Manual National Instruments SCXI-1321

16 pages 0.37 mb
Download

Go to site of 16

Summary
  • National Instruments SCXI-1321 - page 1

    © 1998–2003 National Instruments Corp. All rights reserved. LabVIEW™, National Instru ments™, NI™, ni.com™, NI-DAQ ™, and SCXI™ are t rademarks of National Instrument s Corporation. Product and com pany names mentioned herein are trademarks or trade names of their respective companies. For patent s covering National Instruments produ ...

  • National Instruments SCXI-1321 - page 2

    SCXI-1321 T erminal Block Installation Guide 2 ni.com italic Italic text denotes variables, emphasis, a cross reference, or an int roduction to a key concept. This font also denote s te xt that is a placeholder for a word or v alue that you must supply . monospace T ext in this font denotes text or ch aracters that you should enter from the keyboar ...

  • National Instruments SCXI-1321 - page 3

    © National Instruments Cor poration 3 SCXI-1321 T erminal Block Installa tion Guide Connecting the Signals Note Refer to the Read Me First: Safety and Radio-Frequency Interference document before removing equipment covers, or connecting or disconnecting any signal wires. To connect the signal to the terminal block, complete the following steps, re ...

  • National Instruments SCXI-1321 - page 4

    SCXI-1321 T erminal Block Installation Guide 4 ni.com Figure 1. SCXI-132 1 Parts Locator Diagram 1 Strain-Relief Bar 2 Strain-Relief Screws 3 Safety-Ground Lug 4 Mating Connector 5 Thumbscrew 6 Top-Cover Screws 7 Top Cover Back Vie w F ront View 2 1 3 4 5 7 6 ...

  • National Instruments SCXI-1321 - page 5

    © National Instruments Cor poration 5 SCXI-1321 T erminal Block Installa tion Guide Figure 2. SCXI-1321 Circuit Parts Locator Diagram Installing the T erminal Block To connect the terminal block to the SCXI mo dule front connector, complete the follow ing steps: 1. Connect the module front connector to its connector on the terminal block. 2. Make ...

  • National Instruments SCXI-1321 - page 6

    SCXI-1321 T erminal Block Installation Guide 6 ni.com Note T o minimize the temperature gradient inside the termin al block, move the SCXI chassis aw ay from any extr eme temperature dif ferential. Specifications All specifications are typical at 25 °C unless otherwise specified . Electrical Cold-junction sensor Accuracy 1 ........... ............ ...

  • National Instruments SCXI-1321 - page 7

    © National Instruments Cor poration 7 SCXI-1321 T erminal Block Installa tion Guide Environmental Operating temperature........................ .... 0 to 50 °C Storage temperature ............. ................. . –20 to 70 °C Humidity ............. .............. .............. ....... 10 to 90% RH, noncondensing Maximum altitude ............ ...

  • National Instruments SCXI-1321 - page 8

    SCXI-1321 T erminal Block Installation Guide 8 ni.com Note Refer to the Declaration of Conformity (DoC) for this product fo r any additional regulatory compliance in formation. T o obtain th e DoC for this p roduct, click Declarations of Conformity Information at ni.com/hardref.nsf/ . Performed or Supported Signal Conditioning This section provides ...

  • National Instruments SCXI-1321 - page 9

    © National Instruments Cor poration 9 SCXI-1321 T erminal Block Installa tion Guide The factory default v alue of all the nulling resistors on the terminal blo ck is 39 k Ω . Note These resistors are socketed for easy re placement. These sockets best f it a 1/4 W resistor lead size. T o determ ine the nulling range, use the f ollowing formula wh ...

  • National Instruments SCXI-1321 - page 10

    SCXI-1321 T erminal Block Installation Guide 10 ni.com Assuming a strain-gauge range with a gauge factor of GF = 2 and a quarter-bridge conf iguration, this range corresponds to ±1,498 µε , as given by the strain formula for a quarter-bridge strain-gauge configuration: where Figure 3. Quarter-Bridge Nulling Circuit ε 4 V r – GF 12 V r + () -- ...

  • National Instruments SCXI-1321 - page 11

    © National Instruments Cor poration 11 SCXI-1321 T erminal Bloc k Installa tion Guide Figure 4. Half-Bridge Nulling Circuit CH+ CH+ CH– EX+ EX– + – + – R 1 R 2 4.5 k Ω 4.5 k Ω R SCAL 301 k Ω R Tr i m 10 k Ω R Null 39.1 k Ω (Screw Adjusts P otentiometer) R L R L R L R 4 (gauge) EX+ CH+ CH+ CH– EX+ EX– R Tr i m SCXI-1121 SCXI-1 ...

  • National Instruments SCXI-1321 - page 12

    SCXI-1321 T erminal Block Installation Guide 12 ni.com Figure 5. Full-Bridge Nulling Cir cuit Shunt Calibration Shunt calibration prov ides a method of adjusti ng for gain error to improve accuracy. The shunt-calibration circui try configuration places a shunting resistor in parallel with the resis tive element connected between EX+ and CH+ (elemen ...

  • National Instruments SCXI-1321 - page 13

    © National Instruments Cor poration 13 SCXI-1321 T erminal Bloc k Installa tion Guide z is the module slot of the SCXI-1121 w is the channel o f the module that you want to engage th e shunt and take measurements For e xample, if you want to measure th e volta ge at channel 0 with the shunt resistor enabled, use the SCXI channel string ob0 ! sc1 ! ...

  • National Instruments SCXI-1321 - page 14

    SCXI-1321 T erminal Block Installation Guide 14 ni.com gauge with a gauge factor of GF = 2, V exc = 3.3 V , and R = 120 Ω prod uces the following: V change = 0.3321 mV Replacing the strained voltage by V change in the quarter-bridge strain equation produces an equ i valent 199 µε of change. Refer to the NI-DAQ User Manual for more informat ion ...

  • National Instruments SCXI-1321 - page 15

    © National Instruments Cor poration 15 SCXI-1321 T erminal Bloc k Installa tion Guide Y ou can read a temperature at the begi nning of the test and use that value with the data that follows. Using this method assumes that there are no temperature v ariations during the measurement period. If there are temperature variations, the measure m ents can ...

  • National Instruments SCXI-1321 - page 16

    321924C-01 J an03 *321924C-01* W3 Nulling circuit of Channel 2 is enabled (factory default setting) Nulling circuit of Channel 2 is disabled W4 Nulling circuit of Channel 3 is enabled (factory default setting) Nulling circuit of Channel 3 is disabled Table 4. Jumper Settings of the Nu lling Circuits (Continued) Jum p e r Pos i t i o n Description D ...

Manufacturer National Instruments Category Switch

Documents that we receive from a manufacturer of a National Instruments SCXI-1321 can be divided into several groups. They are, among others:
- National Instruments technical drawings
- SCXI-1321 manuals
- National Instruments product data sheets
- information booklets
- or energy labels National Instruments SCXI-1321
All of them are important, but the most important information from the point of view of use of the device are in the user manual National Instruments SCXI-1321.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals National Instruments SCXI-1321, service manual, brief instructions and user manuals National Instruments SCXI-1321. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product National Instruments SCXI-1321.

A complete manual for the device National Instruments SCXI-1321, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use National Instruments SCXI-1321 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of National Instruments SCXI-1321.

A complete National Instruments manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual National Instruments SCXI-1321 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the National Instruments SCXI-1321, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the National Instruments SCXI-1321, that we can find in the current document
3. Tips how to use the basic functions of the device National Instruments SCXI-1321 - which should help us in our first steps of using National Instruments SCXI-1321
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with National Instruments SCXI-1321
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of National Instruments SCXI-1321 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning National Instruments SCXI-1321?

Use the form below

If you did not solve your problem by using a manual National Instruments SCXI-1321, ask a question using the form below. If a user had a similar problem with National Instruments SCXI-1321 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)