    FPD1000-HP SERIES HIGH PRESSURE Positive Displacement Flowmeters ...

  • Omega HIGH FPD1000-HP - page 2

    ...

  • Omega HIGH FPD1000-HP - page 3

    1. Use thread sealant on all pipe thr eads. 2. Ensur e the meter is installed so that ro tor shafts are always in a horizontal plane (See Figure 1). Flow is bi-dir ec- tional. ▲ ! To t h e owner: Thank you for purchasing an OMEGA FPD Series Flowmeter . Please take a few min- utes to read thr ough the manual befor e in- stalling and operating your ...

  • Omega HIGH FPD1000-HP - page 4

    Reed Relay Specifications •T wo wire SPST N/O. •S witching voltage 150 VDC; maximum current 0.25 Amps. •R a t i ng 3 watts. •D uty cycle 20% ON, 80% OFF . Disassembly: 1. Ensure the fluid supply to the meter has been disconnected, and the line pres- sure has been r eleased before disas- sembly . 2. Remove the four screws and r emove the pul ...

  • Omega HIGH FPD1000-HP - page 5

    3 Symptom Probable Cause Corrective Action FLUID WILL NOT 1. Foreign matter blocking Dismantle meter , clean rotors. Strainer must FLOW THROUGH rotors be fitted in-line. THE METER 2. Line strainer blocked Clean strainer . 3. Damaged rotors Replace rotors. Strainer must be fitted in-line. 4. Meter connections over - Re-adjust connections. tightened ...

  • Omega HIGH FPD1000-HP - page 6

    4 DISPLA Y P AR TS LISTING Model: FPD1202-HP Item Rec. Part or Set No. Qty . Parts (Order fr om this column only) Part Description 11 MS1NS Meter Body Assembly NPT 28 u MS367S Bolt Set - Socket Head 31 MS170H Pulser Cap (BSP) 31 MS170HN Pulser Cap (NPT) 44 uM S115S Bolt Set - Socket Head 61 MS3HES Meter Cap & Hall Effect Assembl y * 61 MS3HPS M ...

  • Omega HIGH FPD1000-HP - page 7

    5 SPECIFICA TIONS Pulse Meter Models Flow Ranges (LPM or GPM) Above 5 centipoise 2 to 100 / 0.53 to 26.4 Below 5 centipoise 5 to 100 / 1.32 to 26.4 Accuracy of Reading ± 1% Maximum V iscosity 1000 Centipoise Maximum Operating Pr essure 55160 kPa / 8000 PSI / 551 Bar Maximum Operating T emperature 120 ° C / 248 ° F Pulse T ype Hall Effect Sensor ...

  • Omega HIGH FPD1000-HP - page 8

    6 DISPLA Y P AR TS LISTING Model: FPD1203-HP Item Rec. Part or Set No. Qty . Parts (Order fr om this column only) Part Description 11 MS2NS Meter Body Assembly NPT 28 u MS367S Bolt Set - Socket Head 31 MS170H Pulser Cap (BSP) 31 MS170HN Pulser Cap (NPT) 44 uM S115S Bolt Set - Socket Head 61 MS3HES Meter Cap & Hall Effect Assembl y * 61 MS3HPS M ...

  • Omega HIGH FPD1000-HP - page 9

    7 SPECIFICA TIONS Pulse Meter Models Flow Ranges (LPM or GPM) Above 5 centipoise 15 to 500 / 4.0 to 132 Below 5 centipoise 25 to 500 / 6.0 to 132 Accuracy of Reading ± 1% Maximum Viscosity * 1000 Centipoise Maximum Operating Pr essure 55160 kPa / 8000 PSI / 551 Bar Maximum Operating T emperature 120 ° C / 248 ° F Pulse T ype Hall Effect Sensor / ...

  • Omega HIGH FPD1000-HP - page 10

    ...

  • Omega HIGH FPD1000-HP - page 11

    ...

  • Omega HIGH FPD1000-HP - page 12

    M- 41 2 5 / 0 4 05 ...

Manufacturer Omega Category Automobile Parts

Documents that we receive from a manufacturer of a Omega HIGH FPD1000-HP can be divided into several groups. They are, among others:
- Omega technical drawings
- HIGH FPD1000-HP manuals
- Omega product data sheets
- information booklets
- or energy labels Omega HIGH FPD1000-HP
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega HIGH FPD1000-HP.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega HIGH FPD1000-HP, service manual, brief instructions and user manuals Omega HIGH FPD1000-HP. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega HIGH FPD1000-HP.

Similar manuals

A complete manual for the device Omega HIGH FPD1000-HP, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega HIGH FPD1000-HP by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega HIGH FPD1000-HP.

A complete Omega manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega HIGH FPD1000-HP - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega HIGH FPD1000-HP, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega HIGH FPD1000-HP, that we can find in the current document
3. Tips how to use the basic functions of the device Omega HIGH FPD1000-HP - which should help us in our first steps of using Omega HIGH FPD1000-HP
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega HIGH FPD1000-HP
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omega HIGH FPD1000-HP in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omega HIGH FPD1000-HP?

Use the form below

If you did not solve your problem by using a manual Omega HIGH FPD1000-HP, ask a question using the form below. If a user had a similar problem with Omega HIGH FPD1000-HP it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)