  • Emerson 00825-0100-4593 - page 1

    www.rosemount.com ¢00825-0100-4360N¤ Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 Step 1: Mount the T ransmitter Step 2: Consider Housing Rotation Step 3: Connect the Wiring and Power Up Step 4: Set Switches Step 5: Configuration Step 6: T rim the T ransmitt er Product Certifications Sta r t End Rosemount 1 151 Pres ...

  • Emerson 00825-0100-4593 - page 2

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 2 W ARNING Explosions could result in death or serious injury: Installation of this tr ansmitter in a n explosive env iron ment must be in accordan ce wit h t he approp riat e local, national, and inter national st andards, code s, and pract ices. Please review the approval ...

  • Emerson 00825-0100-4593 - page 3

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 3 S TE P 1: M OUNT THE T RANSMI TTER A. Applications Liquid Flow Application s 1. Place taps to the side of the line. 2. Mount beside or below the taps. Gas Flow Applicatio ns 1. Place taps in the top or side of the line. 2. Mount beside or above the taps. Steam Flow Applic ...

  • Emerson 00825-0100-4593 - page 4

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 4 S TE P 1 CONTI NUED ... B. Optional Mounting Brackets When installing the transmitter to one of the optional mounting brackets, torque the bracket bolts to 125 in.-lbs. (0,9 N-m). Pipe Mount Panel Mount (1) (1) Panel bolts are customer supplied. Flat Mount 4593_Rev_BA.fm P ...

  • Emerson 00825-0100-4593 - page 5

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 5 S TE P 1 CONTI NUED ... C. O-rings with Flange Adapters W ARNING Whenever the flanges or adapters are removed, vi su ally inspect the o-rings. Replace them if there are any signs of damage, such as nicks or cuts. If you replace the o-rings, re-torque the flange bolts and ...

  • Emerson 00825-0100-4593 - page 6

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 6 S TE P 2: C ON SIDER H OUSING R OT ATION T o improve field access or to better view the optional LCD d isplay: 1. Loosen the housing lock nut. 2. Rotate the housing clockwise to the desired posit ion – up to 90° from its o riginal position. Over rotating will damage the ...

  • Emerson 00825-0100-4593 - page 7

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 7 S TE P 3: C ON NECT THE W IRING AND P OWER U P Use the following steps to wire the transmitter: 1. Remove the housing cover on the side marked TERMINALS on the namepla te. 2. Connect the positive lead to the “+” termin a l and the negative lead to the “–” termin ...

  • Emerson 00825-0100-4593 - page 8

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 8 S TE P 3 CONTI NUED ... Power Supply The dc power supply should provide power with less than tw o percent ripple. The total resistance load is the sum of the resistance of the signal leads and the load resistance of the controller , indicator , and related pieces. Note tha ...

  • Emerson 00825-0100-4593 - page 9

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 9 S TE P 4: C ON FIGURE THE T RANSMITTER Failure Mode Alarm Switch 1. Remove the housing cover . 2. Locate the failure mode switch (see Figure 3). 3. Move the switch to the desired alarm setting. T o set the failure mo de to high alarm, position the switch toward “HI.” ...

  • Emerson 00825-0100-4593 - page 10

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 10 S TE P 5: C AL IBRATE THE T RANSMITTER V erify T ransmitter Configuration NOTE: A check ( ✓ ) indicates the basic configuration para meters. At minimum, these parameters should be verified as part of the configuration and startup procedure. T able 1. HART Communicator F ...

  • Emerson 00825-0100-4593 - page 11

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 11 S TE P 5 CONTI NUED ... Configure LCD Displa y Figure 4. Sample 1 151 LCD Display NOTE The LCD display time-out is approximately 16 se conds. If keys ar e not pressed within this period, the indicator reverts to reading the signal. Position the Decimal Point and Select t ...

  • Emerson 00825-0100-4593 - page 12

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 12 S TE P 5 CONTI NUED ... Set the Display Equivalent to a 4 mA Sign al 1. Unscrew the retaining ring shown in Figure 4 and remove the LCD display cover . 2. Press the left button for two seconds. 3. T o decrement the display numb ers, press the le ft configuration button an ...

  • Emerson 00825-0100-4593 - page 13

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 13 S TE P 6: T RIM THE T RANSMITTE R NOTE T ransmitte rs are shipped fully calibrated per reques t or by the factory d efault of full scale (span = upper range limit). Full T rim A full trim is a two-point sensor calibrat ion where two end- point pressures are applied, and ...

  • Emerson 00825-0100-4593 - page 14

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 14 S TE P 6 CONTI NUED ... Using the T ransmitter Zero Adjustment Buttons Perform the following steps to perform a rerange using the zer o adjustment buttons (see Figure 3). 1. Apply a pressure equivalent to the lower calibra ted value on the high side of the transmitter . 2 ...

  • Emerson 00825-0100-4593 - page 15

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 15 P RODUCT C E RTIFICATI ONS Approved Manufacturi ng Locations Rosemount Inc. — Chanhassen, Minnesota, USA Fisher-Rosemount Gmb H & Co. — Wessling, Germany Emerson Process Management Asia Pacific Private Limited — Singapore Beijing Rosemount Far East Instrument C ...

  • Emerson 00825-0100-4593 - page 16

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 16 Canadian Certifications Canadian St andards Association (CSA) Approvals E6 Explosion Proof for Class I, Division 1, Gr oups C and D; Class II, Division 1, Group s E, F , and G; Class III, Division 1 Hazardous Loc ations. Suit able fo r C la ss I , Di v is i on 2, Groups A ...

  • Emerson 00825-0100-4593 - page 17

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 17 N1 A TEX T y pe N and Dust Certification Certificate Number: BAS99A TEX3293X A TEX Marking: II 3 GD Ex nL IIC T5 (-40 °C  T amb  40 °C) EX nL IIC T4 (-40 °C  T amb  80 °C) Dust Rating: T90 °C (T amb = -20 °C to 40 °C) Ui = 45 Vdc ma ximum IP66 Special ...

  • Emerson 00825-0100-4593 - page 18

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 18 N7 T ype N Certificate Number: Aus Ex 122X Ex n IIC T5 (T a = 80 °C) / T6 IP66 Special Conditions for safe use (x): 1. It is a condition of safe use that a nomi nal voltage of 30 V for Ex n application shall not be exceeded. 2. It is a condition of safe use that when the ...

  • Emerson 00825-0100-4593 - page 19

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 19 DoC 4593_Rev_BA.fm Page 19 Friday, August 7, 2009 3:38 PM ...

  • Emerson 00825-0100-4593 - page 20

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 20 4593_Rev_BA.fm Page 20 Friday, August 7, 2009 3:38 PM ...

  • Emerson 00825-0100-4593 - page 21

    Quick Installation Guid e 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 21 4593_Rev_BA.fm Page 21 Friday, August 7, 2009 3:38 PM ...

  • Emerson 00825-0100-4593 - page 22

    Quick Installation Guide 00825-0100-4593, Rev BA July 2009 Rosemount 1 151 22 4593_Rev_BA.fm Page 22 Friday, August 7, 2009 3:38 PM ...

メーカー Emerson カテゴリー Sleep Apnea Machine

Emerson 00825-0100-4593のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 00825-0100-4593の取扱説明書
- Emersonの製品カード
- パンフレット
- またはEmerson 00825-0100-4593の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Emerson 00825-0100-4593の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Emerson 00825-0100-4593の取り付け説明書、サービスマニュアル、簡易説明書、またはEmerson 00825-0100-4593のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Emerson 00825-0100-4593の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Emerson 00825-0100-4593デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEmerson 00825-0100-4593を使用するのを助ける技術的文書のことです。説明書は通常、全てのEmerson 00825-0100-4593ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Emersonの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Emerson 00825-0100-4593の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEmerson 00825-0100-4593の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Emerson 00825-0100-4593に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Emerson 00825-0100-4593デバイスの基本機能を使うにあたってのヒント。 Emerson 00825-0100-4593のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Emerson 00825-0100-4593に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEmerson 00825-0100-4593のメーカー/サービスへの連絡先に関する情報。

Emerson 00825-0100-4593についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもEmerson 00825-0100-4593の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがEmerson 00825-0100-4593で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)