject to technical modifi cation without prior notice. Product Description Automatic Blood Pressure Monitor Model OMRON M6W(HEM-7213-E(V)) Display LCD Digital Display Measurement Method Oscillometric method Measurement Range Pressure : 0 to 299 mmHg Pulse: 40 to 180/min. Accuracy Pressure: ± 3 mmHg Pulse: ± 5% of di ...

  • Omron M6W Wide Cuff - page 23

    21 6. Techni cal Data EN • This device fulfil s the p rovis ions of EC dire ctiv e 93/42 /EEC (Me dical Device Directi ve). • This blood pres sure moni tor is desig ned acc ording to the Europea n S tandard E N1060, Non - invasiv e sphygm omanom eters P art 1: Ge neral Re quiremen ts and Par t 3: Supp lementary requir ements fo r electro mechan ...

  • Omron M6W Wide Cuff - page 24

    22 6. Techni cal Data Import ant in formation regardi ng Electro Mag netic Comp atibility (EMC) With the increased num ber of e lectronic devices such as PC’s and mobile (c ellular) telephones, medical devic es in use may be suscep tible to electromagnetic interference from other dev ices. Electromagnetic interference may result in incorrect oper ...

  • Omron M6W Wide Cuff - page 25

    23 EN 7. Some Use f ul I nformation about Blood Pressure What is Blood Pressure? Blood pres sure i s a meas ure of the force of blood flowing again st the walls of t he arter ies. Arterial blood pr essure is con stantly chan ging dur ing the c ourse of the hea rt’s cycle. The hig hest pre ssure in t he cy cle is ca lled th e Systoli c Bloo d Pres ...

  • Omron M6W Wide Cuff - page 26

    24 7. Some Useful In formation a bout Bl ood Pressur e Why is it a Good Thing to me asure Blood Pressure at Home? Many fac tors s uch as ph ysic al acti vity , anxiety , or the time of day , ca n influ ence your blo od pressure . A single measur ement may not be s ufficient for an accu rate diag nosis. T hus it i s best to try and measure your blo ...

メーカー Omron カテゴリー Sphygmomanometer

Omron M6W Wide Cuffのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- M6W Wide Cuffの取扱説明書
- Omronの製品カード
- パンフレット
- またはOmron M6W Wide Cuffの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Omron M6W Wide Cuffの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Omron M6W Wide Cuffの取り付け説明書、サービスマニュアル、簡易説明書、またはOmron M6W Wide Cuffのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Omron M6W Wide Cuffの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Omron M6W Wide Cuffデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがOmron M6W Wide Cuffを使用するのを助ける技術的文書のことです。説明書は通常、全てのOmron M6W Wide Cuffユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Omronの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Omron M6W Wide Cuffの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははOmron M6W Wide Cuffの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Omron M6W Wide Cuffに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Omron M6W Wide Cuffデバイスの基本機能を使うにあたってのヒント。 Omron M6W Wide Cuffのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Omron M6W Wide Cuffに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるOmron M6W Wide Cuffのメーカー/サービスへの連絡先に関する情報。

Omron M6W Wide Cuffについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもOmron M6W Wide Cuffの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがOmron M6W Wide Cuffで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください