    www .omega.com e-mail: info@omega.com U ser ’ s Guide HX93A, HX93DA SERIES RH/T emperature T ransmitter MADE IN Shop online at ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 2

    Servicing Nor th America: USA: One Omega Drive, Box 4047 ISO 9001 Certified Stamford CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca For immediate technical or application assistance: USA and Canada: Sales Ser ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 3

    TABLE OF CONTENTS HX93A, HX93DA SERIES RH/Temperature Transmitter i Section Page 1. General Description .................................................................... 1 2. Unpacking .................................................................................... 1 3. Theory of Operation .................................................... ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 4

    FIGURES HX93A, HX93DA SERIES RM/Temperature Transmitter ii Figure Page Figure 1 Basic Transmitter Set-up With Current Loop Output .............................................. 2 Figure 2 Wall Mount Model Dimensions ...................................... 3 Figure 3 Duct Mount And Remote Probe Model Dimensions ....................................... ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 5

    1 HX93A, HX93DA SERIES RH/Temperature Transmitter 1 1. General Description The OMEGA ® HX93A and HX93DA Series Relative Humidity/ Temperature Transmitter’s provide a linearized and temperature compensated output signal of 4 to 20 mA or 0 to 1 Vdc depending upon the model selected for both relative humidity and temperature measurement. The output ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 6

    2 HX15AV-W Wall mount “High Ambient Temperature” application RH/Temperature transmitter (dual 0 to 1 Vdc Output) HX15AC-W Wall mount “High Ambient Temperature” application RH/Temperature transmitter (dual 4 to 20 mA Output) HX15AV-D Duct mount “High Ambient Temperature” application RH/Temperature transmitter (dual 0 to 1 Vdc Output) HX1 ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 7

    3 Complimentary Instruments Power Supply, OMEGA ® Part No.: PSU-93 Csi32 Series Panel Meters and Controllers Recommended Accessories Shielded Transmitter Cable, OMEGA ® Part No.: TX4-100 (100 ft) RH Calibration Kit, OMEGA ® Part No.: HX92-CAL 4. Mounting OMEGA’s HX93A and HX93DA transmitter’s are designed for either wall, duct or remote prob ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 8

    4 5. Electrical Connections All electrical connections and wiring should be performed by a suitably trained professional only. Models: HX93AC, HX93AC-D, HX93AC-RP1, HX93DAC, HX93DAC- D,HX93DAC-RPI (Current Output) TB1 - Relative Humidity TB2 - Temperature 1. + Power Supply 1. + Power Supply 2. No Connection 2. No Connection 3. 4-20 mA Output 3. 4-2 ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 9

    5 Models: HX93AV, HX93AV-D HX93AV-RP1 (Voltage Output) TB1 - Relative Humidity TB2 - Temperature 1. + Power Supply 1. + Power Supply 2. + Output 2. + Output 3. - Power Supply 3. - Power Supply 7. Voltage Transmitter Wiring Examples For voltage output models (0 to 1 Vdc) Figure 5 - Voltage Transmitter Wiring Example 8. Relative Humidity Output Calcu ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 10

    6 9. RH Measured Vs Output Reading 10. Temperature Output Calculations To calculate Temperature by measuring the current or voltage output use the following formulas. For Current output in ºC (-20 to 75ºC) ºC = (Output measured in miliamps – 4) ÷ .1684 - 20 = ºC EXAMPLE: (12.0 mA – 4) ÷ .1684 - 20 = 27.5ºC For Current output ºF (-4 to 1 ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 11

    7 11. Calibration Your transmitter has been factory calibrated to meet or exceed the specifications outlined in this manual. To maintain original specifications it is generally recommended that your transmitter be re- calibrated on an annual basis depending on operating conditions. 12. Calibration Procedure for HX93AC (all styles) Relative Humidity ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 12

    8 13. Calibration Procedure for HX93AC (all styles) Temperature Adjustment Recommended equipment: • RTD Simulator OMEGA ® Model No.: CL510-1 • Handheld Digital Multi-meter, OMEGA ® Model No.: HHM29 • Dc Power Supply, OMEGA ® Model No.: PSU-93 1. Remove enclosure cover 2. Connect the power supply and multimeter as shown in the figure below. ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 13

    9 14. Calibration Procedure for HX93AV (all styles) Relative Humidity Adjustment Recommended equipment: • Humidity Calibration Kit, OMEGA ® Model No.: HX92-CAL • Handheld Digital Multi-meter, OMEGA ® Model No.: HHM29B • Dc Power Supply, OMEGA ® Model No.: PSU-93 1. Remove enclosure cover. 2. Connect transmitter as shown in figure below. 3. ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 14

    10 15. Calibration Procedure for HX93AV (all Styles) Temperature Adjustment Recommended equipment: • 100 Ohm RTD Simulator OMEGA ® Model No.: CL510-1 • Handheld Digital Multi-meter, OMEGA ® Model No.: HHM29 • Dc Power Supply, OMEGA ® Model No.: PSU-93 1. Remove enclosure cover 2. Connect the power supply and multimeter as shown in the figu ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 15

    11 15A. Display Scaling for HX93DAC (all styles) Your HX93DA transmitter comes with dual built-in LCD indicators for displaying both Humidity and Temperature. Before the meter adjustment or re-scaling can be done the transmitters current output should be checked as per the calibration procedure in Section 12 and 13 to verify the display will be adj ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 16

    HX93A, HX93DA SERIES RH/Temperature Transmitter 15A 12 Temperature Display Scaling Recommended equipment: • RTD Simulator OMEGA ® Model No.: CL510-1 • Handheld Digital Multi-meter, OMEGA ® Model No.: HHM29 • Dc Power Supply, OMEGA ® Model No.: PSU-93 For ºF Display 1. Remove enclosure cover 2. Connect the power supply and multimeter as sh ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 17

    13 16. Maintenance If your Humidity transmitter will be used in a dusty environment, the protective sensor filter, if clogged, may be removed for cleaning. Unscrew the protective cover and gently blow compressed air through the filter screen. A soft brush may also be used to remove dirt particles from the screen. If the sensor is subjected to 100% ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 18

    14 Max Loop Resistance: Ohms = (V supply – 6 V)/.02 A Sensor Type: 100 Ohm Platinum RT 18. General Specifications Enclosure Housing: Gray Polycarbonate, (IP 65, DIN 40050) NEMA rated up to 13, UL Listed. Connections Cable Strain Relief: Nylon, Liquid-tight with neoprene gland for 2.29 to 6.73 mm (.09 to .265") diameter cable. Wire Connection ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 19

    W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers ...

  • Omega Engineering RH/Temperature Transmitter HX93A - page 20

    Where Do I Find Ever ything I Need for Pr ocess Measurement and Contr ol? OM EGA…Of Course! Shop online at www .omega.com TEMPERA TURE   Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies   Wire: Thermocouple, RTD & Thermistor   Calibrators & Ice Point References   Recorders, Controllers ...

Производитель Omega Engineering Категория Network Router

Документы, которые мы получаем от производителя устройства Omega Engineering RH/Temperature Transmitter HX93A мы можем разделить на несколько групп. Это в частности:
- технические чертежи Omega Engineering
- инструкции обслуживания RH/Temperature Transmitter HX93A
- паспорта изделия Omega Engineering
- информационные брошюры
- энергетические этикетки Omega Engineering RH/Temperature Transmitter HX93A
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Omega Engineering RH/Temperature Transmitter HX93A.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Omega Engineering RH/Temperature Transmitter HX93A, инструкции обслуживания, короткие инструкции или инструкции пользователя Omega Engineering RH/Temperature Transmitter HX93A. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Omega Engineering RH/Temperature Transmitter HX93A.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Omega Engineering RH/Temperature Transmitter HX93A, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Omega Engineering RH/Temperature Transmitter HX93A пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Omega Engineering RH/Temperature Transmitter HX93A.

Полная инструкция обслуживания Omega Engineering, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Omega Engineering RH/Temperature Transmitter HX93A - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Omega Engineering RH/Temperature Transmitter HX93A, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Omega Engineering RH/Temperature Transmitter HX93A, которое найдем в данном документе
3. Советы по использованию основных функций устройства Omega Engineering RH/Temperature Transmitter HX93A - которые должны облегчить нам первые шаги во время использования Omega Engineering RH/Temperature Transmitter HX93A
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Omega Engineering RH/Temperature Transmitter HX93A
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Omega Engineering RH/Temperature Transmitter HX93A в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Omega Engineering RH/Temperature Transmitter HX93A?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Omega Engineering RH/Temperature Transmitter HX93A, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Omega Engineering RH/Temperature Transmitter HX93A со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)